文献速递神经儿科影像技术与安全CT

三维3D打印儿童脑CT体模的构建与验证

Development and validation of a novel three-dimensional patient-specific pediatric brain phantom for computed tomography neuroimaging research.
PubMed 文献 · Pediatr Radiol · 2026-09-03

本研究基于5岁儿童头部CT数据,经3-D Slicer分割并3D打印模具,制作环氧树脂基脑实质模拟体模。在80–140 kVp下扫描验证CT值,并采用PHITS蒙特卡洛模拟验证能量响应。结果显示体模保留了96.85%的原始脑体积,120 kVp下CT值与真实脑实质相差4.28 HU。

要点

  • 该体模在120 kVp时CT值为44.94±1.38 HU,接近儿童脑实质的40.66±4.13 HU。
  • 脑模拟材料CT值随管电压升高而增加,从80 kVp的18.55 HU升至140 kVp的51.04 HU。
  • 蒙特卡洛模拟显示衰减系数在15–150 keV诊断能谱范围内与参考值偏差在±5%以内。

核心结论

  • 证实了制作保留解剖几何形态且具有临床相关CT衰减值的三维个体化儿童脑组件的可行性。
  • 为后续将颅骨及其他结构集成为模块化完整头部体模提供了方法学基础。

研究设计与样本

  • 基于1例5岁儿童头部CT数据构建个体化脑体模,原始脑体积1165.89 cm³,体模脑体积1129.15 cm³,体积保留率96.85%。
  • 采用3-D Slicer分割并导出STL文件,3D打印PLA母模后制作硅胶模具,以环氧树脂体系改性制备均质脑模拟软组织材料。
  • 体模在80 kVp、100 kVp、120 kVp、140 kVp四个管电压下扫描,采用ROI测量双侧半球CT值。
  • 采用PHITS蒙特卡洛模拟在15–150 keV单能透射条件下计算衰减系数,并与PhyX-PSD及NIST XCOM参考数据库比较。

关键影像表现/征象

  • 体模保持了双侧大脑半球的基本形态,几何外形忠实于原始患者解剖结构。
  • 在120 kVp匹配管电压下,体模平均CT值为44.94±1.38 HU,原始儿童脑实质为40.66±4.13 HU,绝对差值为4.28 HU。
  • 体模CT值随管电压升高而递增,从80 kVp的18.55±1.14 HU增至140 kVp的51.04±0.93 HU。

诊断标准与关键数值

  • 同一匹配管电压120 kVp下体模与真实脑实质CT值绝对差值为4.28 HU。
  • CT衰减系数在80–140 kVp范围内由0.1664±0.0002降至0.1445±0.0001 cm²/g。
  • PHITS模拟衰减系数在15–150 keV能谱范围内与理论参考值的相对偏差在±5%以内。

临床应用与阅片要点

  • 该体模可用于儿科头颅CT扫描方案的优化与辐射剂量评估,提供可重复的定量基准。
  • 影像科可参考该体模在不同管电压下的CT值变化规律,校准颅脑扫描条件。
  • 后续集成颅骨等结构时需注意模块化拼接对整体CT值均匀性的影响。

局限性

  • 当前体模仅包含均质脑实质组件,尚未集成颅骨及其他组织结构。
  • 原始数据仅来源于1例5岁儿童,个体化解剖代表性有限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Pediatric head computed tomography (CT) is widely used for urgent neuroimaging, but children's higher radiosensitivity makes protocol optimization and quantitative validation essential. Many available phantoms lack patient-specific pediatric anatomy and realistic brain-mimicking attenuation, limiting their relevance for neuro-CT research and benchmarking.

OBJECTIVE: To develop and validate a CT-derived, patient-specific pediatric brain phantom that preserves anatomically faithful geometry and reproduces brain-mimicking soft-tissue attenuation with physics-validated energy dependence.

MATERIALS AND METHODS: A fully anonymized head CT of a 5-year-old child was segmented in 3-D Slicer, exported to standard tessellation language (STL), and partitioned into hemispheres. A polylactic acid (PLA) master was three-dimensional (D)-printed to fabricate a reinforced silicone mold. A homogeneous brain-mimicking soft-tissue surrogate was produced by modifying an epoxy system with acetone. The assembled phantom was scanned at 80, 100, 120, and 140 kilovolt peak (kVp). Region of interest (ROI)-based CT numbers were measured across both hemispheres. Computational attenuation verification was performed using Particle and Heavy Ion Transport code System (PHITS) Monte Carlo simulations with a mono-energetic transmission method over 15-150 kiloelectronvolt (keV) and compared with PhyX-Photon Shielding and Dosimetry (PhyX-PSD) and National Institute of Standards and Technology XCOM database (NIST/XCOM) reference.

RESULTS: The fabricated phantom preserved the main bilateral morphology and retained 96.85% of the original patient-derived brain volume, with volumes of 1,129.15 cm3 and 1,165.89 cm3, respectively. CT-number and CT-derived mass attenuation coefficient values are reported as mean±standard error of the mean (SEM). At the matched tube voltage of 120 kVp, the mean CT number was 44.94±1.38 Hounsfield units (HU) in the phantom and 40.66±4.13 HU in the source pediatric brain parenchyma, corresponding to an absolute difference of 4.28 HU. Across the complete phantom acquisition series, mean CT numbers increased from 18.55±1.14 HU at 80 kVp to 51.04±0.93 HU at 140 kVp. CT-derived mass attenuation coefficients decreased from 0.1664±0.0002 to 0.1445±0.0001 cm2 g-1 over 80-140 kVp. PHITS-derived attenuation coefficients closely matched the theoretical reference, with relative differences within ±5% across the predefined 15-150 keV diagnostic photon-energy range.

CONCLUSIONS: This work establishes the feasibility of producing a 3-D patient-specific pediatric brain component with preserved geometry and clinically relevant bulk CT attenuation, providing a methodological basis for subsequent integration of the skull and other structures into a modular complete-head phantom.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Sekkat, El Hafiane, El Mouden, et al. Development and validation of a novel three-dimensional patient-specific pediatric brain phantom for computed tomography neuroimaging research. Pediatr Radiol. 2026. PMID: 42690320. doi:10.1007/s00247-026-06768-y

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。