液性肝脏体模:模拟人肝黏弹性频散用于超声和磁共振弹性成像

Liquid-Liver Phantom: Mimicking the Viscoelastic Dispersion of Human Liver for Ultrasound- and MRI-Based Elastography.
PubMed 文献 · Invest Radiol · 2022-02-23

研究制备了线性聚合聚丙烯酰胺液性肝脏体模,用以模拟人肝剪切波速度随频率的频散特性。在5至3000 Hz范围内,体模的频散与活体人肝文献报道相符(弹簧壶刚度参数2.14 kPa,幂律指数0.367),批次间、6个月内及每摄氏度温升的刚度变异分别为0.5%、4.1%和-0.63%。作者认为该体模有助于超声和MRI弹性成像的标准化与开发。

要点

  • 液性肝脏体模可跨超声和MRI弹性成像设备使用,模拟人肝刚度频散。
  • 体模弹簧壶刚度参数2.14 kPa(95%CI 2.08-2.19),幂律指数0.367(95%CI 0.362-0.373)。
  • 批次间刚度变异0.5%,6个月内4.1%,每摄氏度温升-0.63%。

核心结论

  • 液性肝脏体模具有用于肝脏弹性成像标准化和开发的有用特性,可跨临床和实验性超声与MRI弹性成像设备使用。

研究设计与样本

  • 体模由线性聚合聚丙烯酰胺(PAAm)制成,按文献报道的活体健康人肝黏弹性校准;用2参数弹簧壶模型拟合剪切波速度频散,测量方法包括剪切流变仪、超声时谐弹性成像、临床磁共振弹性成像(MRE)和台式MRE,频率范围5至3000 Hz;并评估了超声和MRI成像参数、重现性、老化行为和温度依赖性,以及临床弹性成像方法(Aplio i900, Canon;Acuson Sequoia, Siemens;FibroScan, EchoSense)的剪切波速度频带宽度。

诊断标准与关键数值

  • 体模弹簧壶刚度参数2.14 kPa(95%CI 2.08-2.19),弹簧壶幂律指数0.367(95%CI 0.362-0.373)。
  • 批次间刚度变异0.5%(95%CI 0.4%-0.6%),6个月内变异4.1%(95%CI 3.9%-4.5%),每摄氏度温升变异-0.63%(95%CI -0.67%至-0.58%)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Tissue stiffness can guide medical diagnoses and is exploited as an imaging contrast in elastography. However, different elastography devices show different liver stiffness values in the same subject, hindering comparison of values and establishment of system-independent thresholds for disease detection. There is a need for standardized phantoms that specifically address the viscosity-related dispersion of stiffness over frequency. To improve standardization of clinical elastography across devices and platforms including ultrasound and magnetic resonance imaging (MRI), a comprehensively characterized phantom is introduced that mimics the dispersion of stiffness of the human liver and can be generated reproducibly.

MATERIALS AND METHODS: The phantom was made of linear polymerized polyacrylamide (PAAm) calibrated to the viscoelastic properties of healthy human liver in vivo as reported in the literature. Stiffness dispersion was analyzed using the 2-parameter springpot model fitted to the dispersion of shear wave speed of PAAm, which was measured by shear rheometry, ultrasound-based time-harmonic elastography, clinical magnetic resonance elastography (MRE), and tabletop MRE in the frequency range of 5 to 3000 Hz. Imaging parameters for ultrasound and MRI, reproducibility, aging behavior, and temperature dependency were assessed. In addition, the frequency bandwidth of shear wave speed of clinical elastography methods (Aplio i900, Canon; Acuson Sequoia, Siemens; FibroScan, EchoSense) was characterized.

RESULTS: Within the entire frequency range analyzed in this study, the PAAm phantom reproduced well the stiffness dispersion of human liver in vivo despite its fluid properties under static loading (springpot stiffness parameter, 2.14 [95% confidence interval, 2.08-2.19] kPa; springpot powerlaw exponent, 0.367 [95% confidence interval, 0.362-0.373]). Imaging parameters were close to those of liver in vivo with only slight variability in stiffness values of 0.5% (0.4%, 0.6%), 4.1% (3.9%, 4.5%), and -0.63% (-0.67%, -0.58%), respectively, between batches, over a 6-month period, and per °C increase in temperature.

CONCLUSIONS: The liquid-liver phantom has useful properties for standardization and development of liver elastography. First, it can be used across clinical and experimental elastography devices in ultrasound and MRI. Second, being a liquid, it can easily be adapted in size and shape to specific technical requirements, and by adding inclusions and scatterers. Finally, because the phantom is based on noncrosslinked linear PAAm constituents, it is easy to produce, indicating potential widespread use among researchers and vendors to standardize liver stiffness measurements.