北美儿科人群中手部X线片自动骨龄分析的验证
本研究比较了451例儿科患者共586张骨龄X线片的自动软件BoneXpert与3名放射科医师手动评估的骨龄,采用Greulich和Pyle法。自动与手动评估呈强相关(r=0.98;R2=0.96;P<0.0001),平均骨龄差为0.12±0.76岁;按性别、内分泌疾病和种族分层后相关性仍强(P<0.0001);纵向分析中相关性亦强(r=0.7852;R2=0.63;P<0.01)。结论认为自动骨龄评估在北美儿科临床实践大样本中可靠且准确。
要点
- 自动骨龄软件与放射科医师手动评估强相关(r=0.98;R2=0.96;P<0.0001)。
- 平均骨龄差为0.12±0.76岁,提示自动评估与手动评估高度一致。
- 按性别、内分泌疾病和种族分层后相关性仍强(P<0.0001),纵向分析亦强(r=0.7852;R2=0.63;P<0.01)。
核心结论
- 在北美儿科临床实践大样本中,自动骨龄评估(BoneXpert)被证实可靠且准确。
研究设计与样本
- 研究纳入451例患者的586张骨龄X线片,由3名放射科医师于2013至2018年手动评估,并与BoneXpert自动分析比较,采用Greulich和Pyle法。
诊断标准与关键数值
- 自动与手动骨龄评估强相关:r=0.98;R2=0.96;P<0.0001。
- 平均骨龄差为0.12±0.76岁。
- 按性别、常见内分泌疾病(包括生长障碍和早/性早熟)及种族分层后,两种方法仍强相关(P<0.0001)。
- 纵向分析中自动软件与手动骨龄随时间强相关:r=0.7852;R2=0.63;P<0.01。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Radiographic bone age assessment by automated software is precise and instantaneous.
OBJECTIVE: The aim of this study was to evaluate the accuracy of an automated tool for bone age assessment.
MATERIALS AND METHODS: We compared a total of 586 bone age radiographs from 451 patients, which had been assessed by three radiologists from 2013 to 2018, with bone age analysis by BoneXpert, using the Greulich and Pyle method. We made bone age comparisons in different patient groups based on gender, diagnosis and race, and in a subset with repeated bone age studies. We calculated Spearman correlation (r) and accuracy (root mean square error, or R2).
RESULTS: Bone age analyses by automated and manual assessments showed a strong correlation (r=0.98; R2=0.96; P<0.0001), with the mean bone age difference of 0.12±0.76 years. Bone age comparisons by the two methods remained strongly correlated (P<0.0001) when stratified by gender, common endocrine conditions including growth disorders and early/precocious puberty, and race. In the longitudinal analysis, we also found a strong correlation between the automated software and manual bone age over time (r=0.7852; R2=0.63; P<0.01).
CONCLUSION: Automated bone age assessment was found to be reliable and accurate in a large cohort of pediatric patients in a clinical practice setting in North America.
引用
- Bowden, Bowden, Ruess, et al. Validation of automated bone age analysis from hand radiographs in a North American pediatric population. Pediatr Radiol. 2022. PMID: 35325266. doi:10.1007/s00247-022-05310-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。