超声弹性成像ARFI的安全性与生物效应综述
本文为叙述性综述,梳理ARFI超声弹性成像的生物效应证据,重点分析动物实验中长时间推力脉冲带来的潜在机械与热效应风险。结果提示ARFI安全性不能仅凭传统诊断超声的MI/TI标准推断,需结合脉冲时长、累积暴露、组织敏感性及造影剂使用等因素评估。
要点
- ARFI长脉冲可能降低肺毛细血管出血和心律失常机械效应阈值
- 同时使用超声造影剂可能增加心脏机械生物效应易感性
- 骨表面温升在ARFI辐照下可高于肝组织,需关注邻近骨骼区域
核心结论
- ARFI安全性不能仅依据传统诊断超声的MI和TI标准推断
- ARFI特异性因素如脉冲时长、累积暴露、组织易损性和脱靶暴露需纳入安全评估
研究设计与样本
- 叙述性综述,梳理ARFI生物效应直接实验研究并对照AIUM、BMUS、WFUMB国际安全指南
- 证据主要来源于动物模型实验研究
关键影像表现/征象
- ARFI推力脉冲声场可超出预设聚焦区域,在肝脏或乳腺扫查时可能累及邻近肺或心脏
诊断标准与关键数值
- 动物模型证据提示ARFI延长脉冲时长可能降低肺毛细血管出血和心脏节律紊乱的机械效应阈值
- 兔肺模型特定测试条件下未显示ARFI所致肺出血显著增加
- 兔活体测量显示ARFI辐照时骨表面温升高于肝组织
临床应用与阅片要点
- 肝或乳腺ARFI扫查时需注意推力脉冲对邻近肺、心脏的非预期暴露
- 使用超声造影剂时需警惕心脏机械生物效应易感性增加
局限性
- 现有证据主要来自动物模型
- 原文为叙述性综述,未报告系统检索与纳入排除标准
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Acoustic radiation force impulse (ARFI) imaging is an ultrasound elastography technique that enables quantitative assessment of tissue mechanical properties. However, unlike conventional B-mode imaging, ARFI elastography uses longer and higher-intensity push pulses to induce tissue displacement. This feature raises important questions regarding whether the conventional mechanical index (MI) and thermal index (TI) are sufficient to assess the safety of ARFI exposure. In addition, the acoustic field associated with ARFI push pulses may extend beyond the intended focal region, creating the potential for unintended effects to adjacent vulnerable organs, such as the lung or the heart during liver scanning or breast scanning. This narrative review examines the current evidence on the safety of ARFI, with emphasis on direct experimental studies of ARFI bioeffects and their interpretation in the context of international safety guidance from organizations such as the American Institute of Ultrasound in Medicine (AIUM), the British Medical Ultrasound Society (BMUS), and the World Federation for Ultrasound in Medicine and Biology (WFUMB). Available evidence, derived primarily from animal models, suggests that the prolonged pulse duration of ARFI may lower the threshold for certain mechanical bioeffects, including pulmonary capillary hemorrhage and cardiac rhythm disturbances. Concurrent use of ultrasound contrast agents may increase susceptibility to cardiac mechanical bioeffects, whereas no significant increase in ARFI-induced pulmonary hemorrhage was demonstrated in a rabbit lung model under the tested conditions. In vivo rabbit measurements also showed greater temperature rise at the bone surface than in liver tissue during ARFI exposure. These findings indicate that the safety profile of ARFI cannot be inferred solely from criteria developed for conventional diagnostic ultrasound. ARFI-specific factors, including pulse duration, cumulative exposure, tissue vulnerability, off-target exposure, and contrast-agent use, should be considered in safety assessment and clinical practice.
引用
- Sasanuma, Takayama, Rifu, et al. Safety and Bioeffects of Acoustic Radiation Force Impulse (ARFI) in Ultrasound Elastography: A Narrative Review. Ultrasound Med Biol. 2026. PMID: 42731931. doi:10.1016/j.ultrasmedbio.2026.08.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。