文献速递骨肌介入MR

UTE序列用于MRI引导肌骨穿刺活检

Mri-guided musculoskeletal needle biopsy by using ultrashort echo time sequence.
PubMed 文献 · Eur J Radiol · 2026-08-19

本研究在3.0T PET/MR系统上开发并优化UTE序列,用于MRI引导肌骨病灶穿刺活检。通过与2DT1WI、3DT1WI、bSSFP序列比较,评估穿刺针相关磁敏感伪影及图像质量。结果显示UTE序列显著减少穿刺针伪影,临床病灶可见性与穿刺针清晰度良好。

要点

  • UTE序列可将穿刺针相关磁敏感伪影减少最高达70%。
  • UTE序列临床评估病灶可见性评分为3.50±0.40,穿刺针清晰度评分为4.39±0.45。
  • UTE序列在不同进针角度下伪影形态稳定,利于近骨或短T2病灶穿刺。

核心结论

  • 优化后的UTE序列可显著减少MRI引导肌骨穿刺活检中穿刺针相关磁敏感伪影,改善病灶与穿刺针显示。

研究设计与样本

  • 研究类型为离体模型实验结合前瞻性临床应用验证,设备为3.0T PET/MR系统。
  • 离体实验使用牛大腿模型,穿刺针相对于B0场方向以三种角度插入。
  • UTE序列参数为最小TE 0.11 ms,各向同性1.5 mm体素,重建至0.75 mm,扫描时间22.8秒。
  • 两名放射科医师独立测量伪影并采用5分Likert量表评价图像质量与穿刺针清晰度。
  • 统计方法包括Friedman检验、事后Wilcoxon符号秩检验、Bonferroni校正及组内相关系数。

关键影像表现/征象

  • UTE序列在离体模型中产生的穿刺针伪影小于2DT1WI、3DT1WI和bSSFP等常规序列。
  • 90°插入角相对于B0方向时,UTE序列伪影减少最高达70%。
  • UTE序列中穿刺针伪影形态在不同角度和平面之间保持稳定。
  • 临床应用中UTE序列病灶可见性评分为3.50±0.40,穿刺针清晰度评分为4.39±0.45。
  • UTE序列图像质量评价具有良好至优良的观察者间一致性。

诊断标准与关键数值

  • UTE序列优化参数:最小TE为0.11 ms,各向同性1.5 mm体素,重建至0.75 mm,扫描时间22.8秒。
  • 病灶可见性评分3.50±0.40,穿刺针清晰度评分4.39±0.45,均采用5分Likert量表。
  • 伪影减少幅度最高达70%,发生在穿刺针与B0方向呈90°插入角时。

临床应用与阅片要点

  • UTE序列可推荐用于MRI引导下肌骨病灶穿刺活检尤其是近骨或短T2病灶的定位。
  • UTE序列能清晰显示穿刺针全程,有助于提高穿刺准确性与安全性。
  • UTE序列在不同穿刺角度下伪影稳定,可减少术中反复调整序列的需求。

局限性

  • 离体实验结果与实际活体组织中的磁敏感伪影表现可能存在差异。
  • 原文未提供临床病例的具体样本量及病灶类型的详细分布。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To improve the susceptibility artifacts, efficiency and safety during MRI-guided musculoskeletal needle biopsy by using ultrashort echo time (UTE) sequence.

METHODS: A dedicated UTE sequence was developed and implemented on a 3.0-T PET/MR system.In an ex vivo bovine thigh model, a biopsy needle was inserted at three angles relative to the B0 field direction.UTE was compared with conventional sequences, including 2DT1WI, 3DT1WI, and balanced steady-state free precession(bSSFP). Artifact dimensions parallel and perpendicular to the needle were independently measured by two radiologists, and mean values were analyzed.Subsequently, patients with suspected musculoskeletaltumors underwent MRI-guided needle biopsies using the optimized UTE. Two radiologists independently evaluated image quality and needle conspicuity using a 5-point Likert scale.Differenceswere analyzed using the Friedman test with post hoc Wilcoxon signed-rank tests and Bonferroni correction.Interobserver agreement was calculated using intraclass correlation coefficients.

RESULTS: The UTE sequence was optimized with aminimum TE of 0.11 ms, isotropic 1.5-mm voxels (reconstructed to 0.75 mm), and a 22.8-s acquisition time. Compared with conventional sequences, UTE produced smaller artifacts in the phantom model,with up to 70% reduction at the 90° insertion angle relative to B0.Artifact morphology remained stable across angles and planesinUTE sequences. Clinically, UTE provided adequate lesion conspicuity (3.50 ± 0.40) and excellentneedle visibility (4.39 ± 0.45) , with good-to-excellent interobserver agreement.

CONCLUSIONS: Optimized UTE significantly reduces needle-related susceptibility artifacts and improves lesion and needle visualization during MRI-guided musculoskeletalbiopsy.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Wang, Zhu, Qian, et al. Mri-guided musculoskeletal needle biopsy by using ultrashort echo time sequence. Eur J Radiol. 2026. PMID: 42641418. doi:10.1016/j.ejrad.2026.113172

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。