自由呼吸径向k空间采样磁共振成像用于新生儿和婴儿以减少麻醉
回顾性纳入23例自由呼吸径向序列与22例屏气全麻下胸部或腹部MRI的3岁以下儿童,图像质量评分相似;腹部径向T2-W序列运动伪影增加(P<0.001),自由呼吸序列总扫描时间较长(P<0.01)但总检查时间无差异(P=0.94),且减少了镇静和插管。
要点
- 自由呼吸径向T2-W与屏气T2-W SSFSE图像质量评分相似(胸部3.6 vs 3.2,P=0.07;腹部3.9 vs 3.7,P=0.66)。
- 腹部径向T2-W序列运动伪影较屏气序列增加(P<0.001),但胸部及增强T1-W序列无差异。
- 自由呼吸序列总扫描时间较长(P<0.01),但总检查时间无差异(P=0.94),并减少了镇静和插管。
核心结论
- 用于婴幼儿的运动不敏感径向序列图像质量与屏气序列相似,并减少了镇静和插管。
研究设计与样本
- 回顾性纳入3岁以下转诊行胸部或腹部MRI的儿童,自由呼吸组23例、屏气全麻组22例;自由呼吸采用径向序列(T2-W MultiVane XD和增强T1-W 3-D Vane XD),屏气组采用T2-W SSFSE和增强T1-W modified Dixon。
关键影像表现/征象
- 自由呼吸径向T2-W序列总体图像评分与屏气T2-W SSFSE序列相似(胸部3.6 vs 3.2,P=0.07;腹部3.9 vs 3.7,P=0.66)。
- 自由呼吸3-D径向T1-W序列图像质量评分与屏气T1-W序列相似(胸部4.0 vs 3.0,P=0.06;腹部3.7 vs 3.9,P=0.15)。
- 腹部径向T2-W序列运动伪影增加(P<0.001),胸部(P=0.73)及增强T1-W序列(胸部P=0.39;腹部P=0.15)运动伪影无差异。
临床应用与阅片要点
- 自由呼吸径向T2-W和增强T1-W序列在婴幼儿胸腹部MRI中可获得与屏气序列相似的图像质量。
- 腹部径向T2-W序列运动伪影增加,但胸部及增强T1-W序列运动伪影无差异。
- 自由呼吸序列总扫描时间较长(P<0.01),但总检查时间无差异(P=0.94),减少了镇静和插管。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Conventional chest and abdominal MRI require breath-holds to reduce motion artifacts. Neonates and infants require general anesthesia with intubation to enable breath-held acquisitions.
OBJECTIVE: We aimed to validate a free-breathing approach to reduce general anesthesia using a motion-insensitive radial acquisition with respiratory gating.
MATERIALS AND METHODS: We retrospectively enrolled children <3 years old who were referred for MRI of the chest or abdomen. They were divided into two groups according to MRI protocol: (1) breath-held scans under general anesthesia with T2-weighted single-shot fast spin-echo (SSFSE) and contrast-enhanced T1-weighted modified Dixon, and (2) free-breathing scans using radial sequences (T2-W MultiVane XD and contrast-enhanced T1-W three-dimensional [3-D] Vane XD). Two readers graded image quality and motion artifacts.
RESULTS: We included 23 studies in the free-breathing cohort and 22 in the breath-hold cohort. The overall imaging scores for the free-breathing radial T2-W sequence were similar to the scores for the breath-held T2-W SSFSE sequence (chest, 3.6 vs. 3.2, P=0.07; abdomen, 3.9 vs. 3.7, P=0.66). The free-breathing 3-D radial T1-W sequence also had image quality scores that were similar to the breath-held T1-W sequence (chest, 4.0 vs. 3.0, P=0.06; abdomen, 3.7 vs. 3.9, P=0.15). Increased motion was seen in the abdomen on the radial T2-W sequence (P<0.001), but increased motion was not different in the chest (P=0.73) or in contrast-enhanced T1-W sequences (chest, P=0.39; abdomen, P=0.15). The mean total sequence time was longer in free-breathing compared to breath-held exams (P<0.01); however, this did not translate to longer overall exam times (P=0.94).
CONCLUSION: Motion-insensitive radial sequences used for infants and neonates were of similar image quality to breath-held sequences and had decreased sedation and intubation.
引用
- Browne, Malone, Englund, et al. Free-breathing magnetic resonance imaging with radial k-space sampling for neonates and infants to reduce anesthesia. Pediatr Radiol. 2022. PMID: 35169882. doi:10.1007/s00247-022-05298-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。