光子计数探测器CT肝脏虚拟平扫成像:一项系统性体模与患者研究
研究在体模和患者中评估双源光子计数探测器CT(PCD-CT)肝脏虚拟平扫(VNC)图像的准确性。体模中PCD-CT总体平均VNC误差为4.1±3.9 HU,EID-CT在80/Sn150 kV、90/Sn150 kV、100/Sn150 kV下分别为7.5±5、6.3±4.7、6.7±4.8 HU,EID-CT各管电压下VNC误差均显著更高(P<0.001)。15例患者中真平扫与VNC图像CT衰减值无差异(P=0.093),平均VNC误差为3.7±2.2 HU。
要点
- 体模中PCD-CT总体平均VNC误差为4.1±3.9 HU,EID-CT三种管电压组合下均显著更高(P<0.001)。
- 患者中真平扫与VNC图像CT衰减值无差异(P=0.093),平均VNC误差为3.7±2.2 HU。
- PCD-CT的VNC准确性不受辐射剂量、基底物质衰减及肝脏碘含量影响。
核心结论
- 光子计数探测器CT可在体模和患者中重建肝脏VNC图像,衰减值准确,且不依赖于辐射剂量、基底物质衰减和肝脏碘含量。
研究设计与样本
- 体模研究使用含肝实质(1.4 mgI/mL)和19个肝脏病灶(碘含量0-5 mgI/mL)的拟人腹部体模,在临床双源PCD-CT(管电压120 kV)及双源能量积分探测器(EID)CT双能模式(80/Sn150 kV、90/Sn150 kV、100/Sn150 kV)下成像,添加脂肪等效环模拟小、中、大三种体型,每种设置于5、10、15 mGy三个辐射剂量(容积CT剂量指数)下成像。
- 患者部分回顾性纳入15例接受PCD-CT临床扫描的乏血供肝脏病灶患者,比较门静脉期CT重建的VNC图像与真平扫图像中病灶及肝实质的衰减值。
- 统计分析采用方差分析及事后t检验、广义线性模型评估剂量、患者体型、基底物质、碘含量及扫描仪/扫描模式对定量准确性的影响。
诊断标准与关键数值
- 体模中PCD-CT总体平均VNC误差为4.1±3.9 HU。
- 体模中EID-CT总体平均VNC误差在80/Sn150 kV为7.5±5 HU、90/Sn150 kV为6.3±4.7 HU、100/Sn150 kV为6.7±4.8 HU,各管电压设置下EID-CT的VNC误差均显著更高(P<0.001),调整剂量、体型、病灶碘含量及基底物质衰减后仍显著。
- PCD-CT中较小体模尺寸与更高定量准确性相关(P=0.007-0.046),辐射剂量对准确性无影响(P>0.126)。
- EID-CT(而非PCD-CT)中VNC误差随病灶碘含量增加而升高(P<0.001)。
- 患者中真平扫与VNC图像衰减值无差异(P=0.093),平均VNC误差为3.7±2.2 HU。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to assess the accuracy of virtual noncontrast (VNC) images of the liver in a phantom and patients using dual-source photon-counting detector computed tomography (PCD-CT).
MATERIALS AND METHODS: An anthropomorphic abdominal phantom with a liver insert containing liver parenchyma (1.4 mgI/mL) and 19 liver lesions (iodine content 0-5 mgI/mL) was imaged on a clinical dual-source PCD-CT (tube voltage 120 kV) and in the dual-energy mode on a dual-source energy-integrating detector (EID) CT (tube voltage combinations, 80/Sn150 kV, 90/Sn150 kV, and 100/Sn150 kV). Rings of fat-equivalent material were added to the phantom to emulate 3 sizes (small, medium, large). Each setup was imaged at 3 different radiation doses (volume CT dose index: 5, 10, and 15 mGy). Virtual noncontrast images were reconstructed and CT attenuation was measured in each lesion and liver parenchyma. The absolute error of CT attenuation (VNCerror) was calculated using the phantom specifications as reference. In addition, 15 patients with hypodense liver lesions who were clinically scanned on PCD-CT were retrospectively included. Attenuation values in lesions and liver parenchyma in VNC images reconstructed from portal venous phase CT were compared with true noncontrast images. Statistical analysis included analysis of variance with post hoc t tests and generalized linear models to assess the impact of various variables (dose, patient size, base material, iodine content, and scanner/scan mode) on quantification accuracy.
RESULTS: In the phantom, the overall mean VNCerror for PCD-CT was 4.1 ± 3.9 HU. The overall mean VNCerror for EID-CT was 7.5 ± 5, 6.3 ± 4.7, and 6.7 ± 4.8 HU for 80/Sn150 kV, 90/Sn150 kV, and 100/Sn150 kV, respectively, with the VNCerror of EID-CT being significantly higher at all tube voltage settings (P < 0.001), even after adjusting for dose, size, iodine content of the lesion, and attenuation of base material. For PCD-CT, a smaller phantom size was associated with higher quantification accuracy (P = 0.007-0.046), whereas radiation dose did not impact accuracy (P > 0.126). For EID-CT, but not for PCD-CT, VNCerror increased with lesion iodine content (P < 0.001). In patients, there was no difference in attenuation measured on true noncontrast and VNC images (P = 0.093), with a mean VNCerror of 3.7 ± 2.2 HU.
CONCLUSIONS: Photon-counting detector CT allows for the reconstruction of VNC images of the liver both in a phantom and in patients with accurate attenuation values, being independent of dose, attenuation of base material, and liver iodine content.