PubMed 文献 · Radiol Imaging Cancer · 2026-01-01

本研究为前瞻性多中心随机试验,纳入165例胸腹部可疑病变患者,比较CT引导机器人辅助与徒手穿刺活检的安全性与可行性。结果显示机器人组技术成功率更高(97.56% vs 62.65%),靶向误差更小,CT扫描次数与调针次数更少,诊断率均为100%。肺部活检中机器人组并发症率更低(4.88% vs 16.87%),穿刺时间两组相当。

要点

  • 机器人辅助活检技术成功率显著高于徒手操作(97.56% vs 62.65%),可减少穿刺过程中CT扫描次数与调针次数。
  • 两组诊断率均为100%,机器人辅助未能提高诊断产出,但肺部活检并发症率明显降低。
  • 机器人组靶向误差均值1.7 mm优于徒手组4.5 mm,提示几何精度更高,穿刺时间两组无差异。

核心结论

  • 机器人辅助活检在几何精度上优于徒手,技术成功率更高,调针与CT扫描更少,诊断效能非劣效,穿刺时间相当,肺部并发症更少。

研究设计与样本

  • 前瞻性多中心随机对照试验,2023年7月至2024年4月在三家大学教学医院进行,纳入165例胸腹部可疑病变患者,随机分为机器人组(n=82)与徒手组(n=83)。

诊断标准与关键数值

  • 技术成功率:机器人组97.56%(80/82),徒手组62.65%(52/83),P < .001。
  • 靶向误差(平均欧氏偏差):机器人组1.7 mm ± 1.1,徒手组4.5 mm ± 3.9,P < .001。
  • CT扫描次数:机器人组平均4.3 ± 1.9次,徒手组5.2 ± 2.3次,P = .002;调针次数:机器人组0.7 ± 0.7次,徒手组1.6 ± 1.6次,P = .003。
  • 诊断产出:两组均为100%(82/82与83/83);穿刺时间中位数:机器人组5.5分钟 ± 4.3,徒手组4.8分钟 ± 7.0,P = .50。
  • 肺部活检并发症率:机器人组4.88%(4/82),徒手组16.87%(14/83),P = .014。

临床应用与阅片要点

  • 机器人辅助适于对靶向精度要求高的胸腹部穿刺活检,可减少重复扫描和调针次数,降低肺部穿刺并发症。
  • 虽然机器人组技术成功率高,但诊断产出无差异,选择时需权衡设备成本与临床获益。

局限性

  • 原文未报告长期随访或迟发性并发症,结果可能受术者经验与机构设备条件影响。
  • 两组穿刺时间差异无统计学意义,机器人系统未能缩短总操作时间。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To compare safety and feasibility between a novel CT-guided robotic system and the conventional freehand technique for puncture biopsy of thoracoabdominal lesions. Materials and Methods In this prospective multicenter randomized trial, individuals with suspected lesions were enrolled between July 2023 and April 2024 across three university teaching hospitals and randomized to the robot-assisted group (n = 82) or the freehand group (n = 83). Procedure outcomes included the technical success rate, targeting error, number of CT scans and needle adjustments, puncture time, and complications. Descriptive and inferential statistics were calculated. Results A total of 165 participants (mean age, 60 years ± 10 [SD]; 83 male) were included. Compared with the freehand group, the robot-assisted group demonstrated a higher technical success rate (97.56% [80 of 82] vs 62.65% [52 of 83], P < .001), lower targeting error (mean Euclidean deviation: 1.7 mm ± 1.1 vs 4.5 mm ± 3.9, P < .001), and fewer CT scans (mean, 4.3 ± 1.9 vs 5.2 ± 2.3; P = .002) and needle adjustments (mean, 0.7 ± 0.7 vs 1.6 ± 1.6; P = .003). Despite differences in geometric precision, both groups achieved 100% (82 of 82 and 83 of 83) diagnostic yield. The median puncture time was comparable between groups (5.5 minutes ± 4.3 vs 4.8 minutes ± 7.0, P = .50). During lung biopsies, the robot-assisted approach yielded fewer complications compared with the freehand approach (4.88% [four of 82] vs 16.87% [14 of 83], P = .014). Conclusion Compared with the freehand approach, robot-assisted biopsy yielded greater precision and reduced adjustments and complications while demonstrating noninferior diagnostic efficacy and comparable duration. Keywords: Robotic Needle Insertion, Biopsy, Thoracoabdominal Lesions, Robot-assisted Biopsy, CT-guided Intervention, Percutaneous Needle Biopsy, Randomized Controlled Trial, Algorithm Development, CT, Clinical Testing, Interventional-Body, Biopsy/Needle Aspiration, Percutaneous, Thorax, Abdomen/GI, Liver, Lung, Kidney ©RSNA, 2026.