文献速递腹部介入CT

肝锥形束CT迭代运动校正算法在经动脉介入中的评估

Evaluation of an iterative motion-correction algorithm for hepatic cone-beam CT during transarterial interventions.
PubMed 文献 · Eur J Radiol · 2026-08-21

本研究回顾性纳入69例接受TACE或SIRT患者的69套肝CBCT数据,评估迭代运动校正算法对图像质量的影响。每套数据按100迭代增量处理,由3名介入放射科医生独立评分。结果显示200-300迭代水平运动伪影与血管显示最佳,超过500迭代因新伪影增多导致诊断质量下降。

要点

  • 肝CBCT迭代运动校正宜采用200-300迭代,过高迭代会产生新伪影降低诊断质量
  • 迭代水平超过500后诊断质量明显下降,1000迭代时88.4%数据集不可诊断
  • 运动伪影与血管显示评分在200-300迭代时从基线3分提升至4分

核心结论

  • 迭代运动校正在200-300中等迭代水平时显著改善经动脉介入中肝CBCT的图像质量,过度迭代引入新伪影导致质量下降

研究设计与样本

  • 回顾性单中心研究,纳入2018-2021年间69例接受TACE或SIRT患者的69套CBCT数据
  • 每例患者取一套CBCT数据为分析单位,以100迭代增量后处理(It0基线至It1000)
  • 3名介入放射科医生独立评分,采用5分Likert量表评估运动伪影与血管显示,新生成伪影采用0-5分制
  • 采用非参数检验包括Friedman、Wilcoxon符号秩检验、Cochran Q和McNemar检验,Fleiss kappa评估观察者间一致性

关键影像表现/征象

  • 运动伪影中位评分从基线3分改善至200-300迭代时的4分,血管显示同步改善
  • 超过500迭代后因新生成伪影进行性增多致诊断质量下降,1000迭代时非诊断数据集比例达88.4%

诊断标准与关键数值

  • 运动伪影中位评分从基线3分(IQR 1-2)改善至200-300迭代时4分(IQR 1),所有p≤0.001
  • 靶病灶可见率从基线42-44%提升至峰值65-75%,血管树可见率从30-41%提升至54-64%,所有p<0.001
  • 非诊断数据集比例从It300的1.4%持续上升至It1000的88.4%
  • 观察者间一致性Fleiss kappa为0.40-0.70,呈中度至高度一致

临床应用与阅片要点

  • 经动脉介入肝CBCT后处理推荐采用200-300迭代进行运动校正以获得最佳图像质量
  • 阅片时注意过高迭代水平的图像可能产生新伪影,不要误判为真实病变
  • 迭代水平500以上图像诊断可靠性下降,超过1000迭代应谨慎用于诊断

局限性

  • 回顾性单中心研究设计可能存在选择偏倚,样本量69例相对有限

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate an iterative motion-correction algorithm for periinterventional hepatic cone-beam computed tomography (CBCT) regarding its efficacy in reducing motion artifacts, improving vessel depiction, and enhancing diagnostic confidence during transarterial interventions.

MATERIALS AND METHODS: This retrospective single-centre study included 69 CBCT datasets from 69 patients undergoing TACE or SIRT between 2018 and 2021. One CBCT dataset per patient served as the unit of analysis. Each dataset was post-processed with a motion-correction algorithm at 100-iteration increments (It0 = Baseline; It100-It1000). Three interventional radiologists independently assessed motion artifacts (MA) and vessel depiction (VD) using a 5-point Likert scale (higher ratings representing better image quality). Newly generated artifacts (NGA) were assessed on a 0-5 scale ("0″ representing no evidence of NGA, higher ratings representing stronger artifacts). Target lesion (TL) and vessel tree (VT) visibility were defined as yes/no. Non-parametric tests were applied, including Friedman and Wilcoxon signed-rank tests for ordinal ratings and Cochran's Q and McNemar tests for binary visibility outcomes. Interrater reliability was determined using Fleiss' kappa.

RESULTS: Compared with baseline datasets, median ratings for motion artifacts and vessel depiction improved from 3 (IQR 1-2) to 4 (IQR 1) at iteration levels 200-300 (all p ≤ 0.001). Target lesion and vessel tree visibility increased from baseline values of 42-44% and 30-41% to peak rates of 65-75% and 54-64%, respectively, with significant differences across iteration levels for all readers (all p < 0.001). Beyond 500 iterations, diagnostic quality declined due to progressively increasing NGA. The proportion of non-diagnostic datasets increased continuously with higher iteration levels, ranging from 1.4% at It300 to 88.4% at It1000. Interrater agreement across parameters was moderate to substantial (κ = 0.40-0.70).

CONCLUSION: Iterative motion correction significantly improves image quality in hepatic CBCT during transarterial interventions at moderate iteration levels (200-300). Excessive iteration introduces new artifacts, underscoring the importance of optimizing mid-range iteration settings to balance motion correction and artifact generation.