单中心评估对比增强数字化乳腺摄影(CEDM)、数字化乳腺摄影(DM)与数字乳腺断层合成(DBT)的乳腺辐射剂量比较

Single Center Evaluation of Comparative Breast Radiation dose of Contrast Enhanced Digital Mammography (CEDM), Digital Mammography (DM) and Digital Breast Tomosynthesis (DBT).
PubMed 文献 · Acad Radiol · 2022-01-19

本回顾性研究于2019年1月至12月分析本院6214次DM、3662次DBT和173次CEDM检查,比较不同乳腺厚度、年龄和密度患者的方案平均腺体剂量(AGD)。CEDM短方案AGD中位数4.5 mGy,较DM的2.8 mGy高59%,较DM+DBT的6.0 mGy低40%,差异有统计学意义(P<0.001);致密乳腺的CEDM短方案AGD中位数4.2 mGy低于非致密乳腺的4.7 mGy(P=0.039)。

要点

  • CEDM短方案AGD中位数4.5 mGy,较DM的2.8 mGy高59%,较DM+DBT的6.0 mGy低40%(P<0.001)。
  • CEDM短方案每标准乳腺投影AGD均低于国际法规要求的3 mGy限值。
  • 致密乳腺CEDM短方案AGD中位数4.2 mGy,低于非致密乳腺的4.7 mGy(P=0.039)。

核心结论

  • CEDM短方案辐射暴露高于常规DM和DBT,但低于Combo模式;无论乳腺大小,致密乳腺患者接受CEDM短方案的剂量更低。

研究设计与样本

  • 单中心回顾性研究,2019年1月至2019年12月共分析6214次DM、3662次DBT和173次CEDM检查,每侧乳腺共分析23383次检查。

诊断标准与关键数值

  • 各方案总AGD中位数:DM 2.8 mGy、DBT 3.2 mGy、DM+DBT 6.0 mGy、CEDM短方案4.5 mGy、CEDM短方案联合DBT 7.4 mGy、CEDM长方案8.4 mGy、CEDM长方案联合DBT 11.6 mGy。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: The aim of this retrospective study is to compare the radiation dose received during CEDM, short and long protocol (CEDM SP and CEDM LP), with dose received during DM and DBT on patients with varying breast thickness, age and density.

MATERIALS AND METHODS: Between January 2019 and December 2019, patients having 6214 DM, 3662 DBT and 173 CEDM examinations in our department were analyzed. Protocol total single breast AGD has been evaluated for all clinical imaging protocols, extracting AGD values and exposure data from the dose DICOM Structured Report (SR) information stored in the hospital PACS system. Protocol AGD was calculated as the sum of single projection AGDs carried out in every exam for each clinical protocol. A total amount of 23,383 exams for each breast were analyzed. Protocol AGDs, stratified as a function of patient breast compression thickness, age, and breast density were assessed.

RESULTS: The total protocol AGD median values for each protocol are: 2.8 mGy for DM, 3.2 mGy for DBT, 6.0 mGy for DM+DBT, 4.5 mGy for CEDM SP, 7.4 mGy for CEDM SP_DBT (CEDM SP protocol with DBT), 8.4 mGy for CEDM LP and 11.6 mGy for CEDM LP_DBT (CEDM LP protocol with DBT). CEDM SP AGD median value is 59% higher than DM AGD median value and 40% lesser than DM+DBT AGD median; this last difference was statistically confirmed with a p-value <0.001. AGD value for each standard breast CEDM SP projection results to be below 3-mGy limit. AGD value for each standard breast CEDM SP projection results to be below 3 mGy, as required by international legislation. For dense breasts, the AGD median value is 4.2 mGy, with the first and third quartile of 3.3 mGy and 6.0 mGy respectively; for non-dense breasts, the AGD median value is 4.7 mGy, with first and third quartile of 3.5 mGy and 6.3 mGy respectively. The difference between the two groups was statistically tested and confirmed, with a p-value of 0.039.

CONCLUSION: CEDM SP results in higher radiation exposure compared with conventional DM and DBT but lower than the Combo mode. The dose administered during the CEDM SP is lower in patients with dense breasts regardless of their size. An interesting outcome, considering the ongoing studies on CEDM screening in patients with dense breasts.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Bicchierai, Busoni, Tortoli, et al. Single Center Evaluation of Comparative Breast Radiation dose of Contrast Enhanced Digital Mammography (CEDM), Digital Mammography (DM) and Digital Breast Tomosynthesis (DBT). Acad Radiol. 2022. PMID: 35065889. doi:10.1016/j.acra.2021.12.022