文献速递神经介入外周血管DSA

低剂量三维旋转血管造影评估颅内动脉瘤:图像质量与辐射剂量分析

Low-Dose Three-Dimensional Rotational Angiography for Evaluating Intracranial Aneurysms: Analysis of Image Quality and Radiation Dose.
PubMed 文献 · Korean J Radiol · 2022-01-04

回顾性分析146例患者共146次三维旋转血管造影,比较常规方法与低剂量方法(5秒、0.10微戈瑞/帧)的图像质量和辐射剂量。低剂量组图像质量总分及四项标准单项评分均不劣于常规组,剂量面积乘积和空气比释动能显著降低。结论认为低剂量三维旋转血管造影在评估颅内动脉瘤时图像质量不劣于常规方法,辐射剂量降低70%–77%。

要点

  • 低剂量三维旋转血管造影图像质量不劣于常规方法,可用于颅内动脉瘤评估。
  • 低剂量方案每旋转剂量面积乘积1.32 Gy·cm²,常规方案5.87 Gy·cm²。
  • 低剂量方案每旋转空气比释动能0.17 Gy,常规方案0.56 Gy。

核心结论

  • 低剂量三维旋转血管造影在评估颅内动脉瘤时图像质量不劣于常规三维旋转血管造影,且辐射剂量降低70%–77%。

研究设计与样本

  • 回顾性评估146例前瞻性登记患者(男46例、女100例;中位年龄58岁,范围19–81岁)的临床数据与三维旋转血管造影数据集;常规方法79次检查,低剂量方法(5秒、0.10微戈瑞/帧)67次检查。

诊断标准与关键数值

  • 由两名神经介入医师采用3分制对四项评估标准进行主观图像质量评分,总分取四项评分的平均值;以-0.2为非劣效界值(低剂量组评分减常规组评分)进行非劣效性统计检验。

临床应用与阅片要点

  • 低剂量三维旋转血管造影可用于颅内动脉瘤评估,图像质量不劣于常规方法且辐射剂量更低。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This study aimed to evaluate the image quality and dose reduction of low-dose three-dimensional (3D) rotational angiography (RA) for evaluating intracranial aneurysms.

MATERIALS AND METHODS: We retrospectively evaluated the clinical data and 3D RA datasets obtained from 146 prospectively registered patients (male:female, 46:100; median age, 58 years; range, 19-81 years). The subjective image quality of 79 examinations obtained from a conventional method and 67 examinations obtained from a low-dose (5-seconds and 0.10-µGy/frame) method was assessed by two neurointerventionists using a 3-point scale for four evaluation criteria. The total image quality score was then obtained as the average of the four scores. The image quality scores were compared between the two methods using a noninferiority statistical testing, with a margin of -0.2 (i.e., score of low-dose group - score of conventional group). For the evaluation of dose reduction, dose-area product (DAP) and air kerma (AK) were analyzed and compared between the two groups.

RESULTS: The mean total image quality score ± standard deviation of the 3D RA was 2.97 ± 0.17 by reader 1 and 2.95 ± 0.20 by reader 2 for conventional group and 2.92 ± 0.30 and 2.95 ± 0.22, respectively, for low-dose group. The image quality of the 3D RA in the low-dose group was not inferior to that of the conventional group according to the total image quality score as well as individual scores for the four criteria in both readers. The mean DAP and AK per rotation were 5.87 Gy-cm² and 0.56 Gy, respectively, in the conventional group, and 1.32 Gy-cm² (p < 0.001) and 0.17 Gy (p < 0.001), respectively, in the low-dose group.

CONCLUSION: Low-dose 3D RA was not inferior in image quality and reduced the radiation dose by 70%-77% compared to the conventional 3D RA in evaluating intracranial aneurysms.

来源:Korean J Radiol,查看原文。版权归原作者所有。

引用

  1. Ki, Kim, Kim, et al. Low-Dose Three-Dimensional Rotational Angiography for Evaluating Intracranial Aneurysms: Analysis of Image Quality and Radiation Dose. Korean J Radiol. 2022. PMID: 35029071. doi:10.3348/kjr.2021.0162

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。