低剂量胸部CT中深度学习结节检测算法的诊断验证:虚拟筛查场景下优化剂量阈值的确定
研究用68个拟人胸部体模携带329个结节(150个磨玻璃、179个实性,4种大小),以9种管电压/电流组合扫描,评估基于深度学习的计算机辅助诊断系统检测肺结节的表现。最低有效剂量且检测率未受损的组合为80 kV/50 mAs(敏感度97.9%,平均假阳性率1.9,平均CTDIvol 1.2±0.4 mGy,平均有效剂量0.66 mSv)。管电压和电流对检出率影响最大,结节大小和密度无显著影响。
要点
- 80 kV/50 mAs组合下深度学习CAD检测肺结节敏感度97.9%,平均假阳性率1.9。
- 管电压和管电流对深度学习CAD检出率影响最大,结节大小和密度无显著影响。
- 最低有效剂量且检测率未受损的组合为80 kV/50 mAs(平均有效剂量0.66 mSv)。
核心结论
- 本研究提出的最佳管电压/电流组合80 kV/50 mAs与类似研究(多使用传统CAD软件)提出的组合相当;修改管电压和管电流对深度学习CAD软件肺结节检测性能有显著影响,与结节大小和成分无关。
研究设计与样本
- 体模研究:68个拟人胸部体模携带329个结节(150个磨玻璃、179个实性),结节大小4种(5 mm、8 mm、10 mm、12 mm),以9种管电压/电流组合扫描,由市售基于深度学习的CAD系统分析,结果用比例比较,并行逻辑回归评估管电压、管电流、结节大小、密度和位置的影响。
诊断标准与关键数值
- 最低有效剂量且检测率未受损的组合为80 kV/50 mAs,敏感度97.9%,平均假阳性率1.9,平均CTDIvol 1.2±0.4 mGy,平均有效剂量0.66 mSv。
- 逻辑回归显示管电压和管电流对检出率影响最大,结节大小和密度无显著影响。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study was conducted to evaluate the effect of dose reduction on the performance of a deep learning (DL)-based computer-aided diagnosis (CAD) system regarding pulmonary nodule detection in a virtual screening scenario.
METHODS: Sixty-eight anthropomorphic chest phantoms were equipped with 329 nodules (150 ground glass, 179 solid) with four sizes (5 mm, 8 mm, 10 mm, 12 mm) and scanned with nine tube voltage/current combinations. The examinations were analyzed by a commercially available DL-based CAD system. The results were compared by a comparison of proportions. Logistic regression was performed to evaluate the impact of tube voltage, tube current, nodule size, nodule density, and nodule location.
RESULTS: The combination with the lowest effective dose (E) and unimpaired detection rate was 80 kV/50 mAs (sensitivity: 97.9%, mean false-positive rate (FPR): 1.9, mean CTDIvol: 1.2 ± 0.4 mGy, mean E: 0.66 mSv). Logistic regression revealed that tube voltage and current had the greatest impact on the detection rate, while nodule size and density had no significant influence.
CONCLUSIONS: The optimal tube voltage/current combination proposed in this study (80 kV/50 mAs) is comparable to the proposed combinations in similar studies, which mostly dealt with conventional CAD software. Modification of tube voltage and tube current has a significant impact on the performance of DL-based CAD software in pulmonary nodule detection regardless of their size and composition.
KEY POINTS: • Modification of tube voltage and tube current has a significant impact on the performance of deep learning-based CAD software. • Nodule size and composition have no significant impact on the software's performance. • The optimal tube voltage/current combination for the examined software is 80 kV/50 mAs.
引用
- Peters, Huber, Obmann, et al. Diagnostic validation of a deep learning nodule detection algorithm in low-dose chest CT: determination of optimized dose thresholds in a virtual screening scenario. Eur Radiol. 2022. PMID: 35059804. doi:10.1007/s00330-021-08511-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。