文献速递腹部MR

5.0T与1.5T及3.0T上腹部MRI图像质量与诊断信心的前瞻性双队列研究

5.0 T versus 1.5 T and 3.0 T upper abdominal MRI: a prospective two‑cohort study of image quality and diagnostic confidence.
PubMed 文献 · Eur Radiol · 2026-09-28

本研究前瞻性纳入348例上腹部MRI患者队列及31名健康志愿者队列,比较5.0T与1.5T、3.0T的图像质量与诊断信心。结果显示5.0T在解剖细节、信噪比、整体图像质量及患者诊断信心方面显著优于1.5T和3.0T,但心脏搏动伪影、T2WI肝脏信号均匀性及DWI几何变形更明显。

要点

  • 5.0T上腹部MRI临床可行,肝成像解剖细节和信噪比显著优于1.5T与3.0T。
  • 5.0T在T2WI心脏搏动伪影和肝脏信号均匀性评分最低,DWI变形和鬼影更重,阅片时需注意。
  • 5.0T在患者队列中诊断信心最高,提示其对肝脏病变评估有潜在优势。

核心结论

  • 5.0T上腹部MRI以肝脏成像为主,临床可行,在解剖细节、信噪比、整体图像质量及诊断信心方面优于1.5T和3.0T。
  • 5.0T在心脏搏动伪影、T2WI肝脏信号不均匀和DWI几何变形方面需进一步优化。

研究设计与样本

  • 前瞻性双队列研究,患者队列纳入348例连续行上腹部MRI检查的患者。
  • 健康志愿者队列纳入31名受试者,均在同一5.0T、3.0T和1.5T三台设备上接受扫描。
  • 由两名放射科医师独立评估主观图像质量,采用5分Likert量表评价解剖细节、伪影、整体图像质量和诊断信心。
  • 客观指标测量信噪比和对比噪声比,参考标准为不同场强间总体及两两比较。

关键影像表现/征象

  • 5.0T在解剖细节、信噪比和整体图像质量评分方面显著优于1.5T和3.0T。
  • 5.0T在T2WI上心脏搏动伪影和肝脏信号均匀性评分最低,DWI上几何变形和鬼影伪影更明显。
  • 患者队列中5.0T获得最高诊断信心评分。

诊断标准与关键数值

  • 客观指标左/右/平均肝脏信噪比、肌肉/脾脏信噪比、肝脾对比噪声比在5.0T、3.0T、1.5T间呈显著梯度趋势,5.0T最高,均p<0.001。
  • 主观评分方面,5.0T与1.5T和3.0T相比,在T2WI心脏搏动伪影和肝脏信号均匀性上评分最低,DWI变形和鬼影更明显,均p<0.05。
  • 5.0T在解剖细节、信噪比和整体图像质量方面显著优于1.5T和3.0T,均p<0.05。
  • 患者队列中5.0T诊断信心评分最高,p<0.05。

临床应用与阅片要点

  • 5.0T上腹部MRI可用于肝脏临床检查,其高信噪比和解剖细节有利于病灶显示和图像判读。
  • 在T2WI和DWI序列判读时需警惕心脏搏动伪影、肝脏信号不均和几何变形对图像质量的干扰。
  • 报告时可提及5.0T带来的诊断信心提高,但对DWI变形区域的结构评估应结合解剖序列综合判断。

局限性

  • 原文指出5.0T在心脏搏动伪影、T2WI肝脏信号不均匀和DWI几何变形方面需进一步优化。
  • 健康志愿者队列样本量仅为31名。
  • 原文未提及多中心设计及患者队列的疾病构成分层。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare 5.0 T magnetic resonance imaging (MRI) image quality versus 1.5 T and 3.0 T in patients and healthy volunteers, so as to validate its clinical feasibility.

MATERIALS AND METHODS: This prospective study included a patient cohort (348 consecutive patients undergoing upper abdominal MRI) and a healthy volunteer cohort (31 volunteers scanned on all three scanners). Two radiologists independently assessed image quality subjectively (5-point Likert scale: anatomical detail, artifacts, overall image quality, diagnostic confidence) and measured signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective and objective metrics were compared across field strengths using overall and pairwise comparisons.

RESULTS: Objective metrics (left/right/mean liver SNR, muscle/spleen SNR, liver-spleen CNR) showed a significant gradient of 5.0 T > 3.0 T > 1.5 T in both cohorts (all p < 0.001). Subjectively, compared with 1.5 T and 3.0 T, 5.0 T received the lowest scores for cardiac motion artifact and liver signal uniformity on T2-weighted imaging (T2WI) and showed greater distortion and ghosting artifacts on diffusion-weighted imaging (DWI) (all p < 0.05). Nevertheless, 5.0 T was significantly superior in anatomical detail, SNR, and overall image quality (all p < 0.05). In patients, 5.0 T also achieved the highest diagnostic confidence (p < 0.05). Results from healthy volunteers were largely consistent with those from the patient cohort.

CONCLUSION: 5.0 T upper abdominal MRI, with a primary focus on liver imaging, is clinically feasible, offering superior anatomical detail, SNR, overall image quality, and diagnostic confidence over 1.5 T and 3.0 T, though cardiac motion artifacts, liver signal inhomogeneity on T2WI, and geometric distortion on DWI require further optimization.

KEY POINTS: Question Is 5.0-T MRI superior to 1.5 T and 3.0 T for liver imaging, and does its diagnostic benefit differ between diseased and non-diseased livers? Findings 5.0 T provided significantly higher SNR, image quality, and diagnostic confidence than 1.5 T and 3.0 T, despite increased cardiac motion and liver signal inhomogeneity. Clinical relevance 5.0-T MRI significantly improves liver image quality and diagnostic confidence over 3.0 T and 1.5 T, supporting its clinical utility despite increased artifacts that require further optimization.

来源:Eur Radiol,查看原文。版权归原作者所有。