颞骨齿科伪影:迭代金属伪影去除与虚拟单能重建联合应用
回顾性纳入44例头颈部CT严重齿科金属伪影患者,比较标准、VMI、IMAR及VMIIMAR重建。IMAR降低伪影效果优于VMI,VMIIMAR在100keV时对伪影区软组织衰减和可解释性有小幅额外改善,两名读片者分别有56.8%和59.1%的检查首选100keV VMIIMAR。
要点
- IMAR对齿科金属伪影的去除效果明显优于VMI。
- VMIIMAR在100keV时对伪影区软组织的显示有小幅额外改善。
- VMIIMAR在多数检查中被读片者首选,可作为严重齿科伪影的补充重建方案。
关键影像表现/征象
- 标准图像高密度伪影校正衰减为521.0HU,IMAR降至48.2HU,VMIIMAR为32.8-91.0HU。
- 标准图像低密度伪影校正衰减为-455.1HU,IMAR为-37.3HU,VMIIMAR为-17.8至-36.9HU。
- 标准图像伪影区软组织校正噪声为58.7HU,IMAR为11.0HU,VMIIMAR为5.8-45.6HU。
- 软组织可解释性中位数:标准图像1.2,IMAR 3.7,VMIIMAR 2.0-3.8。
诊断标准与关键数值
- 软组织可解释性采用1-5分评分,两名放射科医师独立评估后汇总分析。
- VMIIMAR与IMAR相比,仅在100keV时伪影区软组织校正衰减和软组织可解释性显著改善(p<.05)。
- 两名读片者分别在56.8%和59.1%的检查中首选100keV的VMIIMAR重建。
临床应用与阅片要点
- 对严重齿科金属伪影的头颈部CT,可优先采用IMAR重建以降低伪影。
- 可在IMAR基础上补充100keV的VMIIMAR重建,以进一步改善伪影区软组织的显示。
- VMI单能重建降低齿科金属伪影的效果有限,不宜单独替代IMAR。
局限性
- 本研究为回顾性研究,样本量44例,仅纳入严重齿科金属伪影患者。
- 仅在分滤单源双能CT系统上评估,结论外推至其他双能CT平台需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Head and neck CT can be limited by dental hardware artifact. Both postprocessing-based iterative metal artifact reduction (IMAR) and virtual monoenergetic imaging (VMI) reconstruction in dual-energy CT (DECT) can reduce metal artifact. Their combination is poorly described for single-source DECT systems. OBJECTIVE. The purpose of this study was to compare metal artifact reduction between VMI, IMAR, and their combination (VMIIMAR) in split-filter single-source DECT of patients with severe dental hardware artifact. METHODS. This retrospective study included 44 patients (nine woman, 35 men; mean age, 66.0 ± 10.4 years) who underwent head and neck CT and had severe dental hardware artifact. Standard, VMI, IMAR, and VMIIMAR images were generated; VMI and VMIIMAR were performed at 40, 70, 100, 120, 150, and 190 keV. ROIs were placed to measure corrected attenuation in pronounced hyperattenuating and hypoattenuating artifacts and artifact-impaired soft tissue and to measure corrected artifact-impaired soft-tissue noise. Two radiologists independently assessed soft-tissue interpretability (1-5 scale), and pooled ratings were analyzed. Readers selected the preferred reconstruction for each patient. RESULTS. Mean hyperattenuating artifact-corrected attenuation was 521.0 HU for standard imaging, 496.4-892.2 HU for VMI, 48.2 HU for IMAR, and 32.8-91.0 HU for VMIIMAR. Mean hypoattenuating artifact-corrected attenuation was -455.1 HU for standard imaging, -408.5 to -679.9 HU for VMI, -37.3 for IMAR, and -17.8 to -36.9 HU for VMIIMAR. Mean artifact-impaired soft tissue-corrected attenuation was 10.8 HU for standard imaging, -0.6 to 24.9 HU for VMI, 4.3 HU for IMAR, and -2.0 to 7.8 HU for VMIIMAR. Mean artifact-impaired soft tissue-corrected noise was 58.7 HU for standard imaging, 38.2 to 129.7 HU for VMI, 11.0 HU for IMAR, and 5.8 to 45.6 HU for VMIIMAR. Median soft-tissue interpretability was 1.2 for standard imaging, 1.1-1.2 for VMI, 3.7 for IMAR, and 2.0-3.8 for VMIIMAR. Artifact-impaired soft tissue-corrected attenuation and soft-tissue interpretability significantly improved (p < .05) for VMIIMAR versus IMAR only at 100 keV. The two readers preferred VMIIMAR at 100 keV in 56.8% and 59.1% of examinations. CONCLUSION. For reducing severe artifact due to dental material, IMAR has greater effect than VMI. Though the results for IMAR and VMIIMAR were similar overall, VMIIMAR had a small benefit at 100 keV. CLINICAL IMPACT. VMI and IMAR techniques in split-filter DECT may be combined for clinical head and neck imaging to reduce artifact from dental hardware and improve image quality.
引用
- Schmidt, Grunz, Petritsch, et al. Combination of Iterative Metal Artifact Reduction and Virtual Monoenergetic Reconstruction Using Split-Filter Dual-Energy CT in Patients With Dental Artifact on Head and Neck CT. AJR Am J Roentgenol. 2021. PMID: 34755521. doi:10.2214/AJR.21.26772
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。