文献速递腹部神经技术与安全CT

光子计数CT虚拟单能图像CT值是否仍受管电压影响:体模与患者研究

Influence of X-ray Tube Voltage on Computed Tomography Values and Contrast-Noise Ratios in Virtual Monoenergetic Images in routine clinical Photon Counting Detector Computed Tomography: Does kV Still Matter?
PubMed 文献 · Invest Radiol · 2026-09-15

本研究在光子计数探测器CT上使用腹部体模及脑CTA患者数据,评估不同管电压(70/90/120/140 kV)下虚拟单能图像(VMI)的CT值稳定性与对比噪声比剂量效率。结果显示45–70 keV范围内CT值偏差在预设10%容差内;小体模及低keV时70/90 kV的剂量效率更高。

要点

  • 对比增强相关能量范围45–70 keV内,VMI的CT值对管电压基本不敏感。
  • 小体型患者或头部CTA等小部位检查,低管电压(70/90 kV)可能提高剂量效率。
  • 高keV(如100 keV)及低碘浓度时CT值偏差可超过10%,定量对比需谨慎。

核心结论

  • 在45–70 keV能量范围内,光子计数CT的VMI CT值在所有体模尺寸和管电压下均稳定在预设10%容差以内。
  • 小体模和低keV条件下,70/90 kV较120/140 kV具有更高的辐射剂量效率,但该优势在高keV和大体模时消失。

研究设计与样本

  • 体模研究采用光谱腹部体模QSA-543(含2个可选脂肪环),在光子计数CT(NAEOTOM Alpha)上以70/90/120/140 kV扫描。
  • 体模内含2/5/10/15 mg/mL碘棒、肌肉/脂肪/肝脏等效组织及400 mg/mL钙棒,采用标准腹部协议。
  • 重建多能量图像及40–100 keV(5 keV递增)的VMI,以ROI分析计算不同kV间平均CT值的百分比偏差ΔCT。
  • 患者研究为回顾性,脑CTA分别采用140 kV(n=42)或90 kV(n=25),重建55 keV VMI,比较脑动脉相对脑室的ΔCT与CNRD。
  • 边缘扩散函数和噪声功率谱在体模研究中用于评估不同kV设置下的空间分辨率与噪声特性。

关键影像表现/征象

  • 40–70 keV能量范围内,所有碘棒和钙棒的ΔCT均≤8.4%,其中脂肪和肝脏在45–70 keV时ΔCT≤10%。
  • 100 keV时低浓度碘的CT值偏差增大,5 mg/mL碘的ΔCT可达36%,对应平均CT值54 HU中的19 HU偏差。
  • 小体模40–60 keV时,70/90 kV的CNRD较120/140 kV高10%至13%,边缘扩散函数和噪声功率谱在各kV间一致。
  • 患者55 keV VMI中,90 kV的CNRD较140 kV高17%至29%,但多数差异未达统计学显著。

诊断标准与关键数值

  • 临床可接受的CT值偏差上限预设为ΔCT=10%。
  • 碘和钙的CT值在所有kV设置下于40–70 keV保持稳定,ΔCT≤8.4%。
  • 100 keV时5 mg/mL碘的ΔCT最高达36%,对应偏差19 HU(平均CT值54 HU)。
  • 小体模40–60 keV时70/90 kV的CNRD较120/140 kV高10%至13%。
  • 患者55 keV VMI中90 kV的CNRD较140 kV高17%至29%,但一般无统计学显著性。

临床应用与阅片要点

  • 光子计数CT的对比增强检查中,45–70 keV的VMI CT值可用于跨不同管电压扫描间的定量比较。
  • 当检查仅需VMI而无需其他光谱结果时,小体型患者或头部CTA等小部位可考虑采用70/90 kV低管电压。
  • 阅片时若使用100 keV等高能量VMI对低碘浓度区域做定量,应意识到CT值可能随管电压出现超过10%的偏差。

