可视化微聚焦超声:安全性的共识与能量设备综述
本文汇集七位全球专家对MFU-V及三种其他高强度聚焦超声设备独立测试数据的综述,指出MFU-V具备实时可视化、耦合检测及均匀热凝固点等特性,并形成关于可预防并发症类型及防治技术的专家共识。
要点
- MFU-V具备实时可视化及耦合检测能力,其他测试设备均不具备该特性。
- MFU-V热凝固点大小一致且均匀,焦点能量定位精确,使用中热调节可靠。
- 专家共识确立了超声能量设备相关可预防并发症的类型及防治技术。
核心结论
- 独立测试结果与专家共识结论强烈支持:实时可视化和耦合检测能力为MFU-V独有,有助于预防并发症并提升能量设备的安全性与有效性。
研究设计与样本
- 七位全球专家组成的共识小组,基于对MFU-V及三种其他超声设备的软件、热特性、换能器声学和超声治疗的独立评估数据进行综述并形成共识。
诊断标准与关键数值
- MFU-V具备可视化能力,其他设备缺乏该特性。
- MFU-V能保留患者病史,热凝固点大小一致且均匀,焦点能量定位精确,使用中热调节可靠。
- MFU-V具有均匀的热凝固点、严格的热调节、大焦点增益和短波束长度,这些特性未在任何HIFU设备中同时出现。
局限性
- 其他能量设备在美国以外 marketed 用于类似适应证,但支持这些适应证的已发表研究有限。
- 其他设备均不提供可视化或验证其是否到达表浅肌肉腱膜系统的能力。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Microfocused ultrasound with visualization (MFU-V; Ultherapy® , Merz North America) is US Food and Drug Administration-cleared as a non-invasive procedure that lifts the soft tissue of the neck, submentum and brow, and improves lines and wrinkles on the upper chest. Several other energy-based devices are in use in countries outside the USA where they are marketed for indications similar to those of MFU-V, although published studies supporting these indications are limited and none of the other devices provides visualization or verification they reach the superficial musculoaponeurotic system.
METHODS: Due to the evolving landscape of ultrasound technology as more devices enter the market, seven global thought leaders who are qualified experts on the use of various high-intensity focused ultrasound (HIFU)/MFU-V technologies convened to review data from an independent evaluation of the software, thermal characteristics, transducer acoustics and ultrasound therapy of MFU-V and three other ultrasound-based devices.
RESULTS: The independent testing demonstrated the devices have key differences in several parameters that play a role in safety and effectiveness. Specifically, MFU-V has visualization capability but the other devices lack that feature. Other differences include the retention of patient history, consistent size and uniformity of thermal coagulation points (TCPs), precise localization of energy concentration at the focal point, and reliable thermal regulation during use. The expert panel established a consensus on the types of preventable complications associated with ultrasound-based energy devices and techniques for preventing and treating complications.
CONCLUSIONS: The independent test results of MFU-V/HIFU devices and the consensus panel conclusions provide strong support that real-time visualization and the capability to detect coupling, features found only in MFU-V, help prevent complications and enhance the safety and effectiveness of energy-based devices. The independent evaluation also revealed that MFU-V has several additional features that play key roles in safety and clinical effectiveness, including uniformity of TCPs, tight thermal regulation, large focal gain, and short beam length, that were not found collectively in any of the HIFU devices.