文献速递腹部技术与安全CT

腹部低剂量CT:新型光子计数探测器CT的初步经验及与能量积分探测器CT的比较

Low-dose CT of the abdomen: Initial experience on a novel photon-counting detector CT and comparison with energy-integrating detector CT.
PubMed 文献 · Eur J Radiol · 2022-01-29

连续纳入临床指征行低剂量腹部CT的患者在新型光子计数探测器CT(PCD-CT)上扫描,并与同一时间段BMI匹配的能量积分探测器CT(EID-CT)队列比较。PCD-CT图像噪声显著更低、SNR显著更高、主观图像质量更高,肾盂、输尿管和肠系膜血管的可见性更优,肾上腺可见性无显著差异。

要点

  • PCD-CT低剂量腹部CT噪声更低(24.9±3.3 vs 31.4±5.6 SD HU,p<0.001),SNR更高(2.1±0.3 vs 1.5±0.4,p<0.001)。
  • PCD-CT主观图像质量更高(4.0±0.3 vs 3.1±0.6,p<0.001),肾盂、输尿管和肠系膜血管可见性更优。
  • 随BMI增加(第1至第4 BMI四分位),EID-CT噪声增加和SNR下降幅度大于PCD-CT,而辐射剂量增加幅度相当。

核心结论

  • 在BMI匹配的可比患者中,新型PCD-CT行低剂量腹部CT较EID-CT噪声更低、SNR更高、主观图像质量更高,腹部精细结构可见性更优。

研究设计与样本

  • 连续纳入临床指征行低剂量CT的患者在PCD-CT上扫描,与同一时间段BMI匹配的EID-CT队列比较;共纳入20例患者(平均年龄46.2岁,范围19-77岁;男性13例)。

诊断标准与关键数值

  • 图像质量及腹部结构(肾上腺、肠系膜血管、输尿管、肾盂)可见性采用5点Likert量表评估(1=质量很差/不可检测;5=质量极好/可区分)。
  • PCD-CT图像噪声24.9±3.3 SD HU vs EID-CT 31.4±5.6 SD HU,p<0.001。
  • PCD-CT SNR 2.1±0.3 vs EID-CT 1.5±0.4,p<0.001。
  • PCD-CT主观图像质量4.0±0.3 vs EID-CT 3.1±0.6,p<0.001。
  • 随BMI增加(第1至第4 BMI四分位),噪声变化EID-CT 39% vs PCD-CT 2%,SNR变化EID-CT -33% vs PCD-CT -7%,辐射剂量增加EID-CT 70% vs PCD-CT 59%。

临床应用与阅片要点

  • PCD-CT低剂量腹部CT可提供更低噪声和更高SNR的图像,有助于腹部精细结构(肾盂、输尿管、肠系膜血管)的显示。
  • 肾上腺的可见性在PCD-CT与EID-CT之间无显著差异。
  • 随BMI增加,EID-CT噪声增加和SNR下降幅度大于PCD-CT,而两者辐射剂量增加幅度相当。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To analyze the quantitative and qualitative image quality of low-dose CT scans of the abdomen on a novel photon-counting detector CT (PCD-CT) in comparison with a traditional energy-integrating detector CT (EID-CT).

METHODS: Consecutive patients with clinically indicated low-dose CT were scanned on a PCD-CT and compared with a BMI-matched EID-CT-cohort scanned during the same timeframe. Radiation dose, image noise, and signal-to-noise ratio (SNR) were measured for each patient. Furthermore, image quality and conspicuity of abdominal structures (adrenal glands, mesenteric vessels, ureters, and renal pelvis) were assessed on 5-point Likert-scales (1 = very poor quality/not detectable; 5 = excellent quality/differentiability).

RESULTS: Twenty patients (mean age 46.2 [range: 19-77]; 13 men) were included. Image noise was significantly lower (24.9 ± 3.3 vs. 31.4 ± 5.6 SD HU, p < 0.001) and SNR significantly higher (2.1 ± 0.3 vs. 1.5 ± 0.4; p < 0.001) on the PCD-CT. Subjective image quality was substantially higher (4.0 ± 0.3 vs. 3.1 ± 0.6; p < 0.001) and conspicuity better for the renal pelvis, ureters, and mesenteric vessels on the PCD-CT. There was no significant difference in the conspicuity of the adrenal glands. With increasing BMI (1st-4th BMI quartile), noise increased, and SNR decreased more strongly on the EID-CT than on the PCD-CT (ΔNoise: 39% vs. 2%, ΔSNR: -33% vs. -7% for EID-CT vs. PCD-CT, respectively) while radiation dose increased comparably (70 vs. 59%).

CONCLUSIONS: Low-dose CT scans of the abdomen performed on a novel PCD-CT exhibit reduced noise, higher SNR, increased subjective image quality, and superior conspicuity of abdominal fine structures compared to scans in comparable patients on an EID-CT.

来源:Eur J Radiol,查看原文。版权归原作者所有。