数字化动脉血管结构与力学特性自动评估系统的验证与可行性:超高频超声研究
本研究以超高频超声(VevoMD)采集指动脉扫描,用Carotid Studio软件自动分析并与手动测量比较,评估直径、扩张性和内中膜厚度的Bland-Altman一致性及观察者内/间变异系数。结果显示两种方法间无显著偏倚且一致性界限可接受;舒张末直径与内中膜厚度的观察者内变异系数分别为4.1%和4.2%,观察者间分别为7.3%和5.4%;扩张性的观察者内和观察者间变异系数较高,分别为25.7%和26.7%。作者认为该软件是研究指动脉超高频超声扫描的有效且可重复的工具。
要点
- Carotid Studio自动分析与手动测量指动脉直径、扩张性和内中膜厚度间无显著偏倚。
- 舒张末直径与内中膜厚度的观察者内变异系数分别为4.1%和4.2%。
- 扩张性的观察者内和观察者间变异系数较高,分别为25.7%和26.7%。
核心结论
- Carotid Studio软件是研究指动脉超高频超声扫描的有效且可重复的工具,可能对中型动脉罕见血管疾病及动脉老化病理生理研究有潜在用途。
研究设计与样本
- 验证性研究,采用超高频超声(VevoMD,Visualsonics-Fujifilm,加拿大多伦多)采集指动脉扫描,以手动测量为参考,用Carotid Studio软件(Quipu,意大利比萨)自动分析。
- 一致性评估采用Bland-Altman分析,重复性评估采用变异系数。
关键影像表现/征象
- 超高频超声可清晰显示人小型肌性动脉。
诊断标准与关键数值
- Carotid Studio与手动分析在直径、扩张性和内中膜厚度测量上无趋势或显著偏倚,所有一致性界限均可接受。
- 舒张末直径观察者内变异系数4.1%,观察者间变异系数7.3%。
- 内中膜厚度观察者内变异系数4.2%,观察者间变异系数5.4%。
- 扩张性观察者内变异系数25.7%,观察者间变异系数26.7%。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Ultrahigh-frequency ultrasound (UHFUS) allows sharp visualization of human small muscular arteries. This may help in elucidating some aspects of the pathophysiology of arterial aging, such as the stiffness gradient between large and small conduit arteries and its consequences on the microcirculation, as well as vascular diseases affecting medium-sized arteries. However, UHFUS use is still limited, partly because of the lack of validated tools to quantify vascular structure and mechanical properties of small muscular arteries. In this validation study, scans of digital arteries were obtained with UHFUS (VevoMD, Visualsonics-Fujifilm, Toronto, ON, Canada), analyzed using Carotid Studio software (Quipu, Pisa, Italy) and compared with the manual measurement. Agreement between the two techniques on measures of diameter, distension and intima-media thickness (IMT) was evaluated using Bland-Altman analyses; inter- and intra-operator reproducibility was evaluated using coefficients of variation (CVs). Overall, no trend or significant bias was observed between Carotid Studio and manual analysis. All limits of agreement were acceptable. The intra-observer CV of diastolic diameter and IMT were 4.1% and 4.2%, respectively. The inter-observer CV for diastolic diameter and IMT were 7.3% and 5.4%, respectively. Intra- and inter-observer CVs for distension were higher (25.7% and 26.7%, respectively). These results suggest that the Carotid Studio software is a valid and reproducible tool to study UHFUS scans of digital arteries, with potential utility both in rare vascular diseases of medium-sized arteries and in the study of the pathophysiology of arterial aging in general.
引用
- Poli, Fortier, Khettab, et al. Validation and Feasibility of an Automated System for the Assessment of Vascular Structure and Mechanical Properties in the Digital Arteries: An Ultrahigh-Frequency Ultrasound Study. Ultrasound Med Biol. 2022. PMID: 35058069. doi:10.1016/j.ultrasmedbio.2021.12.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。