超高场7T磁共振成像中基础与高级金属伪影减少技术的体模研究:可行性与有效性

Basic and Advanced Metal-Artifact Reduction Techniques at Ultra-High Field 7-T Magnetic Resonance Imaging-Phantom Study Investigating Feasibility and Efficacy.
PubMed 文献 · Invest Radiol · 2022-01-13

本研究在7T磁共振上对钛合金椎弓根螺钉和椎间融合器体模进行多种序列采集,评估基础(增加接收带宽)与高级(视角倾斜VAT、层编码金属伪影校正SEMAC)技术减少金属伪影的可行性与有效性。结果显示所有技术均减少种植体周围信号空洞和几何畸变,以oBW-VAT和SEMAC最为有效,SEMAC可显著减少圆形结构畸变,但增加SEMAC层编码步数未进一步改善伪影减少。

要点

  • 在7T磁共振上,增加接收带宽联合视角倾斜(oBW-VAT)与SEMAC均可显著减少种植体周围信号空洞和几何畸变。
  • 仅SEMAC能显著减少种植体周围框架圆形结构的畸变。
  • 增加SEMAC序列的层编码步数并未显著进一步减少金属伪影。

核心结论

  • 在7T磁共振上使用基础和高级金属伪影减少方法是可行且有效的;增加接收带宽联合VAT以及SEMAC均能显著减少金属植入物周围的信号空洞和几何畸变。

研究设计与样本

  • 实验性体模研究,对由椎弓根螺钉(体模1)和椎间融合器(体模2)组成的钛合金体模行7T磁共振涡轮自旋回波序列采集,序列包括非优化标准序列、增加接收带宽序列(oBW)、VAT、oBW-VAT联合及SEMAC;由两名经过专科培训的肌骨放射科医师独立评估。

关键影像表现/征象

  • 所有金属伪影减少技术均减少种植体周围信号空洞并减轻几何畸变,其中oBW-VAT和SEMAC最为有效。
  • 与未优化序列相比,oBW-VAT和SEMAC产生的种植体周围信号空洞显著更小(所有P≤0.008),畸变角度显著更小(P≤0.001)。
  • 仅SEMAC能显著减少种植体周围框架圆形形状的畸变(P≤0.006)。
  • 增加SEMAC序列的层编码步数并未导致显著更好的金属伪影减少(所有P≥0.257)。

诊断标准与关键数值

  • 评估指标包括种植体周围信号空洞和几何畸变(角度测量及圆形形状丢失 presence)。
  • 统计采用Friedman检验和Cochran Q检验,经Bonferroni校正进行多重比较,P<0.05视为有统计学意义。

临床应用与阅片要点

  • 在7T磁共振检查中,对于金属植入物附近区域,可采用增加接收带宽联合VAT或SEMAC序列以减少信号空洞和几何畸变。
  • SEMAC对圆形结构畸变的减少有额外优势,但增加层编码步数可能不会进一步改善伪影减少效果。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to demonstrate the feasibility and efficacy of basic (increased receive bandwidth) and advanced (view-angle tilting [VAT] and slice-encoding for metal artifact correction [SEMAC]) techniques for metal-artifact reduction in ultra-high field 7-T magnetic resonance imaging (MRI).

MATERIALS AND METHODS: In this experimental study, we performed 7-T MRI of titanium alloy phantom models composed of a spinal pedicle screw (phantom 1) and an intervertebral cage (phantom 2) centered in a rectangular LEGO frame, embedded in deionized-water-gadolinium (0.1 mmol/L) solution. The following turbo spin-echo sequences were acquired: (1) nonoptimized standard sequence; (2) optimized, that is, increased receive bandwidth sequence (oBW); (3) VAT; (4) combination of oBW and VAT (oBW-VAT); and (5) SEMAC. Two fellowship-trained musculoskeletal radiologists independently evaluated images regarding peri-implant signal void and geometric distortion (a, angle measurement and b, presence of circular shape loss). Statistics included Friedman test and Cochran Q test with Bonferroni correction for multiple comparisons. P values <0.05 were considered to represent statistical significance.

RESULTS: All metal-artifact reduction techniques reduced peri-implant signal voids and diminished geometric distortions, with oBW-VAT and SEMAC being most efficient. Compared with nonoptimized sequences, oBW-VAT and SEMAC produced significantly smaller peri-implant signal voids (all P ≤ 0.008) and significantly smaller distortion angles (P ≤ 0.001). Only SEMAC could significantly reduce distortions of circular shapes in the peri-implant frame (P ≤ 0.006). Notably, increasing the number of slice-encoding steps in SEMAC sequences did not lead to a significantly better metal-artifact reduction (all P ≥ 0.257).

CONCLUSIONS: The use of basic and advanced methods for metal-artifact reduction at 7-T MRI is feasible and effective. Both a combination of increased receive bandwidth and VAT as well as SEMAC significantly reduce the peri-implant signal void and geometric distortion around metal implants.

来源:Invest Radiol,查看原文。版权归原作者所有。