手持超声设备对成人肺炎肺超声的图像质量比较

Image Quality of Handheld Ultrasound Devices for Lung Ultrasound in Adult Patients with Pneumonia.
PubMed 文献 · J Ultrasound Med · 2026-09-24

本研究对成人肺炎患者使用三种市售手持超声设备进行肺超声检查,比较图像质量及专家判读一致性。结果显示 Lumify 与 Vscan Air 的总体图像质量高于 Butterfly iQ+,但图像质量评分的观察者间一致性较低。超声病理征象的观察者间一致性因征象类型而异。

要点

  • Lumify 与 Vscan Air 图像质量总体优于 Butterfly iQ+
  • 图像质量评分观察者间一致性低,报告时应谨慎
  • 胸腔积液观察者间一致性较高,空气支气管征较低

核心结论

  • 手持超声设备中 Vscan Air 与 Lumify 对肺炎患者的图像质量优于 Butterfly iQ+
  • 图像质量的观察者间一致性低,超声病理征象的观察者间一致性总体为中等

研究设计与样本

  • 前瞻性探索性研究,对 11 例成人肺炎患者使用 3 种手持超声设备采集图像
  • 共获得 38 段超声片段,随机化后由 3 名经验丰富的胸部超声医师独立盲法评分
  • 图像质量按 1–5 分评价,临床使用充分性按 1–3 分评价,并记录超声病理征象

关键影像表现/征象

  • 不同手持超声设备在肺炎患者肺超声中显示图像质量差异,Lumify 总分最高,Butterfly iQ+ 最低
  • 超声病理征象判读中胸腔积液观察者间一致性较高,空气支气管征一致性较低

诊断标准与关键数值

  • 总体图像质量评分:Lumify 3.60(95% CI 3.40–3.80),Vscan Air 3.32(95% CI 3.11–3.53),Butterfly iQ+ 2.78(95% CI 2.55–3.01)
  • 图像质量评分的总体观察者间一致性为 27.6%
  • 胸腔积液观察者间 kappa 为 0.58–0.84,空气支气管征观察者间 kappa 为 0.15–0.34

临床应用与阅片要点

  • 不同手持超声设备图像质量存在差异,选择设备时需结合临床需求并重视肺炎相关征象的可判读性
  • 胸腔积液的判读在不同观察者间较稳定,空气支气管征判读一致性差,报告时需注意主观差异
  • 图像质量评分一致性低,提示不应单独依赖质量评分作为设备临床适用性的唯一依据

局限性

  • 探索性研究,样本量小,仅纳入 11 例患者
  • 图像质量评分观察者间一致性低,可能影响质量比较结果的稳定性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Handheld ultrasound devices (HHUDs) have increased accessibility and convenience of ultrasonography, but also raised concern regarding image quality and sufficiency for clinical use. This explorative study aims to compare image quality between 3 commercially available HHUDs in patients with pneumonia, and to compare expert agreement on image quality and sonopathological findings.

METHODS: Patients were scanned with 3 different HHUDs. Clips were stored, anonymized, duplicated, and randomized. Three blinded, experienced thoracic ultrasound operators independently rated the clips for image quality (1-5), sufficiency for clinical use (1-3), and presence of sonopathological findings.

RESULTS: Eleven patients were scanned, yielding 38 ultrasound clips and 76 ratings per rater. Overall image quality was 3.60 (95% CI 3.40-3.80) for Lumify, 3.32 (95% CI 3.11-3.53) for Vscan Air, and 2.78 (95% CI 2.55-3.01) for Butterfly iQ+. Overall agreement on image quality ratings was low (27.6%). Interrater agreement for specific sonopathological findings ranged from slight for air bronchograms (kappa 0.15-0.34) to substantial for pleural effusions (kappa 0.58-0.84).

CONCLUSION: This explorative study demonstrated a higher image quality in Vscan Air and Lumify compared to Butterfly iQ+ in patients with pneumonia. However, interrater agreement on image quality was low, while interrater agreement on sonopathological findings was moderate.