不同管电压下腹部多排CT深度学习图像重建算法:体模研究中的图像质量与辐射剂量评估
在定制体模上比较不同管电压(120、100、80 kVp)和管电流(200、100、60 mA)下深度学习图像重建(DLIR)与迭代重建(IR)及滤波反投影(FBP)的图像质量和辐射剂量。结果显示DLIR较FBP和混合IR降低噪声和伪影、提高整体图像质量,低剂量DLIR在降低约70%辐射剂量时仍可维持与常规剂量FBP相当的图像质量,但图像锐利度有所下降。
要点
- DLIR较FBP和混合IR降低图像噪声、提高对比噪声比和整体图像质量。
- 低剂量DLIR在辐射剂量降低约70%时仍可维持与常规剂量FBP相当的图像质量。
- DLIR图像锐利度较参考图像下降(p < 0.05),但中等强度DLIR整体图像质量更优。
核心结论
- 与FBP和混合IR相比,新型DLIR算法降低噪声和伪影并提高整体图像质量;尽管DLIR图像锐利度下降,低剂量DLIR似乎具有更大的剂量优化潜力。
研究设计与样本
- 体模研究,使用定制体模在120、100、80 kVp管电压和200、100、60 mA管电流下扫描,重建FBP、混合IR(30%和50%)及DLIR(低、中、高强度),以120 kVp/200 mA下FBP图像为参考图像,定性和定量评估图像噪声、对比噪声比、锐利度、伪影和整体图像质量,并用容积CT剂量指数评估辐射剂量。
关键影像表现/征象
- 与参考图像相比,低剂量CT联合DLIR显著降低噪声和伪影并提高整体图像质量(p < 0.05)。
- 低剂量CT联合DLIR图像锐利度下降(p < 0.05)。
- 低剂量CT联合DLIR可维持与常规剂量CT联合FBP相当的图像质量,尤其在使用中等强度水平时。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare the image quality and radiation dose of a deep learning image reconstruction (DLIR) algorithm compared with iterative reconstruction (IR) and filtered back projection (FBP) at different tube voltages and tube currents.
MATERIALS AND METHODS: A customized body phantom was scanned at different tube voltages (120, 100, and 80 kVp) with different tube currents (200, 100, and 60 mA). The CT datasets were reconstructed with FBP, hybrid IR (30% and 50%), and DLIR (low, medium, and high levels). The reference image was set as an image taken with FBP at 120 kVp/200 mA. The image noise, contrast-to-noise ratio (CNR), sharpness, artifacts, and overall image quality were assessed in each scan both qualitatively and quantitatively. The radiation dose was also evaluated with the volume CT dose index (CTDIvol) for each dose scan.
RESULTS: In qualitative and quantitative analyses, compared with reference images, low-dose CT with DLIR significantly reduced the noise and artifacts and improved the overall image quality, even with decreased sharpness (p < 0.05). Despite the reduction of image sharpness, low-dose CT with DLIR could maintain the image quality comparable to routine-dose CT with FBP, especially when using the medium strength level.
CONCLUSION: The new DLIR algorithm reduced noise and artifacts and improved overall image quality, compared to FBP and hybrid IR. Despite reduced image sharpness in CT images of DLIR algorithms, low-dose CT with DLIR seems to have an overall greater potential for dose optimization.
KEY POINTS: • Using deep learning image reconstruction (DLIR) algorithms, image quality was maintained even with a radiation dose reduced by approximately 70%. • DLIR algorithms yielded lower image noise, higher contrast-to-noise ratios, and higher overall image quality than FBP and hybrid IR, both subjectively and objectively. • DLIR algorithms can provide a better image quality, much better than FBP and even better than hybrid IR, while facilitating a reduction in radiation dose.