基于虚拟平扫CT评估肝脂肪变性:首台获批临床使用的光子计数扫描仪初步经验
回顾性纳入140例患者,评估首台临床光子计数CT扫描仪生成的虚拟平扫图像对肝脂肪变性的诊断准确性。按既往报道的截断值,VNC图像上肝脏衰减CT(L)的敏感度94%、特异度87%;调整截断值后各CT指标表现良好,L的敏感度94%、特异度92%。结论认为VNC图像可用于基于图像的肝脂肪变性评估。
要点
- VNC图像上肝脏衰减CT(L)按既往截断值诊断肝脂肪变性的敏感度94%、特异度87%。
- 调整截断值后,L-S的敏感度升至96%,L/S的特异度达99.6%。
- VNC图像上L的阴性预测值达99.5%(既往截断值)或99.6%(调整截断值)。
核心结论
- 首台临床光子计数CT扫描仪获得的光谱数据集可重建出可靠的VNC图像,用于基于图像的肝脂肪变性评估;调整截断值后所有测试的CT指标均表现良好,具有优秀的敏感度和特异度。
研究设计与样本
- 回顾性研究,纳入140例患者,使用首台获批临床使用的光子计数CT扫描仪,计算VNC与真实平扫(TNC)图像上的肝脏衰减CT(L)、肝脾衰减差(L-S)和肝脾衰减比(L/S)。
诊断标准与关键数值
- 按既往报道的截断值,VNC图像诊断肝脂肪变性:L的敏感度94%、特异度87%、PPV 30%、NPV 99.5%;L-S的敏感度52%、特异度99%、PPV 81%、NPV 96%;L/S的敏感度67%、特异度98%、PPV 74%、NPV 94%。
- 按本研究调整的截断值,VNC图像诊断肝脂肪变性:L的敏感度94%、特异度92%、PPV 41%、NPV 99.6%;L-S的敏感度96%、特异度90%、PPV 46%、NPV 99.6%;L/S的敏感度95%、特异度99.6%、PPV 42%、NPV 99.6%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The present study evaluates the diagnostic accuracy of virtual non-contrast (VNC) images acquired with the first clinical photon counting CT scanner for the assessment of hepatic steatosis.
METHOD: 140 patients were retrospectively enrolled in this study. The following CT indices were calculated according to the attenuation on VNC and true non-contrast (TNC) images: Liver attenuation CT (L), difference between liver and spleen attenuation CT (L-S) and the ratio between liver and spleen attenuation CT (L/S).
RESULTS: Considering previously reported cut-off values, the sensitivity, specificity, positive and negative predictive value (PPV, NPV) for the detection of hepatic steatosis on VNC images was 94%, 87%, 30% and 99.5% for CT (L), 52%, 99%, 81% and 96% for CT (L-S) and 67%, 98%, 74% and 94% for CT (L/S). When adjusting the cut-off values according to the present study, the sensitivity, specificity, PPV and NPV for the detection of hepatic steatosis on VNC images was 94%, 92%, 41% and 99.6% for CT (L), 96%, 90%, 46% and 99.6% for CT (L-S) and 95%, 99.6%, 42% and 99.6% for CT (L/S).
CONCLUSIONS: In conclusion, the spectral datasets acquired with the first clinical PCCT scanner enable the reconstruction of VNC images that are reliable for the image-based assessment of hepatic steatosis. By adjusting the cut-off values all tested CT indices perform well and offer an excellent sensitivity as well as specificity.