文献速递腹部技术与安全MR

自适应图像接收线圈技术在3.0T肝脏磁共振成像中的影响:与传统线圈的个体内比较

Impact of adaptive image receive coil technology for liver MR imaging at 3.0 Tesla: Intraindividual comparison with use of conventional coil.
PubMed 文献 · Eur J Radiol · 2022-03-19

回顾性纳入50例连续患者,在3.0T磁共振上使用自适应图像接收(AIR)线圈与传统线圈行肝脏磁共振检查,比较两种线圈的三维脂肪抑制T1加权梯度回波(增强前后)、T2加权单次激发快速自旋回波和扩散加权图像。AIR线圈的信噪比显著高于传统线圈(P<.05),主观图像噪声更低,整体图像质量更优。

要点

  • AIR线圈在增强前后T1加权和扩散加权图像上的主观噪声低于传统线圈(P<.05)。
  • AIR线圈的信噪比显著高于传统线圈(P<.05)。
  • 至少一名读者评价AIR线圈在增强前后T1加权和扩散加权图像上的整体图像质量优于传统线圈(P<.05)。

核心结论

  • 与传统线圈相比,AIR线圈提高了肝脏磁共振成像的信噪比和图像质量。

研究设计与样本

  • 回顾性研究,纳入50例连续患者,在3.0T磁共振系统上使用AIR线圈和传统线圈进行随访肝脏磁共振检查以评估肝细胞癌。

诊断标准与关键数值

  • 一名放射科医生测量信噪比,另外两名放射科医生采用3分法独立评估主观图像噪声、伪影、信号均匀性和整体图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: A newly developed Adaptive Image Receive (AIR) coil is designed to be more flexible to conform to the human body habitus, and may improve image quality by reducing the distance between the coil element and the imaging subject. This study evaluated the AIR coil's usefulness for liver MR imaging at 3.0 T in comparison with that of a conventional coil retrospectively.

METHOD: The study population comprised 50 consecutive patients, who underwent follow-up liver MR examinations with a 3.0-T MR system using both an AIR coil and a conventional coil to evaluate hepatocellular carcinoma. Three-dimensional fat-suppressed T1-weighted gradient-echo images before and after injection of an MRI contrast agent, T2-weighted single-shot fast spin-echo (SSFSE) images, and diffusion-weighted (DW) images obtained with the AIR coil were compared with corresponding images obtained with the conventional coil. One radiologist measured signal-to-noise ratios (SNRs), while two other radiologists used a 3-point scale to independently assess subjective image noise, artifacts, signal uniformity, and overall image quality.

RESULTS: SNRs for the AIR coil were significantly higher than those for the conventional coil (P <.05). Subjective image noise for the AIR coil on pre- and postcontrast T1-weighted and DW images was lower than for the conventional coil for both readers (P <.05). Overall image quality on pre- and postcontrast T1-weighted and DW images for the AIR coil was better than that for the conventional coil for at least one reader (P <.05).

CONCLUSIONS: In comparison with the conventional coil, AIR coil improved SNR and image quality of liver MR imaging.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Fukui, Onishi, Nakamoto, et al. Impact of adaptive image receive coil technology for liver MR imaging at 3.0 Tesla: Intraindividual comparison with use of conventional coil. Eur J Radiol. 2022. PMID: 35334243. doi:10.1016/j.ejrad.2022.110271

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。