文献速递骨肌技术与安全CT

光子计数CT对骨小梁结构成像的有效空间分辨率优于常规临床CT,与高分辨率外周CT相近

Effective Spatial Resolution of Photon Counting CT for Imaging of Trabecular Structures is Superior to Conventional Clinical CT and Similar to High Resolution Peripheral CT.
PubMed 文献 · Invest Radiol · 2022-03-21

研究将5个人椎体标本在光子计数CT、双能CT和单能CT三种临床CT上各扫描3次,以高分辨率外周CT为参考标准,通过三维点扩散和调制传递函数评估有效空间分辨率。光子计数CT的有效空间分辨率与高分辨率外周CT相当,体积有效分辨率较双能CT提高5.5倍(MTF 10%)和18倍(MTF 5%),较单能CT提高8.7倍(MTF 10%)和20倍(MTF 5%)。

要点

  • 光子计数CT在骨小梁成像中的有效空间分辨率与高分辨率外周CT相当,较常规临床CT提高超过5倍。
  • Br76重建核在高对比(MTF 10%)下径向有效分辨率为16.9 lp/cm(最小可分辨物体295 μm)。
  • Br76在低对比(MTF 5%)下径向有效分辨率为30.5 lp/cm(164 μm),可能受164 μm各向同性体素限制。

核心结论

  • 光子计数CT在骨小梁骨成像中的有效空间分辨率与高分辨率外周CT相当,较常规CT提高超过5倍;对于离体样本且患者辐射剂量可忽略时,光子计数CT可实现临床前细节水平的骨微结构成像。

研究设计与样本

  • 实验研究,5个人椎体标本包埋于环氧树脂中,每个标本扫描3次,使用3台临床CT扫描仪(光子计数CT Naeotom Alpha、双能CT Somatom Force、单能CT Somatom Sensation 40,均为Siemens Healthineers产品),以XtremeCT高分辨率外周CT获取参考标准。

诊断标准与关键数值

  • 高对比(MTF 10%)下光子计数CT径向有效分辨率:Br68核为10.5 lp/cm(最小可分辨物体476 μm),Br76核为16.9 lp/cm(295 μm)。
  • 低对比(MTF 5%)下光子计数CT径向有效分辨率:Br68核为10.8 lp/cm(464 μm),Br76核为30.5 lp/cm(164 μm)。
  • 光子计数CT两个核的轴向有效分辨率在27.0 lp/cm(185 μm)至29.9 lp/cm(167 μm)之间。
  • Br76核的体积有效分辨率在61.9(MTF 10%)至222.4(MTF 5%)个元素/立方毫米之间。

局限性

  • Br76核的空间分辨率可能更高,但技术上受164 μm各向同性体素尺寸限制。
  • 结论基于离体人椎体标本,且仅在患者辐射剂量可忽略的条件下评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Photon counting computed tomography (PCCT) might offer an effective spatial resolution that is significantly improved compared with conventional state-of-the-art computed tomography (CT) and even provide a microstructural level of detail similar to high-resolution peripheral CT (HR-pQCT). The aim of this study was to evaluate the volumetric effective spatial resolution of clinically approved PCCT as an alternative to HR-pQCT for ex vivo or preclinical high-resolution imaging of bone microstructure.

MATERIALS AND METHODS: The experiment contained 5 human vertebrae embedded in epoxy resin, which were scanned 3 times each, and on 3 different clinical CT scanners: a PCCT (Naeotom Alpha), a dual-energy CT (Somatom Force [SF]), and a single-energy CT (Somatom Sensation 40 [S40]), all manufactured by Siemens Healthineers (Erlangen, Germany). Scans were performed with a tube voltage of 120 kVp and, to provide maximum scan performance and minimum noise deterioration, with exposures of 1500 mAs (SF), 2400 mAs (S40), and 4500 mAs (PCCT) and low slice increments of 0.1 (PCCT) and 0.3 mm (SF, S40). Images were reconstructed with sharp and very sharp bone kernels, Br68 and Br76 (PCCT), Br64 (SF), and B65s and B75h (S40). Ground truth information was obtained from an XtremeCT scanner (Scanco, Brüttisellen, Switzerland). Voxel-wise comparison was performed after registration, calibration, and resampling of the volumes to isotropic voxel size of 0.164 mm. Three-dimensional point spread- and modulation-transfer functions were calculated with Wiener's deconvolution in the anatomical trabecular structure, allowing optimum estimation of device- and kernel-specific smoothing properties as well as specimen-related diffraction effects on the measurement.

RESULTS: At high contrast (modulation transfer function [MTF] of 10%), radial effective resolutions of PCCT were 10.5 lp/cm (minimum resolvable object size 476 μm) for kernel Br68 and 16.9 lp/cm (295 μm) for kernel Br76. At low contrast (MTF 5%), radial effective spatial resolutions were 10.8 lp/cm (464 μm) for kernel Br68 and 30.5 lp/cm (164 μm) for kernel Br76. Axial effective resolutions of PCCT for both kernels were between 27.0 (185 μm) and 29.9 lp/cm (167 μm). Spatial resolutions with kernel Br76 might possibly be still higher but were technically limited by the isotropic voxel size of 164 μm. The effective volumetric resolutions of PCCT with kernel Br76 ranged between 61.9 (MTF 10%) and 222.4 (MTF 5%) elements per cubic mm. Photon counting CT improved the effective volumetric resolution by factor 5.5 (MTF 10%) and 18 (MTF 5%) compared with SF and by a factor of 8.7 (MTF 10%) and 20 (MTF 5%) compared with S40. Photon counting CT allowed obtaining similar structural information as HR-pQCT.

CONCLUSIONS: The effective spatial resolution of PCCT in trabecular bone imaging was comparable with that of HR-pQCT and more than 5 times higher compared with conventional CT. For ex vivo samples and when patient radiation dose can be neglected, PCCT allows imaging bone microstructure at a preclinical level of detail.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Thomsen, Horstmeier, Niehoff, et al. Effective Spatial Resolution of Photon Counting CT for Imaging of Trabecular Structures is Superior to Conventional Clinical CT and Similar to High Resolution Peripheral CT. Invest Radiol. 2022. PMID: 35318968. doi:10.1097/RLI.0000000000000873

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。