文献速递腹部MR

三种MRCP采集方法定量胆道测量的比较

Comparison of quantitative 3D magnetic resonance cholangiography measurements obtained using three different image acquisition methods.
PubMed 文献 · Abdom Radiol (NY) · 2021-11-02

该回顾性研究比较了3D FSE SENSE、3D CS-FSE和3D GRASE三种MRCP采集方法获得的定量胆道测量参数。结果显示3D FSE与3D CS-FSE在左右肝管直径、胆管体积、胆管数目和长度及扩张总长度上具有良好至优秀的一致性(ICC 0.83-0.93),而3D GRASE仅胆管体积和扩张总数与3D FSE一致性良好。3D CS-FSE可在显著缩短采集时间下提供与3D FSE相当的胆管直径和整体胆管定量指标。

要点

  • 3D CS-FSE与3D FSE的胆管直径及整体定量指标一致性良好,可替代后者。
  • 3D GRASE与3D FSE仅胆管体积和扩张总数一致性良好,其他指标一致性不足。
  • 运动伪影导致后处理失败率在3D GRASE中最高(40%),需注意图像质量。

核心结论

  • 3D CS-FSE MRCP在显著缩短采集时间下,可产生与3D FSE MRCP相当的胆管直径指标和整体胆管定量。
  • 3D GRASE MRCP与3D FSE MRCP仅胆管体积和扩张总数一致性良好,其他指标一致性不足。

研究设计与样本

  • 采用MRCP+ Tuning Threshold算法分割3D胆道模型并量化多项指标。
  • 使用单测量双向混合效应组内相关系数、Bland-Altman分析和Wilcoxon符号秩检验进行比较。

诊断标准与关键数值

  • 3D FSE与3D CS-FSE在左右肝管直径、胆管体积、胆管数目和长度、扩张总长度上ICC为0.83-0.93。
  • 3D FSE与3D GRASE仅在胆管体积(ICC 0.75)和扩张总数(ICC 0.77)上一致性良好。

临床应用与阅片要点

  • 日常阅片可采用3D CS-FSE序列替代3D FSE进行胆道定量测量,以缩短检查时间。
  • 使用3D GRASE进行胆道定量时需谨慎解读直径和长度等指标,因其与3D FSE一致性有限。
  • 注意运动伪影可能导致后处理失败,3D GRASE失败率较高(40%),需评估图像质量。

局限性

  • 回顾性研究设计。
  • 部分数据集因运动伪影被排除,可能引入选择偏倚。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare quantitative biliary measurements obtained with three different magnetic resonance cholangiopancreatography (MRCP) acquisition methods.

METHODS: This retrospective study was IRB-approved. Patients with combinations of clinically indicated 3D FSE MRCP with sensitivity encoding (SENSE), 3D FSE SENSE MRCP with compressed sensing (CS-FSE; acceleration factor 8), and 3D gradient and spin-echo (GRASE) MRCP, acquired between October 2018 and March 2020, were included. The MRCP + Tuning Threshold algorithm (Perspectum Ltd., Oxford, UK) was used to segment 3D biliary models from MRCP data, with multiple metrics quantified from the models. Single measure, two-way, mixed-effects intra-class correlations, Bland-Altman analyses, and Wilcoxon signed-rank tests were used to compare quantitative measurements.

RESULTS: From 160 MRCP datasets (25 3D FSE, 67 3D CS-FSE, 68 3D GRASE) in 69 patients, 48 datasets (7 [28%] 3D FSE, 14 [21%] 3D CS-FSE, 27 [40%] 3D GRASE) failed post-processing due to motion artifacts. The remaining 112 MRCP datasets (18 3D FSE, 53 3D CS-FSE, 41 3D GRASE) from 60 patients were included in the analysis. There was good to excellent agreement between 3D FSE and 3D CS-FSE MRCP for diameter of the left and right hepatic ducts, biliary volume, number and length of ducts, and total length of dilations (ICC: 0.83-0.93). The only metrics that exhibited good agreement between 3D FSE and 3D GRASE MRCP were biliary volume (ICC: 0.75) and total number of dilations (ICC: 0.77).

CONCLUSION: 3D CS-FSE MRCP produces comparable biliary diameter metrics and global duct quantification to 3D FSE MRCP at a significantly reduced acquisition time.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Mahalingam, Ralli, Trout, et al. Comparison of quantitative 3D magnetic resonance cholangiography measurements obtained using three different image acquisition methods. Abdom Radiol (NY). 2021. PMID: 34729626. doi:10.1007/s00261-021-03330-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。