文献速递腹部技术与安全CT

量子迭代重建用于腹部光子计数探测器CT可改善图像质量

Quantum Iterative Reconstruction for Abdominal Photon-counting Detector CT Improves Image Quality.
PubMed 文献 · Radiology · 2022-02-01

本研究在体模和接受门静脉期腹部光子计数探测器CT的肿瘤患者中,评估量子迭代重建(QIR)用于虚拟单能图像(60 keV)和多色图像(T3D)的图像质量与最佳强度水平。回顾性纳入50例患者,结果显示QIR-4较未使用QIR时噪声指数降低45%(60 keV)和44%(T3D),肝脏CNR提高74%和69%,且不改变噪声纹理和CT衰减值,四位阅片者均认为QIR-4病灶显示最佳。

要点

  • 门静脉期腹部PCD CT中,QIR-4较QIR-off降低噪声指数45%(60 keV)和44%(T3D)。
  • QIR-4较QIR-off提高肝脏CNR 74%(60 keV)和69%(T3D),脂肪和肝脏平均衰减无差异。
  • 四位阅片者均评价QIR-4在病灶显示上优于其他强度(P值范围<.001至.04)。

核心结论

  • 在门静脉期腹部光子计数探测器CT中,高强度水平的量子迭代重建(QIR;Siemens Healthcare)通过降低噪声、提高对比噪声比和病灶显示改善图像质量,且不损害图像纹理或CT衰减值。

研究设计与样本

  • 回顾性研究,在充水体模中测量噪声功率谱,并连续纳入2021年3月至4月间接受门静脉期腹部PCD CT的肿瘤患者50例(平均年龄59岁±16[标准差],31例女性)。

关键影像表现/征象

  • 体模中噪声功率谱显示各重建间噪声纹理不变,最大空间频率差异为0.01每毫米。
  • 60 keV和T3D图像中,QIR-4较QIR-off的全局噪声指数分别降低45%和44%(均P<.001)。
  • QIR-4较QIR-off的肝脏CNR分别提高74%(60 keV)和69%(T3D)(均P<.001)。
  • 脂肪和肝脏的平均衰减在重建间无差异证据(P=.79-.84),体素水平亦无差异。
  • 定性评估中,QIR-4在60 keV和T3D的每个类别均优于所有其他重建(P值范围<.001至.01)。

诊断标准与关键数值

  • 以未使用QIR的重建(QIR-off)为参考标准,比较QIR 1-4四个强度水平(指标测试)。
  • 噪声指数在充水体模中测量,图像质量定量指标包括全局噪声指数、CNR和体素水平CT衰减差异。
  • 定性评估包括噪声与纹理、伪影、诊断置信度和整体质量;病灶显示由四位阅片者评级。

临床应用与阅片要点

  • 在门静脉期腹部PCD CT中,采用高强度QIR(QIR-4)可在不改变噪声纹理和CT衰减值的情况下降低噪声并提高肝脏CNR。
  • QIR-4较其他强度水平提高低密度肝脏病灶的显示,四位阅片者均评价其更优(P值范围<.001至.04)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background An iterative reconstruction (IR) algorithm was introduced for clinical photon-counting detector (PCD) CT. Purpose To investigate the image quality and the optimal strength level of a quantum IR algorithm (QIR; Siemens Healthcare) for virtual monoenergetic images and polychromatic images (T3D) in a phantom and in patients undergoing portal venous abdominal PCD CT. Materials and Methods In this retrospective study, noise power spectrum (NPS) was measured in a water-filled phantom. Consecutive oncologic patients who underwent portal venous abdominal PCD CT between March and April 2021 were included. Virtual monoenergetic images at 60 keV and T3D were reconstructed without QIR (QIR-off; reference standard) and with QIR at four levels (QIR 1-4; index tests). Global noise index, contrast-to-noise ratio (CNR), and voxel-wise CT attenuation differences were measured. Noise and texture, artifacts, diagnostic confidence, and overall quality were assessed qualitatively. Conspicuity of hypodense liver lesions was rated by four readers. Parametric (analyses of variance, paired t tests) and nonparametric tests (Friedman, post hoc Wilcoxon signed-rank tests) were used to compare quantitative and qualitative image quality among reconstructions. Results In the phantom, NPS showed unchanged noise texture across reconstructions with maximum spatial frequency differences of 0.01 per millimeter. Fifty patients (mean age, 59 years ± 16 [standard deviation]; 31 women) were included. Global noise index was reduced from QIR-off to QIR-4 by 45% for 60 keV and by 44% for T3D (both, P < .001). CNR of the liver improved from QIR-off to QIR-4 by 74% for 60 keV and by 69% for T3D (both, P < .001). No evidence of difference was found in mean attenuation of fat and liver (P = .79-.84) and on a voxel-wise basis among reconstructions. Qualitatively, QIR-4 outperformed all reconstructions in every category for 60 keV and T3D (P value range, <.001 to .01). All four readers rated QIR-4 superior to other strengths for lesion conspicuity (P value range, <.001 to .04). Conclusion In portal venous abdominal photon-counting detector CT, an iterative reconstruction algorithm (QIR; Siemens Healthcare) at high strength levels improved image quality by reducing noise and improving contrast-to-noise ratio and lesion conspicuity without compromising image texture or CT attenuation values. © RSNA, 2022 Online supplemental material is available for this article. See also the editorial by Sinitsyn in this issue.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Sartoretti, Landsmann, Nakhostin, et al. Quantum Iterative Reconstruction for Abdominal Photon-counting Detector CT Improves Image Quality. Radiology. 2022. PMID: 35103540. doi:10.1148/radiol.211931

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。