三维与二维磁共振弹性成像技术评估肝脏硬度的同一患者内比较
回顾性评估57例肝脏磁共振弹性成像检查,比较三维自旋回波回波平面成像(seEPI)与二维梯度回波(GRE)及二维seEPI序列。三维磁共振弹性成像对伪影敏感性更低,可提供更大肝实质面积,基于全部层面的硬度值与基于4个层面的简略值高度相关(r=0.9;p<0.001)且无显著差异(p=0.927)。
要点
- 三维磁共振弹性成像采用seEPI序列可减少伪影影响,测量可重复性更高。
- 三维磁共振弹性成像感兴趣区面积(50,723 mm²)明显大于二维GRE(5,814 mm²)和二维seEPI(10,642 mm²)。
- 三维磁共振弹性成像基于4个层面的简略硬度值与基于全部层面的硬度值高度相关(r=0.9;p<0.001)且无显著差异(p=0.927)。
核心结论
- 三维磁共振弹性成像因对伪影敏感性更低且可提供更大肝实质面积,能实现更可重复的测量,并支持更全面的肝脏评估。
研究设计与样本
- 回顾性评估57例肝脏磁共振弹性成像检查,包括二维GRE、二维seEPI和三维seEPI序列,由2名放射科医生分析弹性图并绘制多边形感兴趣区。
诊断标准与关键数值
- 3D MRE、简略3D MRE、2D GRE和2D seEPI的感兴趣区平均表面积分别为50,723 mm²、12,669 mm²、5,814 mm²和10,642 mm²。
- 三维磁共振弹性成像基于精选和非精选感兴趣区的硬度值无临床显著差异,平均差异小于0.1 kPa。
- 简略三维硬度值与基于全部层面的三维硬度值高度相关(r=0.9;p<0.001)且无显著差异(p=0.927)。
临床应用与阅片要点
- 三维磁共振弹性成像采用seEPI序列可降低伪影影响,在绘制感兴趣区时可获得更大的肝实质面积。
- 采用基于4个层面的简略三维硬度值测量与基于全部层面测量结果高度一致(r=0.9;p<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate performance of 3D magnetic resonance elastography (MRE) using spin-echo echo-planar imaging (seEPI) for assessment of hepatic stiffness compared with 2D gradient-recalled echo (GRE) and 2D seEPI sequences.
METHODS: Fifty-seven liver MRE examinations including 2D GRE, 2D seEPI, and 3D seEPI sequences were retrospectively evaluated. Elastograms were analyzed by 2 radiologists and polygonal regions of interests (ROIs) were drawn in 2 different fashions: "curated" ROI (avoiding liver edge, major vessels, and areas of wave interferences) and "non-curated" ROI (including largest cross section of liver, to assess the contribution of artifacts). Liver stiffness measurement (LSM) was calculated as the arithmetic mean of individual stiffness values for each technique. For 3D MRE, LSMs were also calculated based on 4 slices ("abbreviated LSM"). Intra-patient variations in LSMs and different methods of ROI placement were assessed by univariate tests. A p-value of < 0.05 was set as a statistically significant difference.
RESULTS: Mean surface areas of the ROIs were 50,723 mm2, 12,669 mm2, 5814 mm2, and 10,642 mm2 for 3D MRE, abbreviated 3D MRE, 2D GRE, and 2D seEPI, respectively. 3D LSMs based on curated and non-curated ROIs showed no clinically significant difference, with a mean difference less than 0.1 kPa. Abbreviated 3D LSMs had excellent correlation with 3D LSMs based on all slices (r = 0.9; p < 0.001) and were not significantly different (p = 0.927).
CONCLUSION: 3D MRE allows more reproducible measurements due to its lower susceptibility to artifacts and provides larger areas of parenchyma, enabling a more comprehensive evaluation of the liver.