低剂量全视野光子计数CT改善冠状动脉钙化检测与定量:一项体模研究
该体模研究系统评估了光谱光子计数CT(SPCCT)与常规CT检测和定量冠状动脉钙化(CAC)的能力,并探索降低辐射剂量的可能性。结果显示SPCCT检测CAC优于CT,且在50%辐射剂量降低时仍保持优势;SPCCT能更准确估计物理体积,减少层厚重建可提高物理质量近似精度。
要点
- SPCCT检测CAC优于常规CT,小体模多检出33%,大体模多检出7%。
- 辐射剂量降低50%并减小层厚时,小体模CAC检出率CT增加108%、SPCCT增加150%。
- 常规方案CAC定量与物理体积偏差大(达134%),SPCCT在低剂量薄层下体积高估降至72%。
核心结论
- SPCCT检测CAC优于常规CT,且在降低辐射剂量时仍保持优势;SPCCT能改善物理体积估计;两种CT采用薄层重建均能提高物理质量近似精度。
研究设计与样本
- 体模研究,使用含CAC的圆柱形插入物置于拟人胸部体模内,采用常规临床CAC CT协议采集和重建数据,并额外以降低管电流实现50%辐射剂量降低、减小层厚采集数据;以已知物理量作为参考标准。
诊断标准与关键数值
- 钙化可检测定义为连续三个相邻体素超过CAC评分阈值130 HU;CAC定量以体积评分和质量评分与已知物理量比较。
局限性
- 大体模尺寸下噪声水平干扰CAC检测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of the current study was to systematically assess coronary artery calcium (CAC) detection and quantification for spectral photon-counting CT (SPCCT) in comparison to conventional CT and, in addition, to evaluate the possibility of radiation dose reduction.
METHODS: Routine clinical CAC CT protocols were used for data acquisition and reconstruction of two CAC containing cylindrical inserts which were positioned within an anthropomorphic thorax phantom. In addition, data was acquired at 50% lower radiation dose by reducing tube current, and slice thickness was decreased. Calcifications were considered detectable when three adjacent voxels exceeded the CAC scoring threshold of 130 Hounsfield units (HU). Quantification of CAC (as volume and mass score) was assessed by comparison with known physical quantities.
RESULTS: In comparison with CT, SPCCT detected 33% and 7% more calcifications for the small and large phantoms, respectively. At reduced radiation dose and reduced slice thickness, small phantom CAC detection increased by 108% and 150% for CT and SPCCT, respectively. For the large phantom size, noise levels interfered with CAC detection. Although comparable between CT and SPCCT, routine protocols CAC quantification showed large deviations (up to 134%) from physical CAC volume. At reduced radiation dose and slice thickness, physical volume overestimations decreased to 96% and 72% for CT and SPCCT, respectively. In comparison with volume scores, mass score deviations from physical quantities were smaller.
CONCLUSION: CAC detection on SPCCT is superior to CT, and was even preserved at a reduced radiation dose. Furthermore, SPCCT allows for improved physical volume estimation.
KEY POINTS: • In comparison with conventional CT, increased coronary artery calcium detection (up to 156%) for spectral photon-counting CT was found, even at 50% radiation dose reduction. • Spectral photon-counting CT can more accurately measure physical volumes than conventional CT, especially at reduced slice thickness and for high-density coronary artery calcium. • For both conventional and spectral photon-counting CT, reduced slice thickness reconstructions result in more accurate physical mass approximation.