三维屏气磁共振胰胆管成像采用空间选择性射频激发脉冲的可行性:与并行成像技术及压缩感知方法的前瞻性比较
前瞻性纳入106例疑似胰胆疾病患者,在3T下比较改良三维屏气并行成像MRCP(空间选择性射频激发脉冲)、原三维屏气并行成像MRCP(非选择性激发脉冲)与三维屏气压缩感知MRCP。三种序列的对比度、信噪比和对比噪声比有显著差异(p<0.001);改良序列单次屏气22秒内图像质量与原序列及压缩感知序列相当,信噪比提高且胰胆树显示更优。
要点
- 改良三维屏气并行成像MRCP采用空间选择性射频激发脉冲,单次屏气22秒完成扫描。
- 三种序列的胆总管对比度、信噪比和对比噪声比差异有统计学意义(p<0.001)。
- 改良序列图像质量与原并行成像序列及压缩感知序列相当,信噪比提高且胰胆树显示更优。
核心结论
- 采用空间选择性射频激发脉冲的改良三维屏气并行成像MRCP可在单次屏气22秒内提供与原并行成像MRCP及三维屏气压缩感知MRCP相当的图像质量,并显示信噪比提高和胰胆树可视化更优。
研究设计与样本
- 前瞻性研究,2021年1月至2021年7月纳入106例疑似胰胆疾病患者,在3T下分别行改良三维屏气并行成像MRCP、原三维屏气并行成像MRCP和三维屏气压缩感知MRCP,采用Friedman检验比较三种方案的对比度、信噪比、对比噪声比、整体图像质量和胆管可视化。
诊断标准与关键数值
- 胆总管对比度、信噪比和对比噪声比在三种序列间差异有统计学意义(p<0.001)。
- 两种三维屏气并行成像MRCP的整体图像质量高于三维屏气压缩感知MRCP,但差异无统计学意义。
- 原三维屏气并行成像MRCP的前支和后支可视化评分显著高于三维屏气压缩感知MRCP,改良三维屏气并行成像MRCP与三维屏气压缩感知MRCP之间无显著差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: The purpose of the present study was to evaluate the clinical feasibility of the modified 3D breath-hold magnetic resonance cholangiopancreatography with parallel imaging (3D-BH-PI-MRCP) using a spatially selective radiofrequency excitation pulse in patients with suspected pancreaticobiliary diseases. Moreover, we also compared its image quality with those of the original 3D-BH-PI-MRCP with a nonselective exciting pulse and the 3D breath hold compressed sensing magnetic resonance cholangiopancreatography (3D-BH-CS-MRCP).
MATERIALS AND METHODS: Between January 2021 and July 2021, 106 patients prospectively underwent modified 3D-BH-PI-MRCP, original 3D-BH-PI-MRCP and 3D-BH-CS-MRCP at 3T in this study. The Friedman test was performed to compare the contrast, signal-to-noise-ratio (SNR), and contrast-noise-ratio, overall image quality, and duct visualization among the three protocols.
RESULTS: The contrast ratio, SNR and contrast-to-noise ratio of the common bile duct differed significantly among the three sequences (p < 0.001). Compared to the 3D-BH-CS-MRCP protocol, the overall imaging quality of the two 3D-BH-PI-MRCP was higher but not significantly different. The scores for the anterior and posterior branches visualization were significantly higher in the original 3D-BH-PI-MRCP compared to the 3D-BH-CS-MRCP, but were no significant differences between the modified 3D-BH-PI-MRCP and the 3D-BH-CS-MRCP.
CONCLUSION: The modified 3D-BH-PI-MRCP with a spatially selective radiofrequency excitation pulse could provide comparable image quality to the original 3D-BH-PI-MRCP and the 3D-BH-CS-MRCP during a single breath hold (22 seconds), and showed improved SNR and superior visualization of the pancreaticobiliary tree.