正常脑回波平面成像与经畸变校正的快速自旋回波扩散加权成像表观扩散系数的差异
研究对32例健康检查者的正常脑行1.5-T MRI回波平面成像(EPI)与快速自旋回波(TSE)扩散加权成像,并用回顾性畸变校正技术处理数据。结果显示EPI-DWI测得的ADC显著高于TSE-DWI,且EPI-DWI的脑脊液信噪比显著高于TSE-DWI。作者结论认为在颅底附近如脑干等信噪比易降低或畸变的区域测量ADC时需谨慎。
要点
- EPI-DWI测得的ADC显著高于TSE-DWI。
- EPI-DWI的脑脊液信噪比显著高于TSE-DWI。
- 颅底附近如脑干等信噪比易降低或畸变区域测量ADC需谨慎。
核心结论
- EPI-DWI测得的ADC显著高于TSE-DWI,且EPI-DWI的脑脊液信噪比显著高于TSE-DWI;在颅底附近如脑干等信噪比易降低或畸变的区域测量ADC时需谨慎。
研究设计与样本
- 回顾性研究,纳入32例健康检查者的正常脑,在1.5-T MRI仪器上采集EPI-DWI和TSE-DWI,采用回顾性畸变校正技术(分割、边缘检测、基于Demons算法的非刚性图像配准生成位移场、图像形变)处理数据。
诊断标准与关键数值
- 从b0和b1000图像测量脑的26个部位并计算ADC,采用Wilcoxon符号秩检验比较(P=0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: We performed echo-planar imaging (EPI) and turbo spin-echo (TSE) diffusion-weighted imaging (DWI) using magnetic resonance imaging (MRI) to obtain basic clinical data of the apparent diffusion coefficient (ADC) in various parts of normal brains and compared the datasets using our retrospective distortion correction technique.
MATERIALS AND METHODS: The normal brains of 32 patients who underwent health check were scanned on a 1.5-T MRI instrument using EPI- and TSE-DWI. Distortion was corrected by (1) segmentation: the b0 images were segmented based on the plural threshold values; (2) edge detection: the edge was detected in the images obtained in step (1); (3) non-rigid image registration: non-rigid image registration using Demons algorithm was achieved between the b0 images of EPI-DWI and TSE-DWI, thereby, creating a displacement field; and (4) image warp: the displacement field was applied to the b1000 image to warp. Twenty-six parts of the brain were measured from the images of b0 and b1000 and the ADCs were calculated. The signal-to-noise ratio (SNR) of the cerebrospinal fluid was measured to identify the cause of the difference between the two sequences. These were compared using Wilcoxon signed-rank test (P = 0.05).
RESULTS: The ADC was significantly higher measured by EPI-DWI than by TSE-DWI. The SNR of EPI-DWI was significantly higher than that of the TSE-DWI.
CONCLUSION: Care must be taken when measuring ADCs near the base of the skull, such as the brain stem, where the SNR of the imaging technique is likely to decrease or distort.
引用
- Takatsu, Nakamura, Sagawa, et al. Differences in apparent diffusion coefficients between normal brain echo-planar images and turbo spin-echo diffusion-weighted images with distortion correction. Eur J Radiol. 2022. PMID: 35196616. doi:10.1016/j.ejrad.2022.110202