光子计数CT与锥形束CT在人工耳蜗术后成像的比较

Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging.
PubMed 文献 · Eur J Radiol · 2026-09-18

回顾性研究纳入224例人工耳蜗术后影像数据,比较光子计数CT(PCD CT)与CBCT的图像质量及辐射剂量。PCD CT总体图像质量中位数显著高于CBCT(4.5比2.5),所有协议CNR均更高;超低剂量PCD CT有效剂量0.16 mSv接近CBCT的0.15 mSv,同时保持较高图像质量。

要点

  • PCD CT术后电极评估图像质量显著优于CBCT,伪影更少,优先用于精细定位。
  • 超低剂量PCD CT剂量与CBCT接近,可用于儿童、需复查或行动不便者。
  • PCD CT常规剂量评分最高,低剂量与超低剂量主观评分无显著差异。

核心结论

  • PCD CT人工耳蜗术后图像质量高于CBCT,伪影更少;超低剂量PCD CT有效剂量0.16 mSv接近CBCT的0.15 mSv,仍保留较高图像质量评分。
  • 超低剂量PCD CT可能更有利于更精确的术后评估、个性化设备调整及对放射敏感或无法配合患者的随访。

研究设计与样本

  • PCD CT采用三种剂量协议:常规剂量、低剂量、超低剂量;由两名放射科医师独立采用5分Likert量表评价主观图像质量。
  • 客观图像质量采用对比噪声比(CNR)评估;组间比较采用Mann-Whitney U与Kruskal-Wallis检验,线性混合效应模型分析图像质量预测因素。

关键影像表现/征象

  • PCD CT在全部成像参数上评分均高于CBCT,效应量r为0.66至0.84。
  • PCD CT各协议的CNR均高于CBCT,差异有统计学意义(p<0.001)。
  • PCD CT不同剂量协议间主观图像质量存在差异(H=13.05;p=0.0015;η²_H=0.10),常规剂量最高,低剂量与超低剂量之间无显著差异。

诊断标准与关键数值

  • 总体主观图像质量中位数:PCD CT为4.5,CBCT为2.5,差异有统计学意义(p<0.001;r=0.82)。
  • 估计有效剂量:PCD CT常规剂量0.3 mSv,低剂量0.24 mSv,超低剂量0.16 mSv,CBCT为0.15 mSv。
  • 样本量224例,来源于196例患者,平均年龄58.8±16.5岁,女性118例。

临床应用与阅片要点

  • 改用PCD CT可提高人工耳蜗电极阵列显示清晰度,有助于判断电极位置与有无折叠。
  • 需减少辐射时优先选超低剂量PCD CT,其剂量接近CBCT而图像质量仍优于CBCT。
  • CBCT在电极整体评估中仍可使用,但需注意其主观评分较低且伪影可能干扰精细判断。

局限性

  • 本研究为回顾性设计,存在选择偏倚可能。
  • 样本来自单中心,结果外推性可能受限。
  • 主观图像质量评分为Likert分级,存在阅片者主观差异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare image-quality and radiation dose between photon-counting CT (PCD CT) and cone-beam CT (CBCT) for postoperative cochlear implantation (CI) imaging.

METHODS AND MATERIALS: This retrospective study included 224 postoperative CI datasets from July 2019 to February 2026, including 112 PCD CT scans using three dosage protocols (regular dose [RD], low dose [LD], ultra-low dose [ULD]) and 112 120-kV CBCT scans. Two radiologists independently assessed image-quality using 5-point Likert scales. Objective image-quality was assessed using contrast-to-noise ratio (CNR). Groups were compared using Mann-Whitney U and Kruskal-Wallis tests. Predictors of image-quality were analyzed with linear mixed-effects models.

RESULTS: Among 224 datasets from 196 patients (mean age, 58.8 ± 16.5 years; 118 women), PCD CT showed higher subjective overall image-quality than CBCT (median: 4.5 vs. 2.5; p < 0.001; r = 0.82) and higher ratings across all other imaging parameters (p < 0.001; r ≤ 0.66-0.84). Image-quality differed between PCD CT protocols (H = 13.05; p = 0.0015; η2_H = 0.10), with higher ratings for RD, while LD and ULD did not differ. CNR was higher across all PCD CT protocols than for CBCT (p < 0.001). Estimated effective dose was 0.3 mSv for RD, 0.24 mSv for LD and 0.16 mSv for ULD, and 0.15 mSv for CBCT. ULD showed improved image-quality with an estimated effective dose close to that of CBCT.

CONCLUSION: PCD CT was associated with higher image-quality and fewer artifacts than CBCT. ULD PC DCT retained higher image-quality ratings at a numerically similar dose, and may be advantageous for more precise postoperative evaluation, personalized device adjustments, and follow-up for radiosensitive or immobile patients.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Burck, Tillmanns, Bernatz, et al. Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging. Eur J Radiol. 2026. PMID: 42785130. doi:10.1016/j.ejrad.2026.113256

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。