记忆障碍患者自动脑容量测量:常规与超快速3D T1加权MRI序列使用两种软件包的比较

Automated Brain Volumetry in Patients With Memory Impairment: Comparison of Conventional and Ultrafast 3D T1-Weighted MRI Sequences Using Two Software Packages.
PubMed 文献 · AJR Am J Roentgenol · 2022-01-05

该回顾性研究纳入36例记忆障碍患者,比较常规3D T1加权成像与超快速3D EPI T1加权序列的自动脑容量测量,使用NeuroQuant和FreeSurfer软件分析。结果显示多数脑区两种序列间一致性高至几乎完美,但多数区域存在显著平均差异和系统偏倚,Bland-Altman分析显示宽的一致性界限。结论认为两种序列脑容量测量存在显著差异和系统偏倚,超快速序列可能用于因严重运动伪影无法使用常规序列的患者。

要点

  • 常规与超快速3D T1序列的脑容量测量在多数脑区一致性高至几乎完美。
  • 多数脑区两种序列间存在显著平均差异和系统偏倚,一致性界限宽。
  • 超快速序列可能用于因严重运动伪影无法使用常规序列的患者。

核心结论

  • 常规与超快速3D T1加权序列的脑容量测量存在显著差异和系统偏倚。

研究设计与样本

  • 回顾性研究纳入36例记忆障碍患者(25女11男,平均年龄68.4岁),行3T脑MRI,包括常规3D T1加权反转恢复梯度回波序列(层厚1.0 mm,采集时间3分4秒)和针对运动患者的各向同性超快速3D EPI T1加权序列(层厚1.2 mm,采集时间30秒);排除5例因严重运动伪影致常规序列质量不足者;使用NeuroQuant 3.0和FreeSurfer 7.1.1软件自动测量每侧半球9个脑区体积。

诊断标准与关键数值

  • 使用NeuroQuant,左侧半球超快速序列在4个区域体积显著大于常规序列,常规序列在3个区域体积显著大于超快速序列。
  • 使用NeuroQuant,双侧皮质灰质平均偏倚(超快速减常规)和95% LOA最大:左侧-50.61 cm3(-56.27 cm3,-44.94 cm3),右侧-50.02 cm3(-54.88 cm3,-45.16 cm3)。
  • 在16例阿尔茨海默病患者和20例无阿尔茨海默病患者亚组分析中均观察到两序列间变异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Isotropic 3D T1-weighted imaging has long acquisition times, potentially leading to motion artifact and altered brain volume measurements. Acquisition times may be greatly shortened using an isotropic ultrafast 3D echo-planar imaging (EPI) T1-weighted sequence. OBJECTIVE. The purpose of this article was to compare automated brain volume measurements between conventional 3D T1-weighted imaging and ultrafast 3D EPI T1-weighted imaging. METHODS. This retrospective study included 36 patients (25 women, 11 men; mean age, 68.4 years) with memory impairment who underwent 3-T brain MRI. Examinations included both conventional 3D T1-weighted imaging using inversion recovery gradient-recalled echo sequence (section thickness, 1.0 mm; acquisition time, 3 minutes 4 seconds) and, in patients exhibiting motion, an isotropic ultrafast 3D EPI T1-weighted sequence (section thickness, 1.2 mm; acquisition time, 30 seconds). The 36-patient sample excluded five patients in whom severe motion artifact rendered the conventional sequence of insufficient quality for volume measurements. Automated brain volumetry was performed using NeuroQuant (version 3.0, CorTechs Laboratories) and FreeSurfer (version 7.1.1, Harvard University) software. Volume measurements were compared between sequences for nine regions in each hemisphere. RESULTS. Volumes showed substantial to almost perfect agreement between the two sequences for most regions bilaterally. However, most regions showed significant mean differences between sequences, and Bland-Altman analyses showed consistent systematic biases and wide limits of agreement (LOA). For example, for the left hemisphere using NeuroQuant, volume was significantly greater for the ultrafast sequence in four regions and significantly greater for the conventional sequence in three regions, whereas standardized effect size between sequences was moderate for four regions and large for one region. Using NeuroQuant, mean bias (ultrafast minus conventional) and 95% LOA were greatest in cortical gray matter bilaterally (-50.61 cm3 [-56.27 cm3, -44.94 cm3] for the left hemisphere; -50.02 cm3 [-54.88 cm3, -45.16 cm3] for the right hemisphere). The variation between the two sequences was observed in subset analyses of 16 patients with and 20 patients without Alzheimer disease. CONCLUSION. Brain volume measurements show significant differences and systematic biases between the conventional and ultrafast sequences. CLINICAL IMPACT. In patients in whom severe motion artifact precludes use of the conventional sequence, the ultrafast sequence may be useful to enable brain volume-try. However, the current conventional 3D T1-weighted sequence remains preferred in patients who can tolerate the standard examination.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Heo, Baek, Skare, et al. Automated Brain Volumetry in Patients With Memory Impairment: Comparison of Conventional and Ultrafast 3D T1-Weighted MRI Sequences Using Two Software Packages. AJR Am J Roentgenol. 2022. PMID: 34985311. doi:10.2214/AJR.21.27043