儿童及年轻成人肝脏铁定量:三维多回波Dixon序列与常规二维T2*弛豫法的比较
回顾性纳入2015年7月至2020年1月因肝脏铁定量行MRI的38例患者共54次检查,以1.5T常规T2*弛豫法测得的肝脏铁浓度为参考标准,评估Siemens Healthineers LiverLab三维多回波Dixon技术的诊断准确性。结果显示该技术估测T2*的观察者间一致性极好(CCC 0.99),当肝脏铁浓度≤8 mg/g时与参考标准相关性很强(读者1 r=0.93、读者2 r=0.92,均P<0.001),但铁过载加重时一致性变差、相关性很弱。
要点
- 三维多回波Dixon估测肝脏T2*的观察者间一致性极好(CCC 0.99,CI 0.99-1.00)。
- 肝脏铁浓度≤8 mg/g时,Dixon法与T2*弛豫法测得的肝脏铁浓度相关性很强(读者1 r=0.93、读者2 r=0.92,均P<0.001)。
- 肝脏铁浓度升高时Bland-Altman图显示一致性变差、相关性很弱(读者1 r=0.05,P=0.84;读者2 r=0.17,P=0.44)。
核心结论
- 基于厂商的三维多回波Dixon技术在肝脏T2*估测中观察者间相关性极好;当肝脏铁过载为轻至中度(肝脏铁浓度≤8 mg/g)时,该方法估测的肝脏铁浓度与常规T2*弛豫法相关性很强。
研究设计与样本
- 回顾性研究,查询放射信息系统纳入2015年7月至2020年1月因肝脏铁定量行MRI的38例患者共54次检查(22例女性),平均年龄11.8岁(标准差5.3岁),所有患者均在1.5T扫描仪上行T2*弛豫法估测肝脏铁浓度,并加做Siemens Healthineers LiverLab(德国埃尔朗根)三维多回波Dixon序列,由两名读者独立在扫描仪生成的R2*和T2*图上手绘自由勾画感兴趣区估测肝脏R2*和T2*,以常规T2*弛豫法测得的肝脏铁浓度为参考标准。
诊断标准与关键数值
- 参考标准肝脏铁浓度范围1.1-21.1(中位数6.8)mg/g肝脏干重;三维多回波Dixon估测T2*的观察者间一致性CCC 0.99(置信区间0.99-1.00);肝脏铁浓度平均值≤8 mg/g时Bland-Altman图所有观测值聚集在零偏倚线附近,读者1 r=0.93、读者2 r=0.92(均P<0.001);肝脏铁浓度升高时观测值越过一致性界限下限、相关性很弱(读者1 r=0.05,P=0.84;读者2 r=0.17,P=0.44)。
临床应用与阅片要点
- 在儿童及年轻成人肝脏铁定量中,三维多回波Dixon可在扫描仪上内联/自动后处理,无需第三方软件,其估测T2*的观察者间一致性极好;当肝脏铁浓度为轻至中度(≤8 mg/g)时,该方法测得的肝脏铁浓度与常规T2*弛豫法相关性很强;肝脏铁浓度升高时两种方法一致性变差、相关性很弱。
局限性
- 原文为回顾性研究。
- 肝脏铁浓度升高时Bland-Altman图显示一致性差、相关性很弱(读者1 r=0.05,P=0.84;读者2 r=0.17,P=0.44)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Magnetic resonance imaging (MRI)-based liver iron quantification is the standard of care to guide chelation therapy in children at risk of hemochromatosis. T2* relaxometry is the most widely used technique but requires third-party software for post-processing. Vendor-provided three-dimensional (3-D) multi-echo Dixon techniques are now available that allow inline/automated post-processing.
OBJECTIVE: The purpose of our study was to evaluate the diagnostic accuracy of a volumetric multi-echo Dixon technique using conventional T2* relaxometry as the reference standard in a pediatric and young adult population.
MATERIALS AND METHODS: In this retrospective study, we queried the radiology information system to identify all MRIs performed for liver iron quantification from July 2015 to January 2020. All patients had undergone T2* relaxometry on a 1.5-tesla (T) scanner for liver iron concentration (LIC) estimation. In addition, a 3-D multi-echo Dixon was performed using Siemens Healthineers LiverLab (Erlangen, Germany). Two readers independently estimated liver R2* and T2* on the multi-echo Dixon by drawing free-hand regions of interest on the scanner-generated R2* and T2* maps. Conventional T2*-relaxometry-based LIC was the reference standard. We estimated interobserver agreement by concordance correlation coefficient (CCC). We used Bland-Altman analysis and Pearson correlation coefficient (r) to compare LIC by the two methods.
RESULTS: Fifty-four MRIs on 38 patients (22 females) were available for analysis. Mean patient age was 11.8 years (standard deviation [SD] 5.3 years). Reference standard LIC ranged 1.1-21.1 (median 6.8) mg/g dry weight of liver. The concordance between readers for T2* estimation using 3-D multi-echo Dixon was substantial (CCC 0.99, confidence interval 0.99-1.00). Bland-Altman plot showed that all observations were clustered around the zero bias line if the LIC average was ≤8 mg/g, and r was very strong (reader 1 r=0.93, reader 2 r=0.92, both P-values <0.001). With increasing LIC, there was a pattern of poor agreement on the Bland-Altman plot, with observations crossing the lower limits of agreement, and r was very weak (reader 1 r=0.05, P-value 0.84; reader 2 r=0.17, P-value 0.44).
CONCLUSION: Vendor-based 3-D multi-echo Dixon allows for excellent interobserver correlation in liver T2* estimation. LIC estimated by this method has a very strong correlation with conventional T2* relaxometry if liver iron overload is mild-moderate (LIC ≤8 mg/g).
引用
- Bhimaniya, Arora, Sharma, et al. Liver iron quantification in children and young adults: comparison of a volumetric multi-echo 3-D Dixon sequence with conventional 2-D T2* relaxometry. Pediatr Radiol. 2022. PMID: 35384483. doi:10.1007/s00247-022-05352-4
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。