超声计算机辅助诊断系统鉴别乳腺肿块的诊断效能:一项前瞻性多中心研究

The diagnostic performance of ultrasound computer-aided diagnosis system for distinguishing breast masses: a prospective multicenter study.
PubMed 文献 · Eur Radiol · 2022-01-23

这项前瞻性多中心研究纳入2019年4月至2020年11月间计划行病理诊断的901例乳腺肿块患者,评估超声CAD软件鉴别良恶性及避免不必要活检的价值。CAD长轴切面准确率89.6%、敏感度94.2%、特异度87.0%;以BI-RADS 4a为界值CAD的AUC高于有经验和经验较少的放射科医生,应用CAD后BI-RADS 4类和5类的不必要活检率显著下降。

要点

  • CAD软件长轴切面鉴别乳腺肿块良恶性的准确率为89.6%、敏感度为94.2%、特异度为87.0%。
  • 以BI-RADS 4a为界值,CAD的AUC为0.906,高于有经验医生的0.734和经验较少医生的0.696(均p<0.001)。
  • 应用CAD后BI-RADS 4类和5类不必要活检率分别由33.0%降至11.9%、由37.8%降至14.5%。

核心结论

  • 超声CAD软件可作为鉴别乳腺肿块良恶性的有效辅助诊断工具,并减少不必要活检。

研究设计与样本

  • 前瞻性多中心研究,纳入2019年4月至2020年11月间计划行病理诊断的901例患者共901个乳腺肿块,获取超声图像、视频、CAD分析和BI-RADS分类,计算诊断效能并与放射科医生比较,临床试验注册号NCT03887598。

诊断标准与关键数值

  • CAD软件长轴切面准确率89.6%、敏感度94.2%、特异度87.0%、PPV 80.4%、NPV 96.3%;短轴切面分别为89.0%、91.4%、87.7%、80.8%、94.7%。

临床应用与阅片要点

  • 以BI-RADS 4a为界值时CAD的AUC为0.906,高于有经验医生0.734和经验较少医生0.696(均p<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the diagnostic value of computer-aided diagnosis (CAD) software on ultrasound in distinguishing benign and malignant breast masses and avoiding unnecessary biopsy.

METHODS: This prospective, multicenter study included patients who were scheduled for pathological diagnosis of breast masses between April 2019 and November 2020. Ultrasound images, videos, CAD analysis, and BI-RADS were obtained. The AUC, accuracy, sensitivity, specificity, PPV, and NPV were calculated and compared with radiologists.

RESULTS: Overall, 901 breast masses in 901 patients were enrolled in this study. The accuracy, sensitivity, specificity, PPV and NPV of CAD software were 89.6%, 94.2%, 87.0%, 80.4%, and 96.3, respectively, in the long-axis section; 89.0%, 91.4%, 87.7%, 80.8%, and 94.7%, respectively, in the short-axis section. With BI-RADS 4a as the cut-off value, CAD software has a higher AUC (0.906 vs 0.734 vs 0.696, all p < 0.001) than both experienced and less experienced radiologists. With BI-RADS 4b as the cut-off value, CAD software showed better AUC than less experienced radiologists (0.906 vs 0.874, p < 0.001), but not superior to experienced radiologists (0.906 vs 0.883, p = 0.057). After the application of CAD software, the unnecessary biopsy rate of BI-RADS categories 4 and 5 was significantly decreased (33.0% vs 11.9%, 37.8% vs 14.5%), and the malignant rate of biopsy in category 4a was significantly increased (11.6% vs 40.7%, 7.4% vs 34.9%, all p < 0.001).

CONCLUSIONS: CAD software on ultrasound can be used as an effective auxiliary diagnostic tool for differential diagnosis of benign and malignant breast masses and reducing unnecessary biopsy.

CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov (NCT03887598) KEY POINTS: • Prospective multicenter study showed that computer-aided diagnosis software provides greater diagnostic confidence for differentiating benign and malignant breast masses. • Computer-aided diagnosis software can help radiologists reduce unnecessary biopsy. • The management of patients with breast masses becomes more appropriate.

来源:Eur Radiol,查看原文。版权归原作者所有。