光子计数探测器CT的血管钙化去除算法:基于心脏运动体模的评估

Photon-Counting Detector CT-Based Vascular Calcium Removal Algorithm: Assessment Using a Cardiac Motion Phantom.
PubMed 文献 · Invest Radiol · 2022-01-13

本体模研究评估第一代双源光子计数探测器CT(PCD-CT)上一项基于光谱数据的钙化去除重建算法(PureLumen)。在模拟50-100次/分心脏运动的胸部体模中,含25%和50%钙化狭窄的血管体模显示,PureLumen较单纯虚拟单能图像(Mono)使中位狭窄率更接近真实值(50%病变:50.0% vs 56.0%;25%病变:31.5% vs 36.0%,P<0.05)。结论认为PureLumen能有效减少钙化斑块所致晕状伪影并提高图像可判读性,且在模拟心率达80次/分时保持性能;尚需在体临床研究确认。

要点

  • PureLumen在50%和25%钙化狭窄病变中测得的中位狭窄率均较Mono更接近真实值。
  • 在模拟心率达80次/分时,PureLumen图像质量视觉评估为良好。
  • 该算法基于光谱基物质分解去除钙贡献,旨在保留其他物质不变。

核心结论

  • 本体模研究显示,基于第一代双源PCD-CT的新型钙化去除重建算法(PureLumen)可有效减少重度钙化斑块引起的晕状伪影并提高图像可判读性,且在模拟心率达80次/分时保持性能;尚需在体临床研究确认该重建在冠状动脉CT血管造影质量和准确性方面的获益。

研究设计与样本

  • 体模研究,使用第一代双源光子计数探测器CT系统(NAEOTOM Alpha,Siemens Healthineers),采用前瞻性心电触发双源序列模式采集。

关键影像表现/征象

  • 两个碘对比剂充盈的血管体模(直径4 mm)分别含25%和50%面积直径狭窄(PDS)的钙化病变,置于模拟心脏运动(50-100次/分)的胸部体模中。
  • 图像在RR间期80%处重建,分别采用65 keV的虚拟单能图像(Mono)和附加钙化去除的PureLumen。
  • PureLumen基于光谱基物质分解为碘和钙,旨在重建无钙贡献的图像并保持其他物质贡献不变。
  • 两名读者独立评估所有重建图像序列中各血管插入物的狭窄程度。

诊断标准与关键数值

  • 50%病变:Mono中位PDS为56.0%(52.0%,57.0%),PureLumen中位PDS为50.0%(48.5%,51.0%)。
  • 25%病变:Mono中位PDS为36.0%(29.5%,39.5%),PureLumen中位PDS为31.5%(30.5%,34.0%)。
  • 两种病变大小的Mono与PureLumen结果差异均有统计学意义(P < 0.05),PureLumen中位值更接近50%和25%病变的真实狭窄尺寸。
  • 依据心率的图像质量视觉评估显示,PureLumen在心率达80次/分时图像质量良好。

临床应用与阅片要点

  • PureLumen可减少重度钙化斑块所致晕状伪影,可能有助于提高冠状动脉CT血管造影的图像可判读性。
  • 在模拟心率达80次/分时该算法仍保持性能,但更高心率下的表现需结合原始数据判断。
  • 本研究为体模研究,作者指出尚需在体临床研究确认该重建在冠状动脉CT血管造影质量和准确性方面的获益。

局限性

  • 原文结论指出尚需在体临床研究确认该重建在冠状动脉CT血管造影质量和准确性方面的获益。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The diagnostic performance of coronary computed tomography angiography is known to be negatively affected by the presence of severely calcified plaques in the coronary arteries. In this article, the performance of a novel image reconstruction algorithm (PureLumen) based on spectral CT data of a first-generation dual-source photon-counting detector computed tomography (PCD-CT) system was assessed in a phantom study. PureLumen tries to remove only the calcified contributions from the image while leaving the rest unmodified.

MATERIALS AND METHODS: The study uses 2 iodine contrast filled vessel phantoms (diameter 4 mm) filled with different concentrations of iodine and equipped with calcified stenosis inserts. Each phantom features 2 separate calcified lesions of 25% and 50% percentage diameter stenosis (PDS) size. The vessel phantoms were mounted inside an anthropomorphic thorax phantom attached to an artificial motion device, simulating realistic cardiac motion at heart rates between 50 beats per minute and 100 beats per minute. Acquisitions were performed using a prospectively electrocardiogram triggered dual-source sequence mode on a PCD-CT system (NAEOTOM Alpha, Siemens Healthineers). Images were reconstructed at 80% of the RR interval with virtual monoenergetic images (Mono) and with additional calcium-removal (PureLumen), both at 65 keV. PureLumen is based on a spectral base material decomposition into iodine and calcium, which aims to reconstruct images without calcium contributions, while leaving all other material contribution unchanged. Stenosis grade was assessed individually for each vessel insert in all reconstructed image series by 2 readers.

RESULTS: The measured median PDS values for the 50% lesion were 56.0% (52.0%, 57.0%) for the Mono case and 50.0% (48.5%, 51.0%) for PureLumen. The 25% lesion median PDS values were 36.0% (29.5%, 39.5%) for Mono and 31.5% (30.5%, 34.0%) for PureLumen. Both lesion sizes demonstrate a significant difference between Mono and PureLumen in their result (P < 0.05) with PureLumen median values being closer to the actual true stenosis size for the 50% and 25% lesion. A visual assessment of the image quality depending on the heart rate yielded good image quality up to a heart rate of 80 beats per minute in the PureLumen case.

CONCLUSIONS: This phantom study shows that a novel calcium-removal image reconstruction algorithm (PureLumen) using a first-generation dual-source PCD-CT effectively decreases blooming artifacts caused by heavily calcified plaques and improves image interpretability. It also shows that PureLumen retains its performance in the presence of motion with simulated heart rates up to 80 beats per minute. Future in vivo clinical studies are needed to confirm the benefits of this type of reconstruction in terms of coronary computed tomography angiography quality and accuracy.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Allmendinger, Nowak, Flohr, et al. Photon-Counting Detector CT-Based Vascular Calcium Removal Algorithm: Assessment Using a Cardiac Motion Phantom. Invest Radiol. 2022. PMID: 35025834. doi:10.1097/RLI.0000000000000853

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。