儿童颅缝早闭CT:低剂量方案在维持诊断图像质量方面的机构经验
本研究评估了某机构用低剂量CT方案(100 kV、5 mAs、迭代重建)替代常规儿童脑CT方案诊断颅缝早闭的依从性、辐射剂量降低和图像质量。结果显示低剂量方案依从性高,CTDIvol降至1.1 mGy,有效剂量0.06-0.22 mSv,CTDIvol降低98%,有效剂量最多降低67倍,两位儿科神经放射学家对全部50例低剂量CT图像诊断信心高且认为无需增加辐射剂量。
要点
- 低剂量颅缝早闭CT方案(100 kV、5 mAs、迭代重建)可大幅降低辐射剂量。
- CTDIvol为1.1 mGy,有效剂量0.06-0.22 mSv,与4视图头颅X线相当。
- 儿科神经放射学家对低剂量CT诊断颅缝早闭的信心高,认为无需增加剂量。
诊断标准与关键数值
- 低剂量CT方案CTDIvol为1.1 mGy(所有0-9岁患者),有效剂量范围0.06-0.22 mSv。
- CTDIvol降低98%,有效剂量最多降低67倍。
- 两位放射学家对全部50例低剂量CT研究诊断颅缝早闭的信心高,且认为无需增加辐射剂量。
临床应用与阅片要点
- 在儿童颅缝早闭诊断和随访中,可考虑采用100 kV、5 mAs结合迭代重建的低剂量CT方案。
- 低剂量CT有效剂量与4视图头颅X线相当,可作为替代常规脑CT的选项。
局限性
- 单中心研究,样本量有限(低剂量组135例,图像质量子集50例)。
- 仅两位儿科神经放射学家参与图像质量主观评分。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Children with craniosynostosis may undergo multiple computed tomography (CT) examinations for diagnosis and post-treatment follow-up, resulting in cumulative radiation exposure.
OBJECTIVE: To reduce the risks associated with radiation exposure, we evaluated the compliance, radiation dose reduction and clinical image quality of a lower-dose CT protocol for pediatric craniosynostosis implemented at our institution.
MATERIALS AND METHODS: The standard of care at our institution was modified to replace pediatric head CT protocols with a lower-dose CT protocol utilizing 100 kV, 5 mAs and iterative reconstruction. Study-ordered, protocol-utilized and radiation-dose indices were collected for studies performed with routine pediatric brain protocols (n=22) and with the lower-dose CT protocol (n=135). Two pediatric neuroradiologists evaluated image quality in a subset (n=50) of the lower-dose CT studies by scoring visualization of cranial structures, confidence of diagnosis and the need for more radiation dose.
RESULTS: During the 30-month period, the lower-dose CT protocol had high compliance, with 2/137 studies performed with routine brain protocols. With the lower-dose CT protocol, volume CT dose index (CTDIvol) was 1.1 mGy for all patients (0-9 years old) and effective dose ranged from 0.06 to 0.22 mSv, comparable to a 4-view skull radiography examination. CTDIvol was reduced by 98% and effective dose was reduced up to 67-fold. Confidence in diagnosing craniosynostosis was high and more radiation dose was considered unnecessary in all studies (n=50) by both radiologists.
CONCLUSION: Replacing the routine pediatric brain CT protocol with a lower-dose CT craniosynostosis protocol substantially reduced radiation exposure without compromising image quality or diagnostic confidence.
引用
- Barreto, Tuna, Rajderkar, et al. Pediatric craniosynostosis computed tomography: an institutional experience in reducing radiation dose while maintaining diagnostic image quality. Pediatr Radiol. 2021. PMID: 34731286. doi:10.1007/s00247-021-05205-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。