不准确的床高设置会影响计算机断层扫描中器官特异性辐射剂量

Inaccurate table height setting affects the organ-specific radiation dose in computed tomography.
PubMed 文献 · Eur J Radiol · 2022-04-15

研究用拟人体模加乳腺体模,在龙门中心及每20 mm改变床高(-40至+40 mm)条件下,比较CTDIvol与玻璃剂量计实测的吸收剂量;结果CTDIvol在自动床高校正(AHC)下基本维持,但晶状体和乳腺的吸收剂量随床高显著变化,-40 mm时两者均高于中心位置。

要点

  • CTDIvol在AHC下不能反映床高偏移导致的器官吸收剂量变化。
  • 床高-40 mm时晶状体和乳腺吸收剂量均高于龙门中心位置。
  • 即使使用AHC,操作者也应恰当摆放患者体位。

核心结论

  • 吸收剂量受床高影响,而CTDIvol可由自动床高校正(AHC)维持在相近水平;操作者在使用AHC时仍应恰当摆放患者。

研究设计与样本

  • 体模研究,使用拟人体模联合乳腺体模,以龙门中心为基准,床高在-40 mm至40 mm之间每20 mm变化,各床高均采用自动床高校正(AHC)采集数据;CTDIvol取自CT控制台及各层管电流值(DICOM标签0018,1151),吸收剂量由植入体模各位置的玻璃剂量计测量。

关键影像表现/征象

  • 肺部管电流值在床高+40 mm时低于其他床高。
  • CTDIvol在+40 mm时略低于中心位置(12.78 mGy vs. 13.42 mGy,p < 0.05)。
  • 其他床高的CTDIvol值几乎相同(13.30-13.40 mGy)。
  • 晶状体和乳腺的吸收剂量与龙门中心位置相比有显著差异(分别为-27.27%-17.77%和-24.31%-12.83%)。
  • 与中心位置相比,晶状体和乳腺在床高-40 mm时吸收剂量均更高。

诊断标准与关键数值

  • CTDIvol在床高+40 mm为12.78 mGy,中心位置为13.42 mGy(p < 0.05)。
  • 其他床高CTDIvol值为13.30-13.40 mGy。
  • 晶状体吸收剂量相对中心位置变化范围为-27.27%-17.77%。
  • 乳腺吸收剂量相对中心位置变化范围为-24.31%-12.83%。

临床应用与阅片要点

  • 阅片或做剂量评估时不应仅凭CTDIvol判断器官吸收剂量,床高偏移可改变晶状体和乳腺等器官的实际吸收剂量。
  • 使用自动床高校正(AHC)时仍需关注患者体位摆放,因为AHC维持CTDIvol但未消除床高对器官吸收剂量的影响。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: We aimed to prove that the locally absorbed doses in tissues and organs are affected by inaccurate table height in computed tomography.

MATERIALS AND METHODS: We compared the volume CT dose index (CTDIvol) and the absorbed doses using an anthropomorphic phantom combined with a breast phantom. The phantom was set at the gantry center, from which the table height was changed every 20 mm between-40 mm and 40 mm. Data acquisition was performed using auto table height correction (AHC) for each table height. The CTDIvol was obtained from the CT console and the tube current value for each image slice (DICOM tag: 0018, 1151). The absorbed dose was measured by a glass dosimeter that was implanted at various positions in the phantom.

RESULTS: The tube current values in the lung were lower at a table height of + 40 mm than those at other heights. The CTDIvol was slightly lower at + 40 mm than at the center (12.78 mGy vs. 13.42 mGy, p < 0.05). The CTDIvol values were almost the same at the other table heights (13.30-13.40 mGy). The absorbed doses at the lens and mammary gland were significantly different from those at the gantry center (-27.27%-17.77% and -24.31%-12.83%, respectively). Compared with the center, both the lens and mammary gland had higher absorbed doses at a table height of -40 mm.

CONCLUSION: The absorbed dose was affected by the table height, but the CTDIvol was maintained by AHC. The operator should appropriately position patients even when using AHC.