噪声降低对虚拟单能谱图像辐射剂量降低潜力的影响:深度学习图像重建与混合迭代重建的体模图像与常规120 kVp图像比较

Impact of noise reduction on radiation dose reduction potential of virtual monochromatic spectral images: Comparison of phantom images with conventional 120 kVp images using deep learning image reconstruction and hybrid iterative reconstruction.
PubMed 文献 · Eur J Radiol · 2022-02-05

研究用成人体部体模,在12、9、6 mGy三个剂量水平下比较深度学习图像重建(DLIR)与混合迭代重建(HIR)对虚拟单能谱(VMS)图像和常规120 kVp图像质量的影响。DLIR降噪优于HIR;VMS-DLIR空间分辨率随对比度显著变化;50%剂量降低时VMS-DLIR和120 kVp-DLIR的图像质量在中对比度任务中可能下降、高对比度任务中可能提升。

要点

  • DLIR降噪幅度大于HIR,低空间频率噪声较FBP分别降低45.6%和24.4%。
  • VMS-DLIR的50%任务传递函数值随对比度变化明显:中对比度-13.2%、高对比度+25.3%。
  • 50%剂量降低时,VMS-DLIR与120 kVp-DLIR图像质量在中对比度任务可能下降、高对比度任务可能提升。

核心结论

  • DLIR降噪优于HIR;VMS-DLIR空间分辨率随对比度显著变化;50%剂量降低时VMS-DLIR和120 kVp-DLIR图像质量在中对比度任务可能下降、高对比度任务可能提升。

研究设计与样本

  • 体模研究,使用模拟成人体部的圆柱形体模,对比度设为60 HU(中对比度)和300 HU(高对比度),CT采集在12、9、6 mGy三个剂量水平下进行,图像经滤波反投影(FBP)、DLIR和HIR重建。

诊断标准与关键数值

  • DLIR和HIR低空间频率噪声幅度较FBP分别降低45.6%和24.4%;50%任务传递函数值DLIR中对比度VMS图像变化-13.2%、高对比度VMS图像变化+25.3%,中对比度120 kVp图像变化-2.0%、高对比度120 kVp图像变化+9.3%;与12 mGy FBP图像的系统性能值相比,6 mGy DLIR图像系统性能值在中对比度下降、高对比度升高。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To assess the effects of deep learning image reconstruction (DLIR) and hybrid iterative reconstruction (HIR) on the image quality of virtual monochromatic spectral (VMS) images and to investigate the dose reduction potential of the VMS and conventional 120 kVp images.

METHODS: A cylindrical phantom simulating an adult abdomen was used. The contrast was set to 60 (medium) and 300 (high) Hounsfield units. CT acquisitions were performed at three dose levels: 12, 9, and 6 mGy. Images were reconstructed via filtered back projection (FBP), DLIR, and HIR. The noise power spectrum (NPS) and task transfer function (TTF) were measured, and the system performance (SP) function was calculated (TTF2/NPS).

RESULTS: The noise magnitudes at low spatial frequencies with DLIR and HIR were lower than that with FBP by 45.6% and 24.4%, respectively. Compared to the FBP results, the TTF values at 50% with DLIR at medium and high contrast changed by -13.2% and +25.3% with the VMS images and -2.0% and +9.3% with the 120 kVp images, respectively. In the VMS and 120 kVp images, compared to the SP values of 12 mGy FBP images, SP values of 6 mGy DLIR images decreased at medium contrast and increased at high contrast.

CONCLUSIONS: DLIR achieved better noise reduction than HIR. The spatial resolution of VMS-DLIR varied significantly depending on the contrast. The image quality of VMS-DLIR and 120 kVp-DLIR potentially decrease in medium contrast tasks and increase in high contrast tasks with 50% dose reduction.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Masuda, Yamada, Minamishima, et al. Impact of noise reduction on radiation dose reduction potential of virtual monochromatic spectral images: Comparison of phantom images with conventional 120 kVp images using deep learning image reconstruction and hybrid iterative reconstruction. Eur J Radiol. 2022. PMID: 35168172. doi:10.1016/j.ejrad.2022.110198