新生儿数字化胸部X线摄影——是否应使用附加铜滤过
本研究通过受控析因设计实验,在体模上探讨附加铜滤过(0、0.1、0.2 mm Cu)对新生儿移动胸部X线摄影图像质量和辐射剂量的影响。结果显示附加铜滤过使VGA、CNR和ESD均显著降低(p<0.01);0.1 mm Cu时ESD降低60%,VGA降低14%,CNR降低20%。使用0.1 mm Cu并相应增加kV和mAs可在相同剂量下获得更好图像质量(直接探测器)或在降低剂量下保持相同图像质量(托盘内),因此推荐使用0.1 mm Cu。
要点
- 附加铜滤过显著降低体模表面入射剂量,0.1 mm Cu时ESD降低60%。
- 0.1 mm Cu配合增加kV和mAs可在相同剂量下改善图像质量或降低剂量下保持图像质量。
- 推荐新生儿胸部摄影使用0.1 mm Cu并相应增加kV和mAs。
核心结论
- 使用附加铜滤过(同时增加mAs)可显著降低辐射剂量,且对图像质量无不利影响;推荐新生儿胸部摄影使用0.1 mm Cu并相应增加kV和mAs。
研究设计与样本
- 受控析因设计实验,使用 anthropomorphic phantom(人体等效体模),变量包括铜滤过(0、0.1、0.2 mm Cu)、曝光参数、源-像距和图像受体位置(直接/托盘)。
诊断标准与关键数值
- 图像质量采用绝对视觉分级分析(VGA)和对比噪声比(CNR)评估,入射表面剂量(ESD)用电离室在中心X射线束中测量。
- 使用附加铜滤过时VGA、CNR和ESD均显著降低(p<0.01)。
- 0.1 mm Cu时,ESD降低60%,VGA降低14%,CNR降低20%。
- 与无铜滤过时培养箱成像的最佳参数组合相比,0.1 mm Cu配合增加kV和mAs在直接探测器下同等剂量获得更好图像质量,在托盘内降低剂量下获得同等图像质量。
临床应用与阅片要点
- 作者推荐新生儿胸部成像使用0.1 mm Cu并相应增加kV和mAs。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Copper filtration removes lower energy X-ray photons, which do not enhance image quality but would otherwise contribute to patient radiation dose. This study explores the use of additional copper filtration for neonatal mobile chest imaging.
METHODS: A controlled factorial-designed experiment was used to determine the effect of independent variables on image quality and radiation dose. These variables included: copper filtration (0 Cu, 0.1 Cu and 0.2 Cu), exposure factors, source-to-image distance and image receptor position (direct / tray). Image quality was evaluated using absolute visual grading analysis (VGA) and contrast-to-noise ratio (CNR) and entrance surface dose (ESD) was derived using an ionising chamber within the central X-ray beam.
RESULTS: VGA, CNR and ESD significantly reduced (p < 0.01) when using added copper filtration. For 0.1 Cu, the percentage reduction was much greater for ESD (60%) than for VGA (14%) and CNR (20%), respectively. When compared to the optimal combinations of parameters for incubator imaging using no copper filtration, an increase in kV and mAs when using 0.1-mm Cu resulted in better image quality at the same radiation dose (direct) or, equal image quality at reduced dose (in-tray). The use of 0.1-mm Cu for neonatal chest imaging with a corresponding increase in kV and mAs is therefore recommended.
CONCLUSION: Using additional copper filtration significantly reduces radiation dose (at increased mAs) without a detrimental effect on image quality.
ADVANCES IN KNOWLEDGE: This is the first study, using an anthropomorphic phantom, to explore the use of additional Cu for digital radiography neonatal chest imaging and therefore helps inform practice to standardise and optimise this imaging examination.