在重症监护病房婴儿胸部X线摄影中采用高峰值千伏技术降低辐射剂量
该回顾性研究纳入376张婴儿胸部X线片,按峰值千伏是否≥60 kV分为高kV组(107张)与低kV组(269张),比较两组空气比释动能、有效剂量与图像质量评分。高kV组平均空气比释动能22.9 µGy,低于低kV组的56.6 µGy(P<0.0001),两组有效剂量平均差值为11.68(P<0.0001),图像质量评分平均差值为0.06(±0.03,P=0.10),差异无统计学意义。结论认为高峰值千伏技术可能在不损害图像诊断价值的前提下显著降低辐射剂量。
要点
- 高kV组(≥60 kV)平均空气比释动能为22.9 µGy,低kV组(<60 kV)为56.6 µGy,差异有统计学意义(P<0.0001)。
- 两组图像质量评分平均差值为0.06(±0.03,P=0.10),差异无统计学意义。
- 高峰值千伏技术可能在不损害图像诊断价值的前提下显著降低辐射剂量。
核心结论
- 高峰值千伏技术可能在不损害图像诊断价值的前提下实现辐射剂量的统计学显著降低。
研究设计与样本
- 回顾性研究,纳入376张于本机构儿科重症监护病房拍摄的胸部X线片,按峰值千伏是否≥60 kV分为高kV组(107张)与低kV组(269张),由两名经过专科培训的儿科放射科医生盲法评估图像质量。
诊断标准与关键数值
- 高kV组(≥60 kV)平均空气比释动能22.9 µGy(范围11.8-34.4),低kV组(<60 kV)平均空气比释动能56.6 µGy(范围30.7-81.9),两组比较P<0.0001;两组有效剂量平均差值为11.68(P<0.0001);两组图像质量评分平均差值为0.06(±0.03,P=0.10),差异无统计学意义。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Chest radiography is an important tool in the care of infants in intensive care units. Image optimization must be monitored to minimize radiation exposure in this susceptible population.
OBJECTIVE: To examine the use of a high tube peak kilovoltage technique to achieve radiation dose reduction while maintaining adequate image quality.
MATERIALS AND METHODS: A retrospective study was conducted. Radiation doses of chest radiographs performed in the pediatric intensive care units in our institution were calculated. The radiographs were divided into two groups based on the value of the peak kilovoltage used: above and below 60 kilovolts (kV). Image quality was blindly assessed by two fellowship-trained pediatric radiologists. Air kerma, effective dose and quality score for the high versus the low peak kilovoltage group were compared and analyzed.
RESULTS: The study included 376 radiographs. One hundred and seven radiographs were performed using peak kilovoltage values equal to or above 60 kV and 269 radiographs were performed using values under 60 kV. The average air kerma for the lower peak kilovoltage group was 56.6 microgray (µGy) (30.7-81.9) vs. 22.9 µGy (11.8-34.4) for the higher peak kilovoltage group (P<0.0001). The mean difference in effective dose between the groups was 11.68 (P<0.0001). The mean difference for the quality score was 0.06 (±0.03, P=0.10), not statistically significant.
CONCLUSION: A high peak kilovoltage technique may enable a statistically significant radiation dose reduction without compromising the diagnostic value of the image.
引用
- Yahav-Dovrat, Elbakri, Rozovsky, et al. Radiation dose reduction for chest radiography of infants in intensive care units using a high peak kilovoltage-technique. Pediatr Radiol. 2022. PMID: 35106646. doi:10.1007/s00247-021-05262-x