MagSeed无线定位乳腺病灶:300例的影像学视角
回顾性收集苏格兰西部三个NHS board两年数据,300例女性共置入309枚MagSeed,平均术前7.2天置入,99%成功置入,成功定位病灶100%切除,定位后乳腺X线摄影显示97.3%位置准确,标本X线显示45.1%病灶居中、35.7%偏心超过1mm、14.8%位于标本边缘,再切除率18.3%。
要点
- MagSeed置入后乳腺X线摄影可确认位置,97.3%定位准确,未见延迟迁移。
- 标本X线显示45.1%病灶居中,35.7%偏心超过1mm,14.8%位于标本边缘。
- 再切除率18.3%,乳腺X线摄影引导与超声引导定位准确性无差异。
核心结论
- MagSeed是乳腺保乳手术中准确可靠的定位方法,手术结局良好。
研究设计与样本
- 回顾性研究,收集苏格兰西部三个NHS board两年数据,300例女性共置入309枚MagSeed,在乳腺X线摄影或超声引导下于手术当日或术前最多30天置入,复习定位后乳腺X线摄影、术中标本X线及手术结局。
关键影像表现/征象
- 定位后乳腺X线摄影显示MagSeed影像学上准确放置于97.3%的病例。
- 标本X线显示病灶位于中央者占45.1%,偏心且距边缘超过1mm者占35.7%,位于标本边缘者占14.8%。
- 未观察到MagSeed延迟迁移。
诊断标准与关键数值
- MagSeed成功置入率为99%,成功定位的病灶手术切除率为100%。
- 乳腺X线摄影引导与超声引导定位的准确性无差异。
- 再切除率为18.3%。
临床应用与阅片要点
- 置入后乳腺X线摄影可用于确认MagSeed位置,本组97.3%影像学位置准确。
- 标本X线可评估病灶与标本边缘的关系,本组45.1%居中、35.7%偏心超过1mm、14.8%位于边缘。
- 本组再切除率为18.3%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The purpose of this article is to review the technical and radiological aspects of MagSeed® localisation, to assess its accuracy based on post-localisation mammograms and excision specimen X-rays and to discuss the radiological experience of our institutions.
METHODS: Two-year data were collected retrospectively from three NHS boards from the West of Scotland. A total of 309 MagSeeds® were inserted under mammographic or ultrasonographic guidance in 300 women with unifocal, multifocal and/or bilateral breast lesions at the day of surgery or up to 30 days prior to it. Radiological review of post-localisation mammograms and intraoperative specimen X-rays as well as a review of the surgical outcomes were performed to assess the accuracy and efficacy of the method. Our experience relating to the technique's strengths and downsides were also noted.
RESULTS: The MagSeeds® were inserted on average 7.2 days before surgery. The localisation technique was straight forward for the radiologists. In 99% of the cases, the MagSeed® was successfully deployed and 100% of the successfully localised lesions were excised at surgery. There was no difference in the accuracy of the localisation whether this was mammographically or ultrasonographically guided. On post-localisation mammograms, the MagSeed® was radiologically accurately positioned in 97.3% of the cases. No delayed MagSeed® migration was observed. On the specimen X-rays, the lesion was centrally positioned in 45.1%, eccentric within more than 1 mm from the margin in 35.7% and in 14.8% it was at the specimen's margin. The re-excision rate was 18.3%.
CONCLUSION: The MagSeed® is an accurate and reliable localisation method in breast conserving surgery with good surgical outcomes.
ADVANCES IN KNOWLEDGE: To our knowledge, the radiological aspects of MagSeed® localisation have not been widely described in peer-reviewed journals thus far.
引用
- Constantinidis, Sakellariou, Chang, et al. Wireless localisation of breast lesions with MagSeed. A radiological perspective of 300 cases. Br J Radiol. 2022. PMID: 35201906. doi:10.1259/bjr.20211241
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。