AI辅助超声引导机器人用于快速血管通路建立
结果显示体模穿刺成功率为93%,人体血管定位成功率为97%,但股总动脉段确认率仅为20%。
要点
- 初学者使用该系统体模穿刺成功率达93%,中位时间44秒
- 志愿者股血管定位成功率97%,中位时间25秒
- 股总动脉分叉以上节段确认率仅20%,大孔径动脉操作前需注意
核心结论
- AI-GUIDE-LL使无经验临床人员能够快速、准确完成股血管定位和穿刺,支持在军事前沿及民用即时救治场景中应用。
- 股动脉分叉以上节段特异性定位仍不可靠,尚需进一步改进。
研究设计与样本
- 使用5个人体CT来源腹股沟体模进行股动脉穿刺,每次最多允许3次进针,以抽吸到液体为参考标准。
- 随后在28名健康志愿者中由28名临床人员使用该设备定位股血管并评估股总动脉节段识别。
- 操作成功由专家回顾超声图像和视频记录判定。
关键影像表现/征象
- 体模试验中正确穿刺成功率为93%(140/150次),中位完成时间44秒,平均尝试1.1次。
- 志愿者血管定位成功率为97%(93/96根血管),中位时间25秒。
诊断标准与关键数值
- 体模穿刺成功定义为抽吸到液体,成功率为93%(140/150次)。
- 股总动脉节段确认成功率仅为20%。
临床应用与阅片要点
- 该设备可帮助超声引导经验不足的临床人员在紧急情况下快速定位股血管并进行穿刺。
- 需要大孔径动脉操作或要求分叉以上定位时,不能仅依赖该设备确认节段。
局限性
- 体模试验的参考标准为液体抽吸,与真实临床穿刺条件可能有差异。
- 股总动脉分叉以上节段确认率仅20%,节段特异性定位不可靠。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Vascular cannulation enables life-saving resuscitation in traumatic hemorrhage. The AI-GUIDE-LL is a handheld ultrasound-guided robotic device enabling inexperienced medical providers to obtain femoral vascular access. Device efficacy has been demonstrated in expert-operated porcine models, though performance among inexperienced operators and in humans is unknown. We evaluated the AI-GUIDE-LL's performance among novice clinicians to target human femoral vessels and accurately deploy needles in phantoms.
MATERIALS AND METHODS: This single-site prospective study recruited medical professionals without ultrasound-guided procedural experience. Users attempted femoral artery needle insertion in five human CT-derived inguinal phantoms using fluid aspiration as the reference standard, with three needle insertions permitted per phantom. Subsequently, they used the device to target healthy human volunteer femoral vessels, assessing target vessel localization and common femoral segment identification. Success was defined by expert review of ultrasound images and video recordings.
RESULTS: Thirty clinicians (seven advanced practice providers, five EMTs, nine physicians, and nine registered nurses) achieved correct needle insertion in phantoms in 93% (140/150) of trials, with a median completion time of 44 s and a mean of 1.1 attempts. 28 volunteers were scanned by 28 clinicians. Clinicians successfully targeted 97% (93/96) of vessels with a median time of 25 s. Common femoral segment confirmation was achieved in only 20% of trials.
CONCLUSION: The AI-GUIDE-LL enables rapid, accurate femoral vessel localization and needle insertion among novice clinicians, supporting feasibility for field-forward military and point-of-care civilian settings. Reliable segment-specific localization above the femoral bifurcation, necessary for large-bore arterial procedures, remains an important limitation requiring further work.
KEY POINTS: Question AI-GUIDE-LL is a handheld robotic system that integrates artificial intelligence and ultrasound guidance to assist non-expert users in performing femoral vascular access for rapid point-of-care interventions. Findings Clinicians with limited ultrasound experience successfully localized femoral vessels, achieving 93% phantom needle insertion success and 97% target vessel localization in human trials. Clinical relevance AI-GUIDE-LL may enable clinicians with limited ultrasound training to perform timely femoral cannulation for life-saving interventions, including fluid resuscitation in traumatic hemorrhage, particularly in military or resource-limited environments.