五种移动X线摄影设备散射校正软件的特性分析:体模研究

Characterizing Scatter Correction Software of 5 Mobile Radiography Units: An Anthropomorphic Phantom Study.
PubMed 文献 · Invest Radiol · 2022-01-27

研究用体模对5种移动X线摄影设备的无滤线栅、散射校正软件及物理滤线栅图像进行图像质量评估算法分析,并与3名放射科医师的主观评分验证。结果显示评估算法与观察者结果相关性好(b≥0.89,R2≥0.83);各厂商散射校正软件均可提高无滤线栅图像的7项物理量,部分可达物理滤线栅水平;降低管负荷的散射校正图像对比度、细节、锐度和噪声均高于高管负荷无滤线栅图像。研究提示应用散射校正软件可能降低患者剂量。

要点

  • 图像质量评估算法与3名放射科医师主观评分相关性好(b≥0.89,R2≥0.83)。
  • 各厂商散射校正软件均可提高无滤线栅图像的7项物理量。
  • 部分厂商散射校正软件可使物理量达到物理滤线栅图像水平。

核心结论

  • 概念验证表明,该图像质量评估算法可用于表征不同厂商的不同系列图像;基于物理量,散射校正软件提高对比度、细节、锐度和噪声;实验结果假定应用散射校正软件时可能降低患者剂量。

研究设计与样本

  • 体模研究,使用多功能胸部体模(N1 Lungman,Kyoto Kagaku),在壁式滤线栅系统中用5种不同厂商的平板探测器获取无滤线栅、散射校正软件及物理滤线栅系列图像;图像质量评估算法由本院实施但 originally 由杜克大学医学中心研究组开发;3名有经验的放射科医师进行验证。

关键影像表现/征象

  • 散射校正软件可提高无滤线栅图像的肋肺对比度、膈下肺对比度、肺细节、纵隔细节、肺噪声、纵隔噪声和肋肺锐度7项物理量。
  • 部分厂商的散射校正软件可使物理量达到物理滤线栅图像水平。
  • 降低管负荷的散射校正图像对比度、细节、锐度和噪声均高于高管负荷的无滤线栅图像。

诊断标准与关键数值

  • 图像质量评估算法计算7项物理量:肋肺对比度(RLcontrast)、膈下肺对比度(SLcontrast)、肺细节(Ldetail)、纵隔细节(Mdetail)、肺噪声(Lnoise)、纵隔噪声(Mnoise)、肋肺锐度(RLsharpness)。
  • 观察者结果与图像质量评估算法的线性回归系数b≥0.89,R2≥0.83。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Bedside radiographs are usually obtained gridless, without a physical scatter correction grid because of several limitations. Therefore, multiple manufacturers of mobile radiography systems provide the possibility to apply scatter correction software (SC SW) on those images. The purpose of this study was to characterize different series of radiographs-gridless, SC SW, and physical grid-with an image quality assessment algorithm (IQAA). Furthermore, we investigated the potential dose reduction and the correlation between the output of the IQAA and the human observers.

MATERIALS AND METHODS: We obtained different series of radiographs with an anthropomorphic phantom (multipurpose chest phantom N1 "Lungman," Kyoto Kagaku, Kyoto, Japan). All radiographs were obtained with flat-panel detectors of 5 different manufacturers in a wall bucky system. An IQAA to analyze the radiographs was implemented in our department but was originally developed by the research group of the Duke University Medical Center. Seven physical quantities were calculated by the IQAA: rib-lung contrast (RLcontrast), subdiaphragm-lung contrast (SLcontrast), lung detail (Ldetail), mediastinum detail (Mdetail), lung noise (Lnoise), mediastinum noise (Mnoise), and rib-lung sharpness (RLsharpness). In a proof of concept, the results of the IQAA were validated by 3 experienced radiologists.

RESULTS: Regression coefficients (b) of the linear regression model indicate that the human observer results correlate well with the IQAA (b ≥ 0.89, R2 ≥ 0.83). All manufacturers have SC SW that increases the 7 physical quantities of the gridless images. However, several manufacturers have SC SW that increases the physical metrics to the same level as the physical grid images. The SC SW radiographs obtained with a reduced tube load have an increased level of contrast, detail, sharpness, and noise compared with the gridless images obtained with the higher tube load.

CONCLUSIONS: We have proven in a proof of concept that the originally developed IQAA can be used to characterize different series of images of different manufacturers. Based on the physical quantities, SC SW increases the contrast, detail, sharpness, and noise. The experimental results in this study assume a patient dose reduction could be possible when SC SW is applied.

来源:Invest Radiol,查看原文。版权归原作者所有。