深度学习重建用于超低剂量肺部CT:体模研究中人工磨玻璃结节的体积测量准确性与可重复性
研究在超低剂量(6.3 mAs)320排CT下,用体模中6 mm和10 mm、-660 HU的人工磨玻璃结节比较深度学习重建(DLR)、基于模型的迭代重建(MBIR)与混合迭代重建(HIR)的体积测量准确性。DLR的绝对体积误差百分比(APEvol)低于MBIR和HIR,10 mm中肺野结节DLR为1.36%,显著低于HIR的8.01%(p=0.0022)和MBIR的7.30%(p=0.0053);DLR在观察者间和观察者内一致性上也显示更窄的一致性界限。作者结论为超低剂量CT下DLR对人工磨玻璃结节体积测量的准确性高于MBIR和HIR。
要点
- 10 mm中肺野磨玻璃结节超低剂量CT体积测量APEvol:DLR 1.36%,显著低于HIR 8.01%(p=0.0022)与MBIR 7.30%(p=0.0053)。
- DLR在观察者间及观察者内体积测量一致性上的一致性界限均窄于MBIR和HIR。
- 作者认为超低剂量DLR可在保持肺结节体积测量准确性的同时降低辐射剂量,尤其适用于需长期随访的患者。
核心结论
- 超低剂量CT下,DLR对人工磨玻璃结节体积测量的准确性高于MBIR和HIR,并在观察者间和观察者内一致性上显示更窄的一致性界限。
研究设计与样本
- 体模研究:胸部体模肺尖及中肺野放置直径6 mm和10 mm、-660 HU的人工磨玻璃结节,用320排CT以6.3 mAs超低剂量扫描,分别经DLR、MBIR、HIR重建,半自动测量结节体积并计算绝对体积误差百分比(APEvol),用Tukey-Kramer法比较,用Bland-Altman分析评价观察者间和观察者内变异性。
诊断标准与关键数值
- 体积测量准确性指标为绝对体积误差百分比(APEvol);10 mm中肺野结节APEvol:DLR 1.36% vs HIR 8.01%(p=0.0022) vs MBIR 7.30%(p=0.0053),DLR显著低于HIR和MBIR;观察者间和观察者内一致性以Bland-Altman一致性界限评价,DLR的一致性界限窄于MBIR和HIR。
临床应用与阅片要点
- 作者认为超低剂量DLR可在保持肺结节体积测量准确性的同时降低辐射剂量,尤其适用于需长期随访的患者。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The lung nodule volume determined by CT is used for nodule diagnoses and monitoring tumor responses to therapy. Increased image noise on low-dose CT degrades the measurement accuracy of the lung nodule volume. We compared the volumetric accuracy among deep-learning reconstruction (DLR), model-based iterative reconstruction (MBIR), and hybrid iterative reconstruction (HIR) at an ultra-low-dose setting.
METHODS: Artificial ground-glass nodules (6 mm and 10 mm diameters, -660 HU) placed at the lung-apex and the middle-lung field in chest phantom were scanned by 320-row CT with the ultra-low-dose setting of 6.3 mAs. Each scan data set was reconstructed by DLR, MBIR, and HIR. The volumes of nodules were measured semi-automatically, and the absolute percent volumetric error (APEvol) was calculated. The APEvol provided by each reconstruction were compared by the Tukey-Kramer method. Inter- and intraobserver variabilities were evaluated by a Bland-Altman analysis with limits of agreements.
RESULTS: DLR provided a lower APEvol compared to MBIR and HIR. The APEvol of DLR (1.36%) was significantly lower than those of the HIR (8.01%, p = 0.0022) and MBIR (7.30%, p = 0.0053) on a 10-mm-diameter middle-lung nodule. DLR showed narrower limits of agreement compared to MBIR and HIR in the inter- and intraobserver agreement of the volumetric measurement.
CONCLUSIONS: DLR showed higher accuracy compared to MBIR and HIR for the volumetric measurement of artificial ground-glass nodules by ultra-low-dose CT.
ADVANCES IN KNOWLEDGE: DLR with ultra-low-dose setting allows a reduction of dose exposure, maintaining accuracy for the volumetry of lung nodule, especially in patients which deserve a long-term follow-up.
引用
- Mikayama, Shirasaka, Kojima, et al. Deep-learning reconstruction for ultra-low-dose lung CT: Volumetric measurement accuracy and reproducibility of artificial ground-glass nodules in a phantom study. Br J Radiol. 2021. PMID: 34908478. doi:10.1259/bjr.20210915
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。