心外膜脂肪组织衰减与脂肪衰减指数:体模研究及光子计数探测器CT活体测量
研究在体模和30例患者中评估光子计数探测器CT不同虚拟单能图像能级对心外膜脂肪组织衰减和脂肪衰减指数的影响。体模脂肪插入物参考EID CT 120 kV衰减为-69 HU,PCD CT在70 keV最接近;患者心外膜脂肪衰减随能级升高而增加,FAI亦增加。结论认为VMI能级和对比剂增强影响测量,70 keV可近似常规多色测量。
要点
- PCD CT虚拟单能图像能级影响心外膜脂肪衰减和FAI测量,55至80 keV间衰减随能级升高而增加。
- 体模脂肪插入物在70 keV时PCD CT衰减最接近EID CT 120 kV参考值。
- 对比剂增强扫描与非增强扫描的心外膜脂肪衰减差异随能级升高而减小。
核心结论
- PCD CT的心外膜脂肪衰减和FAI测量受VMI能级和对比剂增强影响。
- 使用70 keV的VMI重建可提供接近常规多色测量的脂肪衰减。
研究设计与样本
- 体模研究:两种尺寸的拟人体模含脂肪插入物,在第二代双源PCD CT和常规能量积分探测器CT 120 kV下成像。
- 回顾性评估30例患者(11女、19男;平均年龄48±10岁;Agatston评分<60),均行ECG门控非增强钙化积分扫描和增强冠状动脉CTA。
- 重建55至80 keV、间隔5 keV的虚拟单能图像,手动测量非增强和增强图像的心外膜脂肪衰减,并用半自动软件计算FAI。
关键影像表现/征象
- 体模脂肪插入物参考EID CT衰减为-69 HU;PCD CT最接近值出现在70 keV(小体模-69 HU、大体模-70 HU)。
- 患者非增强采集心外膜脂肪衰减从55 keV的-111±11 HU增至80 keV的-82±9 HU。
- 患者增强采集心外膜脂肪衰减从55 keV的-104±11 HU增至80 keV的-81±9 HU。
- 非增强与增强扫描的平均衰减差从7±12 HU降至1±10 HU(随能级升高)。
- FAI随能级升高而增加:右冠状动脉从55 keV的-89±8 HU增至80 keV的-77±12 HU;左前降支从-95±11 HU增至-85±11 HU;回旋支从-87±10 HU增至-80±12 HU。
临床应用与阅片要点
- 使用70 keV的VMI重建可使PCD CT脂肪衰减接近常规多色测量。
- 该发现可能有助于标准化PCD CT评估冠状动脉周围炎症以衡量患者心脏风险。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Epicardial adipose tissue (EAT) attenuation is a vascular inflammation marker predictive of adverse cardiac events. The fat attenuation index (FAI) assesses fat attenuation for predefined coronary segments. Photon-counting detector (PCD) CT uses routine virtual monoenergetic image (VMI) reconstructions. VMI energy level may affect EAT attenuation and FAI measurements. OBJECTIVE. The purpose of this article was to assess EAT attenuation and FAI measurements at different monoenergetic energy levels in patients undergoing coronary CTA using a first-generation whole-body dual-source PCD CT scanner. METHODS. An anthropomorphic phantom at two sizes with a fat insert was imaged on a first-generation dual-source PCD CT scanner and, as a reference, on a conventional energy-integrating detector (EID) CT scanner at 120 kV. Thirty patients (11 women, 19 men; mean age, 48 ± 10 years; Agatston score < 60) who underwent an ECG-gated unenhanced calcium-scoring scan and contrast-enhanced coronary CTA by PCD CT were retrospectively evaluated. VMIs from 55 to 80 keV at 5-keV increments were reconstructed. EAT attenuation was manually measured on unenhanced and contrast-enhanced images. FAI was calculated using semiautomated software. RESULTS. The attenuation of the phantom fat insert was -69 HU for the reference EID CT; the closest attenuation for PCD CT was observed at 70 keV for the small (-69 HU) and large (-70 HU) phantoms. In patients, EAT attenuation increased for unenhanced acquisition from -111 ± 11 HU at 55 keV to -82 ± 9 HU at 80 keV and for contrast-enhanced acquisition from -104 ± 11 HU at 55 keV to -81 ± 9 HU at 80 keV. The mean attenuation difference between unenhanced and contrast-enhanced scans decreased with increasing energy level (from 7 ± 12 HU to 1 ± 10 HU). The FAI increased from -89 ± 8 HU at 55 keV to -77 ± 12 HU at 80 keV for the right coronary artery, -95 ± 11 HU at 55 keV to -85 ± 11 HU at 80 keV for the left anterior descending artery, and -87 ± 10 HU at 55 keV to -80 ± 12 HU at 80 keV for the circumflex artery. CONCLUSION. EAT attenuation and FAI measurements using PCD CT are impacted by VMI energy level and contrast enhancement. Use of VMI reconstruction at 70 keV provides fat attenuation approximating conventional polychromatic measurements. CLINICAL IMPACT. The findings may help standardize evaluation of pericoronary inflammation by PCD CT as a measure of patients' cardiac risk.