低剂量CT灌注联合稀疏视图滤波反投影在急性缺血性卒中中的应用
本研究为模拟研究,纳入38例连续急性缺血性卒中患者,通过正向投影原始984视图重建并添加高斯噪声模拟稀疏视图CTP,用滤波反投影重建并生成CBF和Tmax图,评估缺血体积准确性。结果显示164视图CBF<30%及246和164视图Tmax>6s体积与全视图有显著差异,结论提示稀疏视图CTP可能允许剂量降低达3倍。
要点
- 稀疏视图CTP在视图数降至328时,CBF<30%和Tmax>6s缺血体积与全视图无显著差异。
- 视图数降至246时Tmax>6s体积已与全视图显著不同,降至164时CBF<30%体积也显著不同。
- 基于缺血体积准确性,稀疏视图CTP可能允许剂量降低达3倍。
核心结论
- 基于缺血体积准确性,本研究表明稀疏视图CTP可能允许剂量降低达3倍。
研究设计与样本
- 模拟研究,纳入38例连续急性缺血性卒中并行CTP的患者,模拟全视图(984视图)和稀疏视图(492、328、246、164视图)CTP,采用滤波反投影重建,通过去卷积生成CBF和Tmax图。
诊断标准与关键数值
- 缺血体积测量标准:CBF<30%相对对侧半球,Tmax>6s;体积准确性以全视图CTP为参考,采用Friedman检验及事后多重比较校正的成对检验和Bland-Altman分析。
局限性
- 本研究为模拟研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: Radiation dose associated with computed tomography (CT) perfusion (CTP) may discourage its use despite its added diagnostic benefit in quantifying ischemic lesion volume. Sparse-view CT reduces scan dose by acquiring fewer X-ray projections per gantry rotation but is contaminated by streaking artifacts using filtered back projection (FBP). We investigated the achievable dose reduction by sparse-view CTP with FBP without affecting CTP lesion volume estimations.
MATERIALS AND METHODS: Thirty-eight consecutive patients with acute ischemic stroke and CTP were included in this simulation study. CTP projection data was simulated by forward projecting original reconstructions with 984 views and adding Gaussian noise. Full-view (984 views) and sparse-view (492, 328, 246, and 164 views) CTP studies were simulated by FBP of simulated projection data. Cerebral blood flow (CBF) and time-to-maximum of the impulse residue function (Tmax) maps were generated by deconvolution for each simulated CTP study. Ischemic volumes were measured by CBF<30% relative to the contralateral hemisphere and Tmax > 6 s. Volume accuracy was evaluated with respect to the full-view CTP study by the Friedman test with post hoc multiplicity-adjusted pairwise tests and Bland-Altman analysis.
RESULTS: Friedman and multiplicity-adjusted pairwise tests indicated that 164-view CBF < 30%, 246- and 164-view Tmax > 6 s volumes were significantly different to full-view volumes (p < 0.001). Mean difference ± standard deviation (sparse minus full-view lesion volume) ranged from -1.0 ± 2.8 ml to -4.1 ± 11.7 ml for CBF < 30% and -2.9 ± 3.8 ml to -12.5 ± 19.9 ml for Tmax > 6 s from 492 to 164 views, respectively.
CONCLUSION: By ischemic volume accuracy, our study indicates that sparse-view CTP may allow dose reduction by up to a factor of 3.