迈向无透视介入:使用径向心腔内超声进行血管导航

Toward Fluoro-Free Interventions: Using Radial Intracardiac Ultrasound for Vascular Navigation.
PubMed 文献 · Ultrasound Med Biol · 2022-04-27

经导管心血管介入依赖影像引导导航下腔静脉,传统透视技术存在辐射风险;本研究在体模中采用空间追踪心腔内超声结合电磁追踪实时重建血管三维路径并导航导丝,无需辐射成像;以CT扫描为参考, Dice系数为0.79,两模型表面平均距离为1.7 ± 1.12 mm。

要点

  • 心腔内超声结合电磁追踪可在体模中实时重建血管三维路径并引导导丝导航。
  • 与CT参考模型比较,重建血管的Dice系数为0.79,表面平均距离1.7 ± 1.12 mm。
  • 该技术旨在替代透视引导下腔静脉导航,以减少患者和术者的辐射暴露。

核心结论

  • 在体模中,空间追踪心腔内超声结合电磁追踪可实时重建血管三维路径并在无辐射成像条件下导航导丝,几何精度以CT为参考Dice系数0.79、表面平均距离1.7 ± 1.12 mm。

研究设计与样本

  • 先导研究,使用空间追踪心腔内超声在三维虚拟环境中从体模重建血管,并以体模CT扫描评估几何精度。

关键影像表现/征象

  • 在三维虚拟环境中重建体模血管,并展示追踪导丝在体模血管内导航的超声导航虚拟显示。
  • 以体模CT扫描为参考,重建模型与CT模型的Dice系数为0.79。
  • 两模型表面之间的平均距离为1.7 ± 1.12 mm。

诊断标准与关键数值

  • 几何精度评估:以体模CT扫描为参考,Dice系数0.79,两模型表面平均距离1.7 ± 1.12 mm。

临床应用与阅片要点

  • 作者提出使用导管超声结合电磁追踪生成实时血管路径图并导航,以避免有害辐射。
  • 本先导研究在体模中演示了无辐射成像条件下追踪导丝在血管内导航的虚拟显示。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Transcatheter cardiovascular interventions have the advantage of patient safety, reduced surgery time and minimal trauma to the patient's body. Transcathether interventions, which are performed percutaneously, are limited by the lack of direct line of sight with the procedural tools and the patient anatomy. Therefore, such interventional procedures rely heavily on image guidance for navigating toward and delivering therapy at the target site. Vascular navigation via the inferior vena cava, from the groin to the heart, is an imperative part of most transcatheter cardiovascular interventions including heart valve repair surgeries and ablation therapy. Traditionally, the inferior vena cava is navigated using fluoroscopic techniques such as venography and computed tomography venography. These X-ray-based techniques can have detrimental effects on the patient as well as the surgical team, causing increased radiation exposure, leading to risk of cancer, fetal defects and eye cataracts. The use of a heavy lead apron has also been reported to cause back pain and spine issues, thus leading to interventionalist's disc disease. We propose the use of a catheter-based ultrasound augmented with electromagnetic tracking technology to generate a vascular roadmap in real time and perform navigation without harmful radiation. In this pilot study, we used spatially tracked intracardiac echocardiography to reconstruct a vessel from a phantom in a 3-D virtual environment. We illustrate how the proposed ultrasound-based navigation will appear in a virtual environment, by navigating a tracked guidewire within the vessels in the phantom without any radiation-based imaging. The geometric accuracy is assessed using a computed tomography scan of the phantom, with a Dice coefficient of 0.79. The average distance between the surfaces of the two models comes out to be 1.7 ± 1.12 mm.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。