不同预压不降低声触诊组织成像与定量技术对乳腺病灶的诊断价值,尤其病灶与周围组织剪切波速度比值更为稳定

Different precompression does not reduce the diagnostic value of virtual touch tissue imaging and quantification (VTIQ) in breast lesions, especially for the ratio of the shear wave velocity between lesions and surrounding tissues.
PubMed 文献 · Eur J Radiol · 2022-04-04

本单中心前瞻性研究纳入167例乳腺病灶患者,在四种不同预压水平下行声触诊组织成像与定量技术(VTIQ)检查,病灶均经组织病理学证实。结果显示恶性病灶VTIQ参数在各预压水平均显著高于良性病灶(P<0.001),且随预压增加病灶、脂肪及腺体剪切波速度均显著升高,但各参数诊断效能仍较好(AUC=0.765-0.911)。病灶与脂肪及病灶与腺体剪切波速度比值在不同预压水平间无显著差异,变异系数截止值分别为7.42%和8.55%。

要点

  • 恶性乳腺病灶的VTIQ参数在四种预压水平下均显著高于良性病灶(P<0.001)。
  • 预压增加会显著升高病灶、脂肪及腺体的剪切波速度,但不降低VTIQ参数的诊断价值(AUC=0.765-0.911)。
  • 病灶与周围组织剪切波速度比值在不同预压水平下更稳定,变异系数截止值分别为7.42%(病灶/脂肪)和8.55%(病灶/腺体)。

核心结论

  • 预压可增加乳腺病灶、脂肪及腺体组织的硬度,但不降低VTIQ参数对乳腺病灶的诊断价值;在不同预压水平下,病灶与周围组织剪切波速度比值诊断乳腺病灶更稳定。

研究设计与样本

  • 单中心前瞻性研究,纳入167例女性乳腺病灶患者,病灶均经组织病理学证实;在四种不同预压水平下行VTIQ检查,于尽可能相同深度对病灶、周围脂肪及腺体组织重复测量7次剪切波速度。

诊断标准与关键数值

  • 恶性病灶VTIQ参数在各预压水平均显著高于良性病灶(P<0.001)。
  • 各预压水平下VTIQ参数诊断乳腺病灶的AUC为0.765-0.911。
  • 病灶与脂肪剪切波速度比值的变异系数截止值为7.42%。
  • 病灶与腺体剪切波速度比值的变异系数截止值为8.55%。

临床应用与阅片要点

  • 预压会增加病灶、脂肪及腺体组织的剪切波速度,阅片时需结合预压水平解读VTIQ参数。
  • 病灶与周围组织剪切波速度比值在不同预压水平下更稳定,可作为诊断乳腺病灶的参考指标。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate the accuracy of virtual touch tissue imaging and quantification (VTIQ) in the diagnosis of benign and malignant breast lesions under four different precompression levels. The shear wave velocity (SWV) ratios of lesion to surrounding tissue were also added for diagnosis.

METHODS: 167 female patients with breast lesions were included in this single center prospective study. VTIQ was performed under four different precompression levels. The SWV of the lesion, surrounding fat, and gland tissue were measured at the same depth as much as feasible 7 times. The breast lesions studied were all histopathologically confirmed. The VTIQ parameters were compared between precompression levels. The area under the receiver operating characteristic curve (AUC) was calculated to evaluate the diagnostic performance of each parameter of the VTIQ.

RESULTS: The VTIQ parameters of the malignant lesions were significantly higher than those of benign lesions in all precompression levels (P < 0.001). SWV of the lesion, fat, and gland tissue increased significantly with increasing precompression. The VTIQ parameters had great diagnostic performance for breast lesions in all precompression levels (AUC = 0.765-0.911). There was no significant difference between the precompression levels of the lesion-to-fat SWV ratio and the lesion-to-gland SWV ratio in benign and malignant lesions, and the cut-off coefficients of variation were 7.42% and 8.55%, respectively.

CONCLUSIONS: Precompression can increase the stiffness of breast lesions, fat and gland tissues, but does not reduce diagnostic value of VTIQ parameters in the breast. Under different precompression levels, the diagnosis of breast lesions by the ratio of the SWV of the lesion to the surrounding tissues is more stable.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Wang, Zhong, Wang, et al. Different precompression does not reduce the diagnostic value of virtual touch tissue imaging and quantification (VTIQ) in breast lesions, especially for the ratio of the shear wave velocity between lesions and surrounding tissues. Eur J Radiol. 2022. PMID: 35390603. doi:10.1016/j.ejrad.2022.110284

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。