局限性

  • 患者研究为探索性设计,两组间CNRD差异多数无统计学显著性。
  • 体模研究采用单一光子计数CT设备型号,结果对其他平台的外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Conventional energy integrating detector (EID) CT provides polyenergetic images where CT values depend on the tube voltage (kV), hampering quantitative analysis across different CT scans. To mitigate this problem, spectral CT such as photon-counting detector (PCD) CT offers virtual monoenergetic images (VMI) where CT values depend only on the keV level but, ideally, not on the tube voltage. The purpose of this study was to investigate the kV dependence of VMI CT values in PCD-CT across a wide range of tissue types, keV levels, and kV settings for various phantom sizes, and to determine which kV settings yield the optimal contrast-to-noise ratio (CNR) for a given VMI energy and phantom size. The phantom study was complemented with an exploratory patient study.

MATERIALS AND METHODS: A spectral abdomen phantom (QSA-543; QRM) with 2 optional fat rings was scanned on a PCD-CT (NAEOTOM Alpha; Siemens Healthineers). The phantom contained rods with 2/5/10/15 mg/mL iodine (QRM); muscle/fat/liver equivalent tissue; and 400 mg/mL calcium (Gammex Sun Nuclear). A standard abdomen protocol was used at tube voltages of 70/90/120/140 kV. Polyenergetic images and VMIs (range: 40 to 100 keV in 5 keV increments) were reconstructed. For each VMI energy (keV), the percentage deviation ΔCT(%) of the mean CT value across the different kV settings was determined for each rod using region-of-interest (ROI) analysis. A ΔCT of 10% was considered the clinically acceptable upper limit. For each kV setting, the iodine CNR was calculated and normalized to the radiation dose (CTDIvol) to obtain CNRD. Higher CNRD indicated higher radiation dose efficiency. Edge spread functions (EFS) and noise power spectra (NPS) were evaluated for various kV settings. In the retrospective patient study, a standard cerebral CT angiography (CTA) protocol was applied at either 140 kV (n = 42) or 90 kV (n = 25), with VMI reconstructed at 55 keV. Both the ΔCT(%) and the CNRD relative to the ventricles were determined for various cerebral arteries. The 2 groups were compared using unpaired t tests.

RESULTS: For all iodine and calcium rods and all phantom sizes, CT values remained stable within the predefined 10% tolerance across all kV settings in the VMI energy range of 40 to 70 keV, with ΔCT ≤8.4%. At 100 keV, ΔCT increased for low iodine concentration, reaching up to 36%, corresponding to a deviation of 19 HU from the mean CT value of 54 HU for 5 mg/mL iodine. Adipose tissue and liver exhibited ΔCT ≤10% at 45 to 70 keV. At 40 keV, deviations were larger but remained small in absolute terms (11.0 HU for liver and 19.6 HU for adipose tissue in the large phantom). For the small phantom and 40-60 keV, CNRD was 10% to 13% higher for 70/90 kV compared with 120/140 kV. EFS and NPS were consistent across kV settings. This advantage was lost at higher keV and larger phantoms. In patient VMIs at 55 keV, CNRD was 17% to 29% higher at 90 kV than at 140 kV, albeit generally not statistically significant.

CONCLUSIONS: In VMIs acquired with PCD-CT, CT values were stable within the predefined 10% tolerance for all phantom sizes and tube voltages in the energy range of 45 to 70 keV, which is relevant for contrast-enhanced CT. This represents a significant step towards the meaningful comparison of CT values across different CT scans. At low keV and small phantom sizes, scans at 70 or 90 kV demonstrated greater radiation dose efficiency compared with scans at 120 or 140 kV. Therefore, for routine contrast-enhanced CT examinations of small patients or small body parts (eg, CTA of the head), scanning at low tube voltage may be considered if no spectral results other than VMIs are required.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Jeukens, Pinckaers, Martens, et al. Influence of X-ray Tube Voltage on Computed Tomography Values and Contrast-Noise Ratios in Virtual Monoenergetic Images in routine clinical Photon Counting Detector Computed Tomography: Does kV Still Matter? Invest Radiol. 2026. PMID: 42740544. doi:10.1097/RLI.0000000000001312

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。