文献速递腹部技术与安全CT

腹部CT深度学习重建的临床接受度:客观与主观图像质量及低对比度可检测性评估

Clinical acceptance of deep learning reconstruction for abdominal CT imaging: objective and subjective image quality and low-contrast detectability assessment.
PubMed 文献 · Eur Radiol · 2022-01-06

该研究用两种体模在九个剂量水平下行腹部CT采集,比较深度学习重建(DLR)与两类迭代重建(IR)及滤波反投影的图像质量与临床接受度。DLR在低剂量下噪声更低、低对比度可检测性指数更高,且放射科医生对所有评估标准的DLR图像偏好显著(p<0.0001),重建速度较MBIR快12倍。

要点

  • DLR在低剂量(CTDIvol≤2.2/≤4.5 mGy)下噪声低于MBIR,低对比度可检测性指数高于MBIR。
  • DLR图像纹理比MBIR更细腻,更接近HIR。
  • 10名资深放射科医生对所有评估标准均显著偏好DLR图像(p<0.0001)。

核心结论

  • DLR在较低剂量水平下客观低对比度检测与噪声表现改善,且在所评估的重建算法中临床接受度最佳。

研究设计与样本

  • 研究论文,使用两种体模和共九个剂量水平的CT采集,重建采用两种迭代重建算法、DLR及滤波反投影,比较空间分辨率、图像纹理、平均噪声值及客观与主观低对比度可检测性;10名资深放射科医生对10例分别用DLR和IR算法重建的CT检查进行临床接受度评分。

关键影像表现/征象

  • DLR图像纹理比MBIR更细腻,更接近HIR的纹理。
  • 与MBIR相比,DLR在低剂量(CTDIvol≤2.2 mGy)下噪声更低,在低剂量(CTDIvol≤4.5 mGy)下低对比度可检测性指数更高。
  • 在较高剂量下,MBIR的空间分辨率和可检测性优于DLR。
  • 与HIR相比,DLR空间分辨率更高、噪声更低、可检测性指数更高。

诊断标准与关键数值

  • 与MBIR相比,DLR在CTDIvol≤2.2 mGy时噪声更低。
  • 与MBIR相比,DLR在CTDIvol≤4.5 mGy时低对比度可检测性指数更高。
  • DLR的主观低对比度表现与其它算法间未发现显著差异(p≥0.005)。
  • 放射科医生对所有评估标准均更偏好DLR图像(p<0.0001);MBIR在所有评估标准中评分均差于HIR(p<0.0001),噪声项除外(p=0.044)。
  • DLR重建速度较MBIR快12倍。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the image quality and clinical acceptance of a deep learning reconstruction (DLR) algorithm compared to traditional iterative reconstruction (IR) algorithms.

METHODS: CT acquisitions were performed with two phantoms and a total of nine dose levels. Images were reconstructed with two types of IR algorithms, DLR and filtered-back projection. Spatial resolution, image texture, mean noise value, and objective and subjective low-contrast detectability were compared. Ten senior radiologists evaluated the clinical acceptance of these algorithms by scoring ten CT exams reconstructed with the DLR and IR algorithms evaluated.

RESULTS: Compared to MBIR, DLR yielded a lower noise and a higher low-contrast detectability index at low doses (CTDIvol ≤ 2.2 and ≤ 4.5 mGy, respectively). Spatial resolution and detectability at higher doses were better with MBIR. Compared to HIR, DLR yielded a higher spatial resolution, a lower noise, and a higher detectability index. Despite these differences in algorithm performance, significant differences in subjective low-contrast performance were not found (p ≥ 0.005). DLR texture was finer than that of MBIR and closer to that of HIR. Radiologists preferred DLR images for all criteria assessed (p < 0.0001), whereas MBIR was rated worse than HIR (p < 0.0001) in all criteria evaluated, except for noise (p = 0.044). DLR reconstruction time was 12 times faster than that of MBIR.

CONCLUSION: DLR yielded a gain in objective detection and noise at lower dose levels with the best clinical acceptance among the evaluated reconstruction algorithms.

KEY POINTS: • DLR yielded improved objective low-contrast detection and noise at lower dose levels. • Despite the differences in objective detectability among the algorithms evaluated, there were no differences in subjective detectability. • DLR presented significantly higher clinical acceptability scores compared to MBIR and HIR.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Bornet, Villani, Gillet, et al. Clinical acceptance of deep learning reconstruction for abdominal CT imaging: objective and subjective image quality and low-contrast detectability assessment. Eur Radiol. 2022. PMID: 34989850. doi:10.1007/s00330-021-08410-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。