活体兔肝及股骨骨表面暴露于声辐射力脉冲的温度升高测量
本研究在活体兔肝(有无超声造影剂)及股骨骨表面施加声辐射力脉冲(ARFI),用热电偶测量最高温升、5分钟温升与冷却时间,发现骨表面最高温升超过英国医学超声学会(BMUS)声明的允许温升1.5°C,但利用单脉冲脉冲强度积分(PII)与温度参数的线性关系可确定PII使最高温升在允许范围内,并可由PII估算冷却时间。
要点
- 骨表面最高温升可超过BMUS声明的允许温升1.5°C。
- 单脉冲PII与最高温升、5分钟温升及冷却时间呈线性关系。
- 可通过PII确定输出使最高温升处于BMUS允许范围内。
核心结论
- 在诊断超声输出规定限值内测试时,股骨骨表面最高温升超过BMUS声明的允许温升1.5°C;利用PII与温度参数的线性关系可确定PII使最高温升在允许范围内,并可由PII估算冷却时间。
研究设计与样本
- 活体兔实验,对肝脏(有无超声造影剂)及股骨骨表面施加ARFI,用热电偶测量温度参数。
诊断标准与关键数值
- 最高温升超过BMUS声明的允许温升1.5°C。
- 单脉冲PII与最高温升、5分钟温升及冷却时间呈线性关系。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Acoustic radiation force impulse (ARFI) imaging and shear wave elastography use a "push pulse." The push pulse, which is referenced as an ARFI in this study, has a longer duration than that of conventional diagnostic pulses (several microseconds). Therefore, there are concerns regarding thermal safety in vivo. However, few in vivo studies have been conducted using living animals. In this study, to suggest a concept for deciding an ARFI output and cooling time while considering thermal safety, the liver (with and without an ultrasound contrast agent) and femur bone surface of living rabbits were exposed to an ARFI, and the maximum temperature rise, temperature rise for 5-min duration, and cooling time were measured via a thermocouple. While testing within the regulation limits of diagnostic ultrasound outputs, a maximum temperature rise on the femur bone surface exceeded the allowable temperature rise (1.5°C) in the British Medical Ultrasound Society (BMUS) statement. However, using the linear relationships between the pulse intensity integral (PII) of a single pulse and the above three temperature parameters, PII may be determined so that the maximum temperature rise is within the allowable temperature rise in the BMUS statement. The cooling time can be estimated from the PII.
引用
- Nitta, Ishiguro, Sasanuma, et al. In Vivo Temperature Rise Measurements of Rabbit Liver and Femur Bone Surface Exposed to an Acoustic Radiation Force Impulse. Ultrasound Med Biol. 2022. PMID: 35422349. doi:10.1016/j.ultrasmedbio.2022.02.019
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